Changes for page Water Quality Sensors

Last modified by Karry Zhuang on 2025/07/18 16:37

From version 72.14
edited by Karry Zhuang
on 2025/07/16 09:28
Change comment: There is no comment for this version
To version 73.1
edited by Karry Zhuang
on 2025/07/16 09:59
Change comment: There is no comment for this version

Summary

Details

Page properties
Content
... ... @@ -113,7 +113,10 @@
113 113  RS485 signal (K1 default address 0x12; K10 default address 0x11):
114 114  Standard Modbus-RTU protocol, baud rate: 9600; check bit: none; data bit: 8; stop bit: 1
115 115  
116 +(% style="color:red" %)**The following commands are by default for ECK1/ECK10.**
116 116  
118 +(% style="color:red" %)**Please note that EC200 has different commands.**
119 +
117 117  === 1.7.1 Query address ===
118 118  
119 119  
... ... @@ -228,12 +228,16 @@
228 228  0X00 0X00  0X1E 0XEF 0X01 0X14 0X00 0X2B
229 229  )))|(% style="width:72px" %)(((
230 230  0X42
231 -)))|(% style="width:77px" %)0XC8
234 +)))|(% style="width:77px" %)0X59
232 232  
236 +Conductivity calculation: 0X1EEF=7919=>Conductivity=7919μS/cm
237 +Temperature calculation: 0X0114=276=>Temperature=27.6℃
238 +Salinity calculation: 0X002b=43=>Salinity=4.3PSU
233 233  
234 234  
235 235  
236 236  
243 +
237 237  === 1.7.5 Calibration Method ===
238 238  
239 239  ECK1 and ECK10.0