Changes for page Water Quality Sensors
Last modified by Karry Zhuang on 2025/02/18 15:43
Summary
-
Page properties (1 modified, 0 added, 0 removed)
Details
- Page properties
-
- Content
-
... ... @@ -11,28 +11,19 @@ 11 11 12 12 13 13 * **Power Input**: DC7~~30 14 - 15 15 * **Power Consumption** : < 0.5W 16 - 17 17 * **Interface**: RS485. 9600 Baud Rate 18 - 19 19 * **EC Range & Resolution:** 20 20 ** **ECK0.01** : 0.02 ~~ 20 μS/cm 21 21 ** **ECK0.1**: 0.2 ~~ 200.0 μS/cm 22 22 ** **ECK1.0** : 2 ~~ 2,000 μS/cm Resolution: 1 μS/cm 23 23 ** **ECK10.0** : 20 ~~ 20,000 μS/cm Resolution: 10 μS/cm 24 - 25 25 * **EC Accuracy**: ±1% FS 26 - 27 27 * **Temperature Measure Range**: -20 ~~ 60 °C 28 - 29 29 * **Temperature Accuracy: **±0.5 °C 30 - 31 31 * **IP Rated**: IP68 32 - 33 33 * **Max Pressure**: 0.6MPa 34 34 35 - 36 36 == 1.2 Application for Different Range == 37 37 38 38 ... ... @@ -205,8 +205,7 @@ 205 205 206 206 This device uses one-point calibration, and you need to prepare a known E standard solution. When mileage K=1, 1~~2000 uses 1413μS/cm standard solution, and when mileage K=10, 10~~20000 uses 12.88mS/cm standard solution. 207 207 208 -(% style="color:blue" %)**The calibration steps are as follows:** 209 - 199 +**The calibration steps are as follows:** 210 210 (1) Place the electrode in distilled water and clean it. When mileage 1~~2000 uses 1413μS/cm standard solution, enter the following calibration command after the data is stable. 211 211 212 212 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)