Changes for page Water Quality Sensors
Last modified by Karry Zhuang on 2025/07/18 16:37
From version 67.1
edited by Karry Zhuang
on 2025/07/15 18:00
on 2025/07/15 18:00
Change comment:
Uploaded new attachment "1752573643879-991.png", version {1}
To version 68.1
edited by Karry Zhuang
on 2025/07/15 18:34
on 2025/07/15 18:34
Change comment:
There is no comment for this version
Summary
-
Page properties (1 modified, 0 added, 0 removed)
Details
- Page properties
-
- Content
-
... ... @@ -948,11 +948,10 @@ 948 948 02 9A is the turbidity value, converted to decimal, it is 666, and then divided by 10, the actual value is 66.6, 02 9A means the current turbidity value is 66.6 NTU 949 949 950 950 951 -= 6. DR- ECKWaterEC Probe =951 += 6. DR-CL Water CL Probe = 952 952 953 953 == 6.1 Specification: == 954 954 955 - 956 956 * **Power Input**: DC7~~30 957 957 958 958 * **Power Consumption** : 0.19W ... ... @@ -974,17 +974,27 @@ 974 974 975 975 * **Max Pressure**: 0.6MPa 976 976 977 -== 2.2 Wiring ==976 +== 6.2 Wiring == 978 978 979 979 [[image:image-20240720172548-2.png||height="348" width="571"]] 980 980 981 -== 2.3 Mechinical Drawing ==980 +== 6.3 Mechinical Drawing == 982 982 983 983 [[image:1752573238705-910.png||height="694" width="278"]] 984 984 984 +== 6.4 Installation == 985 985 986 +Flow-through installation: Use the matching flow slot for installation. The device and the flow slot are installed tightly. 986 986 988 +The measuring end is completely immersed in the measured liquid to ensure a steady flow rate without bubbles. 987 987 990 +It is recommended that the flow rate be controlled at 30-60Lh to ensure the accuracy of the test. 991 + 992 +[[image:1752573643879-991.png||height="360" width="343"]] 993 + 994 + 995 + 996 + 988 988 = 7. Water Quality Sensor Datasheet = 989 989 990 990 * **[[Water Quality Sensor Transmitter Datasheet>>https://www.dropbox.com/scl/fi/9tofocmgapkbddshznumn/Datasheet_WQS-xB-WQS-xS_Water-Quality-Sensor-Transmitter.pdf?rlkey=wxua12ur9swk30rkqnh2boo9z&st=axga6epf&dl=0]]**