... ... @@ -37,42 +37,31 @@
37 37  == 1.5 Installation ==
38 38  
39 39  
40 -**Electrode installation form**
40 + Do not power on while connect the cables. Double check the wiring before power on.
41 41  
42 -A:Side wall installation
42 +Installation Photo as reference:
43 43  
44 -B:Top flange installation
44 +**~ Submerged installation:**
45 45  
46 -C:Pipeline bend installation
46 +The lead wire of the equipment passes through the waterproof pipe, and the 3/4 thread on the top of the equipment is connected to the 3/4 thread of the waterproof pipe with raw tape. Ensure that the top of the equipment and the equipment wire are not flooded.
47 47  
48 -D:Pipeline bend installation
48 +[[image:image-20240715181933-4.png||height="281" width="258"]]
49 49  
50 -E:Flow-through installation
50 +**~ Pipeline installation:**
51 51  
52 -F:Submerged installation
52 +Connect the equipment to the pipeline through the 3/4 thread.
53 53  
54 -[[image:image-20240716104537-2.png||height="475" width="706"]]
54 +[[image:image-20240715182122-6.png||height="291" width="408"]]
55 55  
56 -**Several common installation methods of electrodes**
56 +**Sampling:**
57 57  
58 -When installing the sensor on site, you should strictly follow the correct installation method shown in the following picture. Incorrect installation method will cause data deviation.
58 +Take representative water samples according to sampling requirements. If it is inconvenient to take samples, you can also put the electrode into the solution to be tested and read the output data. After a period of time, take out the electrode and clean it.
59 59  
60 -A. Several common incorrect installation methods
60 +**Measure the pH of the water sample:**
61 61  
62 -[[image:image-20240717103452-1.png||height="320" width="610"]]
62 +First rinse the electrode with distilled water, then rinse it with the water sample, then immerse the electrode in the sample, carefully shake the test cup or stir it to accelerate the electrode balance, let it stand, and record the pH value when the reading is stable.
63 63  
64 -Error cause: The electrode joint is too long, the extension part is too short, the sensor is easy to form a dead cavity, resulting in measurement error.
65 -
66 66  
67 -[[image:image-20240716105124-4.png||height="326" width="569"]]
68 -
69 -Error cause: Measurement error or instability may occur due to water flow not being able to fill the pipe or air accumulation at high altitudes.
70 -
71 -B. Correct installation method
72 -
73 -[[image:image-20240716105318-5.png||height="330" width="594"]]
74 -
75 -
76 76  == 1.6 Maintain ==
77 77  
78 78  
... ... @@ -147,6 +147,7 @@
147 147  0XE5
148 148  )))
149 149  
139 +
150 150  === 1.7.4 Query data ===
151 151  
152 152  The address of the EC K10 sensor is 11
... ... @@ -214,6 +214,7 @@
214 214  |(% style="width:99px" %)0X11|(% style="width:112px" %)0X06|(% style="width:135px" %)0X00|(% style="width:126px" %)0X26|(% style="width:85px" %)0X00|(% style="width:1px" %)0X02|(% style="width:1px" %)0XEB|(% style="width:1px" %)0X50
215 215  
216 216  
207 +
217 217  = 2. DR-PH01 Water PH Sensor =
218 218  
219 219  == 2.7 RS485 Commands ==