Summary

Details

Page properties
Author
... ... @@ -1,1 +1,1 @@
1 -XWiki.Edwin
1 +XWiki.karry
Content
... ... @@ -3,4 +3,113 @@
3 3  {{toc/}}
4 4  
5 5  
6 -= 1. DR-ECK1 Water EC Probe (0 ~~ 2,000 μS/cm) =
6 += 1. DR-ECK Water EC Probe =
7 +
8 +== 1.1 Specification: ==
9 +
10 +* **Power Input**: DC7~~30
11 +* **Power Consumption** : < 0.5W
12 +* **Interface**: RS485. 9600 Baud Rate
13 +* **EC Range & Resolution:**
14 +** **ECK0.01** : 0.02 ~~ 20 μS/cm
15 +** **ECK0.1**: 0.2 ~~ 200.0 μS/cm
16 +** **ECK1.0** : 2 ~~ 2,000 μS/cm  Resolution: 1 μS/cm
17 +** **ECK10.0** : 20 ~~ 20,000 μS/cm  Resolution: 10 μS/cm
18 +* **EC Accuracy**: ±1% FS
19 +* **Temperature Measure Range**: -20 ~~ 60 °C
20 +* **Temperature Accuracy: **±0.5 °C
21 +* **IP Rated**: IP68
22 +* **Max Pressure**: 0.6MPa
23 +
24 +== 1.2 Application for Different Range ==
25 +
26 +[[image:image-20240714173018-1.png]]
27 +
28 +
29 +== 1.3 Wiring ==
30 +
31 +
32 +== 1.4 Mechinical Drawing ==
33 +
34 +[[image:image-20240714174241-2.png]]
35 +
36 +
37 +== 1.5 Installation ==
38 +
39 +
40 +== 1.6 Maintain ==
41 +
42 +
43 +== 1.7 RS485 Commands ==
44 +
45 +
46 +The address of the EC K10 sensor is 11
47 +
48 +The query data command is 11 03 00 00 00 02 C6 9B
49 +
50 +For example, the returned data is 11 03 04 (% style="color:red" %)**02 AE**(%%) 01 64 8B D0. 02 AE is converted to decimal 686,  K=10, EC: 6860uS/cm
51 +
52 +
53 +The address of the EC K1 sensor is 12
54 +
55 +The query data command is 12 03 00 00 00 02 C6 A8
56 +
57 +For example, the returned data is 12 03 04 (% style="color:red" %)**02 AE**(%%) 01 64 B8 D0. 02 AE is converted to decimal 686,  K=1, EC: 686uS/cm.
58 +
59 +
60 += 2. DR-PH01 Water PH Sensor =
61 +
62 +== 2.7 RS485 Commands ==
63 +
64 +
65 +The address of the pH  sensor is 10
66 +
67 +The query data command is 10 03 00 00 00 01 87 4B. After the query, 7 bytes will be returned.
68 +
69 +For example, the returned data is 10 03 02 (% style="color:red" %)**02 AE**(%%) C4 9B.
70 +
71 +02 AE is the pH value, which is converted into decimal to get 686, and then two decimal places are added to get the actual value. 02 AE means the current pH value is 6.86.
72 +
73 +
74 += 3. DR-ORP1 Water ORP Sensor =
75 +
76 +== 3.7 RS485 Commands ==
77 +
78 +
79 +The address of the ORP sensor is 13
80 +
81 +The query data command is 13 03 00 00 00 01 87 78
82 +
83 +For example, the returned data is 13 03 02 (% style="color:red" %)**02 AE**(%%) 80 9B.
84 +
85 +02 AE is the ORP value, converted to decimal, the actual value is 686, 02 AE means the current ORP value is 686mV
86 +
87 +
88 += 4. DR-DO1 Dissolved Oxygen Sensor =
89 +
90 +== 4.7 RS485 Commands ==
91 +
92 +
93 +The address of the dissolved oxygen sensor is 14
94 +
95 +The query data command is 14 03 00 14 00 01 C6 CB
96 +
97 +After the query, 7 bytes will be returned. For example, the returned data is 14 03 02 (% style="color:red" %)**03 78**(%%) B5 55. 03 78 is the value of dissolved oxygen.
98 +
99 +Converted to decimal, it is 888. Add two decimal places to get the actual value. 03 78 means the current dissolved oxygen is 8.88mg/L
100 +
101 +
102 += 5. DR-TS1 Water Turbidity Sensor =
103 +
104 +== 5.7 RS485 Commands ==
105 +
106 +
107 +The address of the dissolved oxygen sensor is 15
108 +
109 +The query data command is 15 03 00 00 00 01 87 1E
110 +
111 +For example, the returned data is 15 03 02 (% style="color:red" %)**02 9A**(%%) 09 4C
112 +
113 +02 9A is the turbidity value, converted to decimal, it is 666, and then divided by 10, the actual value is 66.6, 02 9A means the current turbidity value is 66.6 NTU
114 +
115 +
Copyright ©2010-2024 Dragino Technology Co., LTD. All rights reserved
Dragino Wiki v2.0