... ... @@ -19,18 +19,20 @@
19 19  * **EC Range & Resolution:**
20 20  ** **ECK0.01** : 0.02 ~~ 20 μS/cm
21 21  ** **ECK0.1**: 0.2 ~~ 200.0 μS/cm
22 -** **ECK1.0** : 0 ~~ 2,000 μS/cm  Resolution: 1 μS/cm
23 -** **ECK10.0** : 10 ~~ 20,000 μS/cm  Resolution: 10 μS/cm
22 +** **ECK1.0** : 2 ~~ 2,000 μS/cm  Resolution: 1 μS/cm
23 +** **ECK10.0** : 20 ~~ 20,000 μS/cm  Resolution: 10 μS/cm
24 24  
25 25  * **EC Accuracy**: ±1% FS
26 +
27 +* **Temperature Measure Range**: -20 ~~ 60 °C
28 +
26 26  * **Temperature Accuracy: **±0.5 °C
27 -* **Working environment:**
28 -** Ambient Temperature: 0–60°C
29 -** Relative Humidity: <85% RH(Specifically refers to the cable male and female)
30 +
30 30  * **IP Rated**: IP68
31 31  
32 32  * **Max Pressure**: 0.6MPa
33 33  
35 +
34 34  == 1.2 Application for Different Range ==
35 35  
36 36  
... ... @@ -40,7 +40,7 @@
40 40  == 1.3 Wiring ==
41 41  
42 42  
43 -[[image:image-20241129142314-1.png||height="352" width="1108"]]
45 +[[image:image-20240720172533-1.png||height="347" width="569"]]
44 44  
45 45  
46 46  == 1.4 Mechinical Drawing ==
... ... @@ -98,6 +98,7 @@
98 98  
99 99  * The equipment should be calibrated before each use. It is recommended to calibrate it every 3 months for long-term use. The calibration frequency should be adjusted appropriately according to different application conditions (degree of dirt in the application, deposition of chemical substances, etc.).
100 100  
103 +
101 101  == 1.7 RS485 Commands ==
102 102  
103 103  
... ... @@ -123,6 +123,7 @@
123 123  |=(% style="width: 100px;background-color:#4F81BD;color:white" %)New address|=(% style="width: 110px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 106px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 93px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 104px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
124 124  |(% style="width:99px" %)0X01|(% style="width:112px" %)0X03|(% style="width:106px" %)0X00|(% style="width:93px" %)0X20|(% style="width:104px" %)0XF0
125 125  
129 +
126 126  === 1.7.2 Change address ===
127 127  
128 128  
... ... @@ -160,6 +160,7 @@
160 160  0X8F
161 161  )))
162 162  
167 +
163 163  === 1.7.4 Query data ===
164 164  
165 165  
... ... @@ -246,6 +246,7 @@
246 246  |=(% style="width: 42px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 68px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 68px;background-color:#4F81BD;color:white" %)CRC16 high
247 247  |(% style="width:99px" %)0X11|(% style="width:112px" %)0X06|(% style="width:135px" %)0X00|(% style="width:126px" %)0X26|(% style="width:85px" %)0X00|(% style="width:1px" %)0X02|(% style="width:1px" %)0XEB|(% style="width:1px" %)0X50
248 248  
254 +
249 249  = 2. DR-PH01 Water PH Sensor =
250 250  
251 251  == 2.1 Specification ==
... ... @@ -267,9 +267,7 @@
267 267  
268 268  * **Temperature measurement error**: ±0.5°C
269 269  
270 -* **Working environment:**
271 -** Ambient Temperature: 0–60°C
272 -** Relative Humidity: <85% RH(Specifically refers to the cable male and female)
276 +* **Temperature Measure Range**: -20 ~~ 60 °C
273 273  
274 274  * **Temperature Accuracy: **±0.5 °C
275 275  
... ... @@ -277,6 +277,7 @@
277 277  
278 278  * **Max Pressure**: 0.6MPa
279 279  
284 +
280 280  == 2.2 Wiring ==
281 281  
282 282  
... ... @@ -338,6 +338,7 @@
338 338  
339 339  * The equipment should be calibrated before each use. For long-term use, it is recommended to calibrate once every 3 months. The calibration frequency should be adjusted appropriately according to different application conditions (degree of dirt in the application, deposition of chemical substances, etc.). After aging, the electrodes should be replaced in time.
340 340  
346 +
341 341  == 2.6 RS485 Commands ==
342 342  
343 343  
... ... @@ -476,14 +476,13 @@
476 476  
477 477  * **Stability**: ≤2mv/24 hours
478 478  
479 -* **Working environment:**
480 -** Ambient Temperature: 0–60°C
481 -** Relative Humidity: <85% RH(Specifically refers to the cable male and female)
485 +* **Equipment working conditions**: Ambient temperature: 0-60°C Relative humidity: <85%RH
482 482  
483 483  * **IP Rated**: IP68
484 484  
485 485  * **Max Pressure**: 0.6MPa
486 486  
491 +
487 487  == 3.2 Wiring ==
488 488  
489 489  
... ... @@ -554,6 +554,7 @@
554 554  |=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)Original address|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 64.75px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)CRC16 high
555 555  |(% style="width:99px" %)0XFE |(% style="width:112px" %)0X03|(% style="width:135px" %)0X00|(% style="width:126px" %)0X50|(% style="width:85px" %)0X00|(% style="width:1px" %)0X00|(% style="width:1px" %)0X51|(% style="width:1px" %)0XD4
556 556  
562 +
557 557  **response:**
558 558  
559 559  (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
... ... @@ -560,6 +560,7 @@
560 560  |=(% style="width: 103.6px;background-color:#4F81BD;color:white" %)New address|=(% style="width: 103.6px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 103.6px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 103.6px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 103.6px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
561 561  |(% style="width:99px" %)0X01|(% style="width:112px" %)0X03|(% style="width:106px" %)0X00|(% style="width:93px" %)0X20|(% style="width:104px" %)0XF0
562 562  
569 +
563 563  === 3.6.2 Change address ===
564 564  
565 565  
... ... @@ -597,6 +597,7 @@
597 597  0X96
598 598  )))
599 599  
607 +
600 600  === 3.6.4 Query data ===
601 601  
602 602  
... ... @@ -649,11 +649,11 @@
649 649  == 4.1 Specification ==
650 650  
651 651  
652 -* **Measuring range**: 0-20mg/L, 050
660 +* **Measuring range**: 0-20mg/L, 0-50°C
653 653  
654 -* **Accuracy**: 3%, ±0.5
662 +* **Accuracy**: 3%, ±0.5°C
655 655  
656 -* **Resolution**: 0.01 mg/L, 0.01
664 +* **Resolution**: 0.01 mg/L, 0.01°C
657 657  
658 658  * **Maximum operating pressure**: 6 bar
659 659  
... ... @@ -661,9 +661,7 @@
661 661  
662 662  * **Power supply voltage**: 5-24V DC
663 663  
664 -* **Working environment:**
665 -** Ambient Temperature: 0–60°C
666 -** Relative Humidity: <85% RH(Specifically refers to the cable male and female)
672 +* **Working environment**: temperature 0-60°C; humidity <95%RH
667 667  
668 668  * **Power consumption**: ≤0.5W
669 669  
... ... @@ -676,10 +676,13 @@
676 676  == 4.3 Impedance requirements for current signals ==
677 677  
678 678  
679 -(% border="1" cellspacing="3" style="background-color:#f2f2f2; width:400px" %)
680 -|(% style="width:132px" %)**Supply Voltage**|(% style="width:67px" %)**9V**|(% style="width:67px" %)**12V**|(% style="width:67px" %)**20V**|(% style="width:67px" %)**24V**
681 -|(% style="width:132px" %)**Max Impedance**|(% style="width:65px" %)**<250Ω**|(% style="width:67px" %)**<400Ω**|(% style="width:67px" %)**<500Ω**|(% style="width:65px" %)**<900Ω**
685 +(% border="1" cellspacing="3" style="background-color:#f2f2f2; width:420px" %)
686 +|=(% style="width: 140px" %)Supply Voltage|=(% style="width: 70px;" %)9V|=(% style="width: 70px;" %)**12V**|=(% style="width: 70px;" %)**20V**|=(% style="width: 70px;" %)**24V**
687 +|(% style="width:137px" %)**Max Impedance**|(% style="width:70px" %)**<250Ω**|(% style="width:68px" %)**<400Ω**|(% style="width:68px" %)**<500Ω**|(% style="width:70px" %)**<900Ω**
682 682  
689 +[[image:image-20240718195414-8.png||height="100" width="575"]]
690 +
691 +
683 683  == 4.4 Mechinical Drawing ==
684 684  
685 685  
... ... @@ -693,6 +693,7 @@
693 693  
694 694  * If the water conditions are complex and you want accurate data, you need to wipe the sensor probe frequently.
695 695  
705 +
696 696  == 4.6 RS485 Commands ==
697 697  
698 698  
... ... @@ -736,6 +736,7 @@
736 736  |=(% style="width: 42px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address low|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
737 737  |(% style="width:99px" %)0X01|(% style="width:72px" %)0X10|(% style="width:64px" %)0X00|(% style="width:68px" %)0X0A|(% style="width:70px" %)0X00|(% style="width:72px" %)0X02|(% style="width:56px" %)0X61|(% style="width:56px" %)0XCA
738 738  
749 +
739 739  === 4.6.3 Query data ===
740 740  
741 741  
... ... @@ -778,7 +778,7 @@
778 778  == 5.1 Specification ==
779 779  
780 780  
781 -* **Measuring range**: 0.1~~1000.0NTU
792 +* **Measuring range**: 0.11000.0NTU
782 782  
783 783  * **Accuracy**: ±5%
784 784  
... ... @@ -786,13 +786,11 @@
786 786  
787 787  * **Stability**: ≤3mV/24 hours
788 788  
789 -* **Output signal**: RS485 (standard Modbus-RTU protocol, device default address: 01)
800 +* **Output signal**: A: 420 mA (current loop)B: RS485 (standard Modbus-RTU protocol, device default address: 01)
790 790  
791 -* **Power supply voltage**: 5~~24V DC (when output signal is RS485), 12~~24V DC (when output signal is 4~~20mA)
802 +* **Power supply voltage**: 524V DC (when output signal is RS485)1224V DC (when output signal is 420mA)
792 792  
793 -* **Working environment:**
794 -** Ambient Temperature: 0–60°C
795 -** Relative Humidity: <85% RH(Specifically refers to the cable male and female)
804 +* **Working environment**: temperature 060°C; humidity ≤ 95%RH
796 796  
797 797  * **Power consumption**: ≤ 0.5W
798 798  
... ... @@ -805,10 +805,11 @@
805 805  == 5.3 Impedance requirements for current signals ==
806 806  
807 807  
808 -(% border="1" cellspacing="3" style="background-color:#f2f2f2; width:400px" %)
809 -|(% style="width:132px" %)**Supply Voltage**|(% style="width:67px" %)**9V**|(% style="width:67px" %)**12V**|(% style="width:67px" %)**20V**|(% style="width:67px" %)**24V**
810 -|(% style="width:132px" %)**Max Impedance**|(% style="width:65px" %)**<250Ω**|(% style="width:67px" %)**<400Ω**|(% style="width:67px" %)**<500Ω**|(% style="width:65px" %)**<900Ω**
817 +(% border="1" cellspacing="3" style="background-color:#f2f2f2; width:420px" %)
818 +|=(% style="width: 140px" %)Supply Voltage|=(% style="width: 70px;" %)9V|=(% style="width: 70px;" %)**12V**|=(% style="width: 70px;" %)**20V**|=(% style="width: 70px;" %)**24V**
819 +|(% style="width:137px" %)**Max Impedance**|(% style="width:70px" %)**<250Ω**|(% style="width:68px" %)**<400Ω**|(% style="width:68px" %)**<500Ω**|(% style="width:70px" %)**<900Ω**
811 811  
821 +
812 812  == 5.4 Mechinical Drawing ==
813 813  
814 814  
... ... @@ -881,9 +881,3 @@
881 881  For example, the returned data is 15 03 02 (% style="color:red" %)**02 9A**(%%) 09 4C
882 882  
883 883  02 9A is the turbidity value, converted to decimal, it is 666, and then divided by 10, the actual value is 66.6, 02 9A means the current turbidity value is 66.6 NTU
884 -
885 -
886 -= 6.  Water Quality Sensor Datasheet =
887 -
888 -
889 -* **[[Water Quality Sensor Transmitter Datasheet>>https://www.dropbox.com/scl/fi/9tofocmgapkbddshznumn/Datasheet_WQS-xB-WQS-xS_Water-Quality-Sensor-Transmitter.pdf?rlkey=wxua12ur9swk30rkqnh2boo9z&st=axga6epf&dl=0]]**
image-20241129142314-1.png
Author
... ... @@ -1,1 +1,0 @@
1 -XWiki.karry
Size
... ... @@ -1,1 +1,0 @@
1 -1.2 MB
Content
Copyright ©2010-2024 Dragino Technology Co., LTD. All rights reserved