... ... @@ -1,1 +1,1 @@
1 -XWiki.Edwin
1 +XWiki.karry
Content
... ... @@ -3,4 +3,96 @@
3 3  {{toc/}}
4 4  
5 5  
6 -= 1. DR-ECK1 Water EC Probe (0 ~~ 2,000 μS/cm) =
6 += 1. DR-ECK Water EC Probe =
7 +
8 +== 1.1 Specification: ==
9 +
10 +* **Power Input**: DC7~~30
11 +* **Power Consumption** : < 0.5W
12 +* **Interface**: RS485. 9600 Baud Rate
13 +* **EC Range & Resolution:**
14 +** **ECK0.01** : 0.02 ~~ 20 μS/cm
15 +** **ECK0.1**: 0.2 ~~ 200.0 μS/cm
16 +** **ECK1.0** : 2 ~~ 2,000 μS/cm  Resolution: 1 μS/cm
17 +** **ECK10.0** : 20 ~~ 20,000 μS/cm  Resolution: 10 μS/cm
18 +* **EC Accuracy**: ±1% FS
19 +* **Temperature Measure Range**: -20 ~~ 60 °C
20 +* **Temperature Accuracy: **±0.5 °C
21 +* **IP Rated**: IP68
22 +* **Max Pressure**: 0.6MPa
23 +
24 +== 1.2 Application for Different Range ==
25 +
26 +[[image:image-20240714173018-1.png]]
27 +
28 +
29 +== 1.3 Wiring ==
30 +
31 +
32 +== 1.4 Mechinical Drawing ==
33 +
34 +[[image:image-20240714174241-2.png]]
35 +
36 +
37 +== 1.5 Installation ==
38 +
39 +
40 +== 1.6 Maintain ==
41 +
42 +
43 +== 1.7 RS485 Commands ==
44 +
45 +The address of the EC K1 sensor is 11
46 +
47 +The query data command is 11 03 00 00 00 02 C6 9B
48 +
49 +For example, the returned data is 11 03 04 02 AE 01 64 8B D0, 02 AE is converted to decimal 686,  K=1, EC: 686uS/cm
50 +
51 +
52 +The address of the EC K1 sensor is 12
53 +
54 +The query data command is 12 03 00 00 00 02 C6 A8
55 +
56 +~; when K=10, EC: 6860uS/cm.
57 +
58 += 2. DR-PH01 Water PH Sensor =
59 +
60 +== 2.7 RS485 Commands ==
61 +
62 +The address of the dissolved oxygen sensor is 12
63 +
64 +The query data command is 14 03 00 14 00 01 C6 CB
65 +
66 +
67 += 3. DR-ORP1 Water ORP Sensor =
68 +
69 +== 3.7 RS485 Commands ==
70 +
71 +
72 +The address of the dissolved oxygen sensor is 13
73 +
74 +The query data command is 14 03 00 14 00 01 C6 CB
75 +
76 +
77 += 4. DR-DO1 Dissolved Oxygen Sensor =
78 +
79 +== 4.7 RS485 Commands ==
80 +
81 +The address of the dissolved oxygen sensor is 14
82 +
83 +The query data command is 14 03 00 14 00 01 C6 CB
84 +
85 +After the query, 7 bytes will be returned. For example, the returned data is 14 03 02 03 78 B5 55. 03 78 is the value of dissolved oxygen.
86 +
87 +Converted to decimal, it is 888. Add two decimal places to get the actual value. 03 78 means the current dissolved oxygen is 8.88mg/L
88 +
89 += 5. DR-TS1 Water Turbidity Sensor =
90 +
91 +== 5.7 RS485 Commands ==
92 +
93 +
94 +The address of the dissolved oxygen sensor is 15
95 +
96 +The query data command is 15 03 00 14 00 01 C6 CB
97 +
98 +