257 257  == 2.2 Wiring ==
258 258  
259 +
259 259  [[image:image-20240720172548-2.png||height="348" width="571"]]
260 260  
261 261  
262 262  == (% style="color:inherit; font-family:inherit" %)2.3 (% style="color:inherit; font-family:inherit; font-size:26px" %)Mechinical Drawing(%%) ==
263 263  
265 +
264 264  [[image:image-20240714174241-2.png]]
265 265  
266 266  
267 267  == 2.4 Installation Notice ==
268 268  
271 +
269 269  Do not power on while connect the cables. Double check the wiring before power on.
270 270  
271 271  Installation Photo as reference:
272 272  
273 -**~ Submerged installation:**
276 +(% style="color:blue" %)**Submerged installation:**
274 274  
275 275  The lead wire of the equipment passes through the waterproof pipe, and the 3/4 thread on the top of the equipment is connected to the 3/4 thread of the waterproof pipe with raw tape. Ensure that the top of the equipment and the equipment wire are not flooded.
276 276  
277 277  [[image:image-20240718191348-6.png]]
278 278  
279 -**~ Pipeline installation:**
282 +(% style="color:blue" %)**Pipeline installation:**
280 280  
281 281  Connect the equipment to the pipeline through the 3/4 thread.
282 282  
283 283  [[image:image-20240718191336-5.png||height="239" width="326"]]
284 284  
285 -**Sampling:**
288 +(% style="color:blue" %)**Sampling:**
286 286  
287 287  Take representative water samples according to sampling requirements. If it is inconvenient to take samples, you can also put the electrode into the solution to be tested and read the output data. After a period of time, take out the electrode and clean it.
288 288  
289 -**Measure the pH of the water sample:**
292 +(% style="color:blue" %)**Measure the pH of the water sample:**
290 290  
291 291  First rinse the electrode with distilled water, then rinse it with the water sample, then immerse the electrode in the sample, carefully shake the test cup or stir it to accelerate the electrode balance, let it stand, and record the pH value when the reading is stable.
292 292  
... ... @@ -306,13 +306,17 @@
306 306  The equipment should be calibrated before each use. For long-term use, it is recommended to calibrate once every 3 months. The calibration frequency should be adjusted appropriately according to different application conditions (degree of dirt in the application, deposition of chemical substances, etc.). After aging, the electrodes should be replaced in time.
307 307  )))
308 308  
312 +
309 309  == 2.6 RS485 Commands ==
310 310  
315 +
311 311  RS485 signaldefault address 0x10
312 312  Standard Modbus-RTU protocol, baud rate: 9600; check bit: none; data bit: 8; stop bit: 1
313 313  
319 +
314 314  === 2.6.1 Query address ===
315 315  
322 +
316 316  send
317 317  
318 318  (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:676.25px" %)
... ... @@ -325,6 +325,7 @@
325 325  |=(% style="width: 50px;background-color:#4F81BD;color:white" %)New address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 106px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 93px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 104px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
326 326  |(% style="width:99px" %)0X01|(% style="width:112px" %)0X03|(% style="width:106px" %)0X00|(% style="width:93px" %)0X20|(% style="width:104px" %)0XF0
327 327  
335 +
328 328  === 2.6.2 Change address ===
329 329  
330 330  For example: Change the address of the sensor with address 1 to 2, master → slave