... ... @@ -51,7 +51,7 @@
51 51  
52 52  F:Submerged installation
53 53  
54 -[[image:image-20240718190121-1.png||height="350" width="520"]]
54 +[[image:image-20240716104537-2.png||height="475" width="706"]]
55 55  
56 56  **Several common installation methods of electrodes**
57 57  
... ... @@ -59,17 +59,18 @@
59 59  
60 60  A. Several common incorrect installation methods
61 61  
62 -[[image:image-20240718190204-2.png||height="262" width="487"]]
62 +[[image:image-20240717103452-1.png||height="320" width="610"]]
63 63  
64 64  Error cause: The electrode joint is too long, the extension part is too short, the sensor is easy to form a dead cavity, resulting in measurement error.
65 +
65 65  
66 -[[image:image-20240718190221-3.png||height="292" width="500"]]
67 +[[image:image-20240716105124-4.png||height="326" width="569"]]
67 67  
68 68  Error cause: Measurement error or instability may occur due to water flow not being able to fill the pipe or air accumulation at high altitudes.
69 69  
70 70  B. Correct installation method
71 71  
72 -[[image:image-20240718190249-4.png||height="287" width="515"]]
73 +[[image:image-20240716105318-5.png||height="330" width="594"]]
73 73  
74 74  
75 75  == 1.6 Maintain ==
... ... @@ -212,6 +212,7 @@
212 212  |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 1px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 high
213 213  |(% style="width:99px" %)0X11|(% style="width:112px" %)0X06|(% style="width:135px" %)0X00|(% style="width:126px" %)0X26|(% style="width:85px" %)0X00|(% style="width:1px" %)0X02|(% style="width:1px" %)0XEB|(% style="width:1px" %)0X50
214 214  
216 +
215 215  = 2. DR-PH01 Water PH Sensor =
216 216  
217 217  == 2.7 RS485 Commands ==
... ... @@ -267,3 +267,4 @@
267 267  
268 268  02 9A is the turbidity value, converted to decimal, it is 666, and then divided by 10, the actual value is 66.6, 02 9A means the current turbidity value is 66.6 NTU
269 269  
272 +