... ... @@ -948,11 +948,10 @@
948 948  02 9A is the turbidity value, converted to decimal, it is 666, and then divided by 10, the actual value is 66.6, 02 9A means the current turbidity value is 66.6 NTU
949 949  
950 950  
951 -= 6. DR-ECK Water EC Probe =
951 += 6. DR-CL Water CL Probe =
952 952  
953 953  == 6.1 Specification: ==
954 954  
955 -
956 956  * **Power Input**: DC7~~30
957 957  
958 958  * **Power Consumption** : 0.19W
... ... @@ -974,17 +974,27 @@
974 974  
975 975  * **Max Pressure**: 0.6MPa
976 976  
977 -== 2.2 Wiring ==
976 +== 6.2 Wiring ==
978 978  
979 979  [[image:image-20240720172548-2.png||height="348" width="571"]]
980 980  
981 -== 2.3 Mechinical Drawing ==
980 +== 6.3 Mechinical Drawing ==
982 982  
983 983  [[image:1752573238705-910.png||height="694" width="278"]]
984 984  
984 +== 6.4 Installation ==
985 985  
986 +Flow-through installation: Use the matching flow slot for installation. The device and the flow slot are installed tightly.
986 986  
988 +The measuring end is completely immersed in the measured liquid to ensure a steady flow rate without bubbles.
987 987  
990 +It is recommended that the flow rate be controlled at 30-60Lh to ensure the accuracy of the test.
991 +
992 +[[image:1752573643879-991.png||height="360" width="343"]]
993 +
994 +
995 +
996 +
988 988  = 7.  Water Quality Sensor Datasheet =