From version < 45.91 >
edited by Xiaoling
on 2024/08/06 17:58
To version < 54.1 >
edited by Karry Zhuang
on 2024/12/21 09:35
>
Change comment: There is no comment for this version

Summary

Details

Page properties
Author
... ... @@ -1,1 +1,1 @@
1 -XWiki.Xiaoling
1 +XWiki.karry
Content
... ... @@ -19,8 +19,8 @@
19 19  * **EC Range & Resolution:**
20 20  ** **ECK0.01** : 0.02 ~~ 20 μS/cm
21 21  ** **ECK0.1**: 0.2 ~~ 200.0 μS/cm
22 -** **ECK1.0** : 2 ~~ 2,000 μS/cm  Resolution: 1 μS/cm
23 -** **ECK10.0** : 20 ~~ 20,000 μS/cm  Resolution: 10 μS/cm
22 +** **ECK1.0** : 0 ~~ 2,000 μS/cm  Resolution: 1 μS/cm
23 +** **ECK10.0** : 10 ~~ 20,000 μS/cm  Resolution: 10 μS/cm
24 24  
25 25  * **EC Accuracy**: ±1% FS
26 26  
... ... @@ -41,7 +41,7 @@
41 41  == 1.3 Wiring ==
42 42  
43 43  
44 -[[image:image-20240720172533-1.png||height="347" width="569"]]
44 +[[image:image-20241129142314-1.png]]
45 45  
46 46  
47 47  == 1.4 Mechinical Drawing ==
... ... @@ -481,7 +481,6 @@
481 481  
482 482  * **Max Pressure**: 0.6MPa
483 483  
484 -
485 485  == 3.2 Wiring ==
486 486  
487 487  
... ... @@ -642,8 +642,6 @@
642 642  0XC3
643 643  )))
644 644  
645 -
646 -
647 647  = 4. DR-DO1 Dissolved Oxygen Sensor =
648 648  
649 649  == 4.1 Specification ==
... ... @@ -674,9 +674,10 @@
674 674  == 4.3 Impedance requirements for current signals ==
675 675  
676 676  
677 -[[image:image-20240718195414-8.png||height="100" width="575"]]
674 +(% border="1" cellspacing="3" style="background-color:#f2f2f2; width:400px" %)
675 +|(% style="width:132px" %)**Supply Voltage**|(% style="width:67px" %)**9V**|(% style="width:67px" %)**12V**|(% style="width:67px" %)**20V**|(% style="width:67px" %)**24V**
676 +|(% style="width:132px" %)**Max Impedance**|(% style="width:65px" %)**<250Ω**|(% style="width:67px" %)**<400Ω**|(% style="width:67px" %)**<500Ω**|(% style="width:65px" %)**<900Ω**
678 678  
679 -
680 680  == 4.4 Mechinical Drawing ==
681 681  
682 682  
... ... @@ -775,7 +775,7 @@
775 775  == 5.1 Specification ==
776 776  
777 777  
778 -* **Measuring range**: 0.11000.0NTU
776 +* **Measuring range**: 0.1~~1000.0NTU
779 779  
780 780  * **Accuracy**: ±5%
781 781  
... ... @@ -783,9 +783,9 @@
783 783  
784 784  * **Stability**: ≤3mV/24 hours
785 785  
786 -* **Output signal**: A: 420 mA (current loop)B: RS485 (standard Modbus-RTU protocol, device default address: 01)
784 +* **Output signal**: RS485 (standard Modbus-RTU protocol, device default address: 01)
787 787  
788 -* **Power supply voltage**: 524V DC (when output signal is RS485)1224V DC (when output signal is 420mA)
786 +* **Power supply voltage**: 524V DC (when output signal is RS485)12-24V DC (when output signal is 420mA)
789 789  
790 790  * **Working environment**: temperature 060°C; humidity ≤ 95%RH
791 791  
... ... @@ -799,14 +799,11 @@
799 799  
800 800  == 5.3 Impedance requirements for current signals ==
801 801  
802 -供电电压 9V 12V 20V 24V
803 -最大阻抗 <250Ω <400Ω <500Ω <900Ω
804 804  
805 -(% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
806 -|=(% style="width: 128px; background-color: rgb(79, 129, 189); color: white;" %)Supply Voltage|=(% style="width: 82px" %)9V|=(% style="width: 103.6px" %)**12V**|=(% style="width: 103.6px" %)**20V**|=(% style="width: 103.6px" %)**24V**
807 -|(% style="width:128px;background-color: rgb(79, 129, 189); color: white;" %)Max Impedance|(% style="width:82px" %)<250Ω|(% style="width:106px" %)<400Ω|(% style="width:93px" %)<500Ω|(% style="width:104px" %)<900Ω
801 +(% border="1" cellspacing="3" style="background-color:#f2f2f2; width:400px" %)
802 +|(% style="width:132px" %)**Supply Voltage**|(% style="width:67px" %)**9V**|(% style="width:67px" %)**12V**|(% style="width:67px" %)**20V**|(% style="width:67px" %)**24V**
803 +|(% style="width:132px" %)**Max Impedance**|(% style="width:65px" %)**<250Ω**|(% style="width:67px" %)**<400Ω**|(% style="width:67px" %)**<500Ω**|(% style="width:65px" %)**<900Ω**
808 808  
809 -
810 810  == 5.4 Mechinical Drawing ==
811 811  
812 812  
... ... @@ -879,3 +879,9 @@
879 879  For example, the returned data is 15 03 02 (% style="color:red" %)**02 9A**(%%) 09 4C
880 880  
881 881  02 9A is the turbidity value, converted to decimal, it is 666, and then divided by 10, the actual value is 66.6, 02 9A means the current turbidity value is 66.6 NTU
877 +
878 +
879 += 6.  Water Quality Sensor Datasheet =
880 +
881 +
882 +* **[[Water Quality Sensor Transmitter Datasheet>>https://www.dropbox.com/scl/fi/9tofocmgapkbddshznumn/Datasheet_WQS-xB-WQS-xS_Water-Quality-Sensor-Transmitter.pdf?rlkey=wxua12ur9swk30rkqnh2boo9z&st=axga6epf&dl=0]]**
image-20241129142314-1.png
Author
... ... @@ -1,0 +1,1 @@
1 +XWiki.karry
Size
... ... @@ -1,0 +1,1 @@
1 +1.2 MB
Content
Copyright ©2010-2024 Dragino Technology Co., LTD. All rights reserved