Changes for page Water Quality Sensors

Last modified by Karry Zhuang on 2025/07/18 16:37

From version 71.7
edited by Karry Zhuang
on 2025/07/16 09:19
Change comment: There is no comment for this version
To version 72.3
edited by Karry Zhuang
on 2025/07/16 09:26
Change comment: There is no comment for this version

Summary

Details

Page properties
Content
... ... @@ -17,21 +17,24 @@
17 17  * **Interface**: RS485. 9600 Baud Rate
18 18  
19 19  * **EC Range & Resolution:**
20 -** **ECK1.0** : 0 ~~ 2,000 μS/cm  Resolution: 1 μS/cm
21 -** **ECK10.0** : 10 ~~ 20,000 μS/cm  Resolution: 10 μS/cm
22 -** **EC200.0** : 1 ~~ 200,000 μS/cm  Resolution: 1 μS/cm
20 +** **ECK1.0 :** 0 ~~ 2,000 μS/cm  Resolution: 1 μS/cm
21 +** **ECK10.0 : **10 ~~ 20,000 μS/cm  Resolution: 10 μS/cm
22 +** **EC200 : **1 ~~ 200,000 μS/cm  Resolution: 1 μS/cm
23 23  * **EC Accuracy**: ±1% FS
24 24  * **Salinity measurement range**
25 -** **EC200** :0~~70PSU Resolution: 0.1PSU
25 +** **EC200 :**0~~70PSU Resolution: 0.1PSU
26 26  * **Temperature measurement range**
27 -** **ECK1/ECK10**:-20~~+60℃; Resolution: 0.1℃
28 -** **EC200** :-5~~+80℃; Resolution: 0.1℃
27 +** **ECK1/ECK10:**-20~~+60℃; Resolution: 0.1℃
28 +** **EC200 :**-5~~+80℃; Resolution: 0.1℃
29 29  * **Temperature Accuracy: **±0.5 °C
30 -* Temperature compensation range
30 +* **Temperature compensation range**
31 +** **ECK1/ECK10:**0~~+60℃ (default compensation temperature 25℃)
32 +** **EC200:**-5~~+80℃ (default compensation temperature 25℃)
33 +* **Temperature compensation coefficient:**Default 0.2
31 31  * **Working environment:**
32 32  ** Ambient Temperature: 0–60°C
33 33  ** Relative Humidity: <85% RH(Specifically refers to the cable male and female)
34 -** ECK200.0 Continuous monitoring of cross-section water quality, aquaculture, sewage treatment, environmental protection, pharmaceuticals, food, tap water, seawater and other high conductivity environments
37 +** ECK200 Continuous monitoring of cross-section water quality, aquaculture, sewage treatment, environmental protection, pharmaceuticals, food, tap water, seawater and other high conductivity environments
35 35  * **IP Rated**: IP68
36 36  * **Max Pressure**: 0.6MPa
37 37  
... ... @@ -49,7 +49,7 @@
49 49  
50 50  == 1.4 Mechinical Drawing ==
51 51  
52 - ECK1 and ECK10  ECK200
55 + ECK1 and ECK10  EC200
53 53  
54 54  
55 55  [[image:image-20240714174241-2.png]] [[image:1752564223905-283.png||height="399" width="160"]]
... ... @@ -209,8 +209,26 @@
209 209  **For example**, the returned data is 12 03 04 (% style="color:red" %)**02 AE**(%%) 01 64 B8 D0. 02 AE is converted to decimal 686,  K=1, EC: 686uS/cm,temperature: 35.6℃ Convert the returned data to decimal and divide by 10.
210 210  
211 211  
212 -ECK200
215 +EC200
213 213  
217 +(% border="1" cellspacing="3" style="background-color:#f2f2f2; width:534.333px" %)
218 +|=(% style="width: 42px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 74px; background-color: rgb(79, 129, 189); color: white;" %)Register Address|=(% style="width: 94px; background-color: rgb(79, 129, 189); color: white;" %)Register length|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 77px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
219 +|(% style="width:99px" %)0X01|(% style="width:112px" %)0X03|(% style="width:74px" %)0X00 0X00|(% style="width:94px" %)0X00 0X04|(% style="width:72px" %)(((
220 +0XC5
221 +)))|(% style="width:77px" %)0XC8
222 +
223 +**response:**
224 +
225 +(% border="1" cellspacing="3" style="background-color:#f2f2f2; width:534.333px" %)
226 +|=(% style="width: 42px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 74px; background-color: rgb(79, 129, 189); color: white;" %)Register Address|=(% style="width: 94px; background-color: rgb(79, 129, 189); color: white;" %)Register length|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 77px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
227 +|(% style="width:99px" %)0X01|(% style="width:112px" %)0X03|(% style="width:74px" %)0X00 0X00|(% style="width:94px" %)0X00 0X04|(% style="width:72px" %)(((
228 +0XC5
229 +)))|(% style="width:77px" %)0XC8
230 +
231 +
232 +
233 +
234 +
214 214  === 1.7.5 Calibration Method ===
215 215  
216 216  ECK1 and ECK10.0
... ... @@ -259,7 +259,7 @@
259 259  
260 260  
261 261  
262 -ECK200.0
283 +EC200
263 263  
264 264  For the device with address 01, use 1413uS/cm standard solution to calibrate the first point. Send frame: 1413. Convert hexadecimal to 585. Write 0001, 00 00, 0585 to 0x0120, 0x0121, 0x0122 respectively.
265 265