... ... @@ -256,40 +256,37 @@
256 256  
257 257  == 2.2 Wiring ==
258 258  
259 -
260 260  [[image:image-20240720172548-2.png||height="348" width="571"]]
261 261  
262 262  
263 263  == (% style="color:inherit; font-family:inherit" %)2.3 (% style="color:inherit; font-family:inherit; font-size:26px" %)Mechinical Drawing(%%) ==
264 264  
265 -
266 266  [[image:image-20240714174241-2.png]]
267 267  
268 268  
269 269  == 2.4 Installation Notice ==
270 270  
271 -
272 272  Do not power on while connect the cables. Double check the wiring before power on.
273 273  
274 274  Installation Photo as reference:
275 275  
276 -(% style="color:blue" %)**Submerged installation:**
273 +**~ Submerged installation:**
277 277  
278 278  The lead wire of the equipment passes through the waterproof pipe, and the 3/4 thread on the top of the equipment is connected to the 3/4 thread of the waterproof pipe with raw tape. Ensure that the top of the equipment and the equipment wire are not flooded.
279 279  
280 280  [[image:image-20240718191348-6.png]]
281 281  
282 -(% style="color:blue" %)**Pipeline installation:**
279 +**~ Pipeline installation:**
283 283  
284 284  Connect the equipment to the pipeline through the 3/4 thread.
285 285  
286 286  [[image:image-20240718191336-5.png||height="239" width="326"]]
287 287  
288 -(% style="color:blue" %)**Sampling:**
285 +**Sampling:**
289 289  
290 290  Take representative water samples according to sampling requirements. If it is inconvenient to take samples, you can also put the electrode into the solution to be tested and read the output data. After a period of time, take out the electrode and clean it.
291 291  
292 -(% style="color:blue" %)**Measure the pH of the water sample:**
289 +**Measure the pH of the water sample:**
293 293  
294 294  First rinse the electrode with distilled water, then rinse it with the water sample, then immerse the electrode in the sample, carefully shake the test cup or stir it to accelerate the electrode balance, let it stand, and record the pH value when the reading is stable.
295 295  
... ... @@ -309,17 +309,13 @@
309 309  The equipment should be calibrated before each use. For long-term use, it is recommended to calibrate once every 3 months. The calibration frequency should be adjusted appropriately according to different application conditions (degree of dirt in the application, deposition of chemical substances, etc.). After aging, the electrodes should be replaced in time.
310 310  )))
311 311  
312 -
313 313  == 2.6 RS485 Commands ==
314 314  
315 -
316 316  RS485 signaldefault address 0x10
317 317  Standard Modbus-RTU protocol, baud rate: 9600; check bit: none; data bit: 8; stop bit: 1
318 318  
319 -
320 320  === 2.6.1 Query address ===
321 321  
322 -
323 323  send
324 324  
325 325  (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:676.25px" %)
... ... @@ -332,7 +332,6 @@
332 332  |=(% style="width: 50px;background-color:#4F81BD;color:white" %)New address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 106px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 93px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 104px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
333 333  |(% style="width:99px" %)0X01|(% style="width:112px" %)0X03|(% style="width:106px" %)0X00|(% style="width:93px" %)0X20|(% style="width:104px" %)0XF0
334 334  
335 -
336 336  === 2.6.2 Change address ===
337 337  
338 338  For example: Change the address of the sensor with address 1 to 2, master → slave