Summary

Details

Page properties
Author
... ... @@ -1,1 +1,1 @@
1 -XWiki.karry
1 +XWiki.Xiaoling
Content
... ... @@ -3,10 +3,13 @@
3 3  {{toc/}}
4 4  
5 5  
6 +
7 +
6 6  = 1. DR-ECK Water EC Probe =
7 7  
8 8  == 1.1 Specification: ==
9 9  
12 +
10 10  * **Power Input**: DC7~~30
11 11  * **Power Consumption** : < 0.5W
12 12  * **Interface**: RS485. 9600 Baud Rate
... ... @@ -21,18 +21,23 @@
21 21  * **IP Rated**: IP68
22 22  * **Max Pressure**: 0.6MPa
23 23  
27 +
28 +
24 24  == 1.2 Application for Different Range ==
25 25  
31 +
26 26  [[image:image-20240714173018-1.png]]
27 27  
28 28  
29 29  == 1.3 Wiring ==
30 30  
37 +
31 31  [[image:image-20240720172533-1.png||height="347" width="569"]]
32 32  
33 33  
34 34  == 1.4 Mechinical Drawing ==
35 35  
43 +
36 36  [[image:image-20240714174241-2.png]]
37 37  
38 38  
... ... @@ -39,19 +39,19 @@
39 39  == 1.5 Installation ==
40 40  
41 41  
42 -**Electrode installation form**
50 +**Electrode installation form:**
43 43  
44 -A:Side wall installation
52 +A: Side wall installation
45 45  
46 -B:Top flange installation
54 +B: Top flange installation
47 47  
48 -C:Pipeline bend installation
56 +C: Pipeline bend installation
49 49  
50 -D:Pipeline bend installation
58 +D: Pipeline bend installation
51 51  
52 -E:Flow-through installation
60 +E: Flow-through installation
53 53  
54 -F:Submerged installation
62 +F: Submerged installation
55 55  
56 56  [[image:image-20240718190121-1.png||height="350" width="520"]]
57 57  
... ... @@ -63,11 +63,11 @@
63 63  
64 64  [[image:image-20240718190204-2.png||height="262" width="487"]]
65 65  
66 -Error cause: The electrode joint is too long, the extension part is too short, the sensor is easy to form a dead cavity, resulting in measurement error.
74 +**Error cause:** The electrode joint is too long, the extension part is too short, the sensor is easy to form a dead cavity, resulting in measurement error.
67 67  
68 68  [[image:image-20240718190221-3.png||height="292" width="500"]]
69 69  
70 -Error cause: Measurement error or instability may occur due to water flow not being able to fill the pipe or air accumulation at high altitudes.
78 +**Error cause: **Measurement error or instability may occur due to water flow not being able to fill the pipe or air accumulation at high altitudes.
71 71  
72 72  B. Correct installation method
73 73  
... ... @@ -82,6 +82,8 @@
82 82  * Cleaning of conductivity electrodes: Organic stains on the electrode can be cleaned with warm water containing detergent, or with alcohol. Calcium and magnesium precipitates are best cleaned with 10% citric acid. The electrode plate or pole can only be cleaned by chemical methods or by shaking in water. Wiping the electrode plate will damage the coating (platinum black) on the electrode surface.
83 83  * The equipment should be calibrated before each use. It is recommended to calibrate it every 3 months for long-term use. The calibration frequency should be adjusted appropriately according to different application conditions (degree of dirt in the application, deposition of chemical substances, etc.).
84 84  
93 +
94 +
85 85  == 1.7 RS485 Commands ==
86 86  
87 87  
... ... @@ -91,9 +91,10 @@
91 91  
92 92  === 1.7.1 Query address ===
93 93  
94 -send
95 95  
96 -(% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
105 +**send**
106 +
107 +(% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
97 97  |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Original address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Quantity high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
98 98  |(% style="width:99px" %)0XFE |(% style="width:72px" %)0X03|(% style="width:64px" %)0X00|(% style="width:68px" %)0X50|(% style="width:70px" %)0X00|(% style="width:72px" %)0X00|(% style="width:56px" %)0X51|(% style="width:56px" %)0XD4
99 99  
... ... @@ -100,14 +100,17 @@
100 100  If you forget the original address of the sensor, you can use the broadcast address 0XFE instead. When using 0XFE, the host can only connect to one slave, which can be used as a method of address query.
101 101  
102 102  
103 -response
114 +**response**
104 104  
105 -(% border="1" cellspacing="3" style="background-color:#f2f2f2; width:561.333px" %)
116 +(% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
106 106  |=(% style="width: 50px;background-color:#4F81BD;color:white" %)New address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 106px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 93px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 104px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
107 107  |(% style="width:99px" %)0X01|(% style="width:112px" %)0X03|(% style="width:106px" %)0X00|(% style="width:93px" %)0X20|(% style="width:104px" %)0XF0
108 108  
120 +
121 +
109 109  === 1.7.2 Change address ===
110 110  
124 +
111 111  For example: Change the address of the sensor with address 1 to 2, master → slave
112 112  
113 113  (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:575.333px" %)