<
From version < 143.3 >
edited by Xiaoling
on 2022/06/10 17:25
To version < 147.2 >
edited by Xiaoling
on 2022/06/10 17:34
>
Change comment: There is no comment for this version

Summary

Details

Page properties
Content
... ... @@ -58,7 +58,6 @@
58 58  * 4000mAh or 8500mAh Battery for long term use
59 59  
60 60  
61 -
62 62  == 1.3  Specification ==
63 63  
64 64  === 1.3.1  Rated environmental conditions ===
... ... @@ -100,7 +100,6 @@
100 100  * Bottom water level monitoring
101 101  
102 102  
103 -
104 104  == 1.6  Pin mapping and power on ==
105 105  
106 106  
... ... @@ -242,7 +242,6 @@
242 242  * If the sensor value lower than 0x0118 (280mm), the sensor value will be invalid. Since v1.1.4, all value lower than 280mm will be set to 0x0014(20mm) which means the value is invalid.
243 243  
244 244  
245 -
246 246  === 2.3.3  Interrupt Pin ===
247 247  
248 248  This data field shows if this packet is generated by interrupt or not. [[Click here>>||anchor="H4.2A0SetInterruptMode"]] for the hardware and software set up.
... ... @@ -813,7 +813,6 @@
813 813  * Blink once when device transmit a packet.
814 814  
815 815  
816 -
817 817  == 2.8  ​Firmware Change Log ==
818 818  
819 819  
... ... @@ -832,7 +832,7 @@
832 832  [[image:image-20220610172003-2.png]]
833 833  
834 834  
835 -== 2.10  Battery Analysis  ==
831 +== 2.10  Battery Analysis ==
836 836  
837 837  === 2.10.1  Battery Type ===
838 838  
... ... @@ -857,80 +857,27 @@
857 857  
858 858  === 2.10.2  Replace the battery ===
859 859  
860 -You can change the battery in the LDDS75.The type of battery is not limited as long as the output is between 3v to 3.6v. On the main board, there is a diode (D1) between the battery and the main circuit. If you need to use a battery with less than 3.3v, please remove the D1 and shortcut the two pads of it so there won’t be voltage drop between battery and main board.
861 -
862 -
863 -The default battery pack of LDDS75 includes a ER18505 plus super capacitor. If user can’t find this pack locally, they can find ER18505 or equivalence, which will also work in most case. The SPC can enlarge the battery life for high frequency use (update period below 5 minutes)
864 -
865 -
866 -
867 -= 3.  LiDAR ToF Measurement =
868 -
869 -== 3.1 Principle of Distance Measurement ==
870 -
871 -The LiDAR probe is based on TOF, namely, Time of Flight principle. To be specific, the product emits modulation wave of near infrared ray on a periodic basis, which will be reflected after contacting object. The product obtains the time of flight by measuring round-trip phase difference and then calculates relative range between the product and the detection object, as shown below.
872 -
873 -[[image:1654831757579-263.png]]
874 -
875 -
876 -
877 -== 3.2 Distance Measurement Characteristics ==
878 -
879 -With optimization of light path and algorithm, The LiDAR probe has minimized influence from external environment on distance measurement performance. Despite that, the range of distance measurement may still be affected by the environment illumination intensity and the reflectivity of detection object. As shown in below:
880 -
881 -[[image:1654831774373-275.png]]
882 -
883 -
884 884  (((
885 -(% style="color:blue" %)**① **(%%)Represents the detection blind zone of The LiDAR probe, 0-10cm, within which the output data is unreliable.
857 +You can change the battery in the LDDS75.The type of battery is not limited as long as the output is between 3v to 3.6v. On the main board, there is a diode (D1) between the battery and the main circuit. If you need to use a battery with less than 3.3v, please remove the D1 and shortcut the two pads of it so there won’t be voltage drop between battery and main board.
886 886  )))
887 887  
888 888  (((
889 -(% style="color:blue" %)**② **(%%)Represents the operating range of The LiDAR probe detecting black target with 10% reflectivity, 0.1-5m.
861 +
890 890  )))
891 891  
892 892  (((
893 -(% style="color:blue" %)**③ **(%%)Represents the operating range of The LiDAR probe detecting white target with 90% reflectivity, 0.1-12m.
865 +The default battery pack of LDDS75 includes a ER18505 plus super capacitor. If user cant find this pack locally, they can find ER18505 or equivalence, which will also work in most case. The SPC can enlarge the battery life for high frequency use (update period below 5 minutes)
894 894  )))
895 895  
896 896  
897 -(((
898 -Vertical Coordinates: Represents the radius of light spot for The LiDAR probe at the different distances. The diameter of light spot depends on the FOV of The LiDAR probe (the term of FOV generally refers to the smaller value between the receiving angle and the transmitting angle), which is calculated as follows:
899 -)))
900 900  
870 += 3.  Configure LLDS12 via AT Command or LoRaWAN Downlink =
901 901  
902 -[[image:1654831797521-720.png]]
903 -
904 -
905 905  (((
906 -In the formula above, d is the diameter of light spot; D is detecting range; β is the value of the receiving angle of The LiDAR probe, 3.6°. Correspondence between the diameter of light spot and detecting range is given in Table below.
907 -)))
908 -
909 -[[image:1654831810009-716.png]]
910 -
911 -
912 912  (((
913 -If the light spot reaches two objects with different distances, as shown in Figure 3, the output distance value will be a value between the actual distance values of the two objects. For a high accuracy requirement in practice, the above situation should be noticed to avoid the measurement error.
874 +Use can configure LDDS75 via AT Command or LoRaWAN Downlink.
914 914  )))
915 -
916 -
917 -
918 -== 3.3 Notice of usage: ==
919 -
920 -Possible invalid /wrong reading for LiDAR ToF tech:
921 -
922 -* Measure high reflectivity object such as: Mirror, Smooth ceramic tile, static milk surface, will have possible wrong readings.
923 -* While there is transparent object such as glass, water drop between the measured object and the LiDAR sensor, the reading might wrong.
924 -* The LiDAR probe is cover by dirty things; the reading might be wrong. In this case, need to clean the probe.
925 -* The sensor window is made by Acrylic. Don’t touch it with alcohol material. This will destroy the sensor window.
926 -
927 -= 4.  Configure LLDS12 via AT Command or LoRaWAN Downlink =
928 -
929 -(((
930 -(((
931 -Use can configure LLDS12 via AT Command or LoRaWAN Downlink.
932 932  )))
933 -)))
934 934  
935 935  * (((
936 936  (((
... ... @@ -949,7 +949,7 @@
949 949  )))
950 950  
951 951  (((
952 -There are two kinds of commands to configure LLDS12, they are:
895 +There are two kinds of commands to configure LDDS75, they are:
953 953  )))
954 954  )))
955 955  
... ... @@ -990,55 +990,49 @@
990 990  
991 991  * (((
992 992  (((
993 -(% style="color:#4f81bd" %)** Commands special design for LLDS12**
936 +(% style="color:#4f81bd" %)** Commands special design for LDDS75**
994 994  )))
995 995  )))
996 996  
997 997  (((
998 998  (((
999 -These commands only valid for LLDS12, as below:
942 +These commands only valid for LDDS75, as below:
1000 1000  )))
1001 1001  )))
1002 1002  
1003 1003  
1004 1004  
1005 -== 4.1  Set Transmit Interval Time ==
948 +== 3.1  Access AT Commands ==
1006 1006  
1007 -Feature: Change LoRaWAN End Node Transmit Interval.
950 +LDDS75 supports AT Command set in the stock firmware. You can use a USB to TTL adapter to connect to LDDS75 for using AT command, as below.
1008 1008  
1009 -(% style="color:#037691" %)**AT Command: AT+TDC**
952 +[[image:image-20220610172924-4.png||height="483" width="988"]]
1010 1010  
1011 -[[image:image-20220607171554-8.png]]
1012 1012  
955 +Or if you have below board, use below connection:
1013 1013  
1014 -(((
1015 -(% style="color:#037691" %)**Downlink Command: 0x01**
1016 -)))
1017 1017  
1018 -(((
1019 -Format: Command Code (0x01) followed by 3 bytes time value.
1020 -)))
958 +[[image:image-20220610172924-5.png]]
1021 1021  
1022 -(((
1023 -If the downlink payload=0100003C, it means set the END Node’s Transmit Interval to 0x00003C=60(S), while type code is 01.
1024 -)))
1025 1025  
1026 -* (((
1027 -Example 1: Downlink Payload: 0100001E ~/~/ Set Transmit Interval (TDC) = 30 seconds
1028 -)))
1029 -* (((
1030 -Example 2: Downlink Payload: 0100003C ~/~/ Set Transmit Interval (TDC) = 60 seconds
1031 -)))
961 +In the PC, you need to set the serial baud rate to (% style="color:green" %)**9600**(%%) to access the serial console for LDDS75. LDDS75 will output system info once power on as below:
1032 1032  
1033 -== 4.2  Set Interrupt Mode ==
1034 1034  
1035 -Feature, Set Interrupt mode for GPIO_EXIT.
964 + [[image:image-20220610172924-6.png||height="601" width="860"]]
1036 1036  
1037 -(% style="color:#037691" %)**AT Command: AT+INTMOD**
1038 1038  
1039 -[[image:image-20220610105806-2.png]]
1040 1040  
968 +== 3.2  Set Transmit Interval Time ==
1041 1041  
970 +Feature: Change LoRaWAN End Node Transmit Interval.
971 +
972 +(% style="color:#037691" %)**AT Command: AT+TDC**
973 +
974 +[[image:image-20220610173409-7.png]]
975 +
976 +
977 +
978 +
1042 1042  (((
1043 1043  (% style="color:#037691" %)**Downlink Command: 0x06**
1044 1044  )))
... ... @@ -1058,7 +1058,7 @@
1058 1058  Example 2: Downlink Payload: 06000003 ~/~/ Set the interrupt mode to rising edge trigger
1059 1059  )))
1060 1060  
1061 -== 4.3  Get Firmware Version Info ==
998 +== 3.3  Get Firmware Version Info ==
1062 1062  
1063 1063  Feature: use downlink to get firmware version.
1064 1064  
image-20220610172924-4.png
Author
... ... @@ -1,0 +1,1 @@
1 +XWiki.Xiaoling
Size
... ... @@ -1,0 +1,1 @@
1 +1.5 MB
Content
image-20220610172924-5.png
Author
... ... @@ -1,0 +1,1 @@
1 +XWiki.Xiaoling
Size
... ... @@ -1,0 +1,1 @@
1 +901.4 KB
Content
image-20220610172924-6.png
Author
... ... @@ -1,0 +1,1 @@
1 +XWiki.Xiaoling
Size
... ... @@ -1,0 +1,1 @@
1 +68.6 KB
Content
image-20220610173409-7.png
Author
... ... @@ -1,0 +1,1 @@
1 +XWiki.Xiaoling
Size
... ... @@ -1,0 +1,1 @@
1 +11.8 KB
Content
Copyright ©2010-2024 Dragino Technology Co., LTD. All rights reserved
Dragino Wiki v2.0