... ... @@ -267,17 +267,16 @@
267 267  **3. Send the calibration command to Sensor 2**:
268 268  
269 269  * Command Format: **0B 06 00 53 [(% style="color:#037691" %)K-value(%%) (2 bytes)] [CRC]**
270 -* Example for K-value 0x02EE: 0B 06 00 53 02 EE F8 5D
271 271  
272 -(% style="color:red" %)**Notes:**
271 +Example for K-value 0x02EE: 0B 06 00 53 02 EE F8 5D
273 273  
274 -~1. Always use the sensor with the most trusted reading as the reference (do not modify its K-value).
273 +Notes:
274 +Always use the sensor with the most trusted reading as the reference (do not modify its K-value).
275 +The default K-value 03E8 (1000) implies no scaling. Values below 03E8 reduce the output; values above increase it.
276 +CRC is auto-generated by the tool (no manual calculation needed).
275 275  
276 -2. The default K-value 03E8 (1000) implies no scaling. Values below 03E8 reduce the output; values above increase it.
277 277  
278 -3. CRC is auto-generated by the tool (no manual calculation needed).
279 279  
280 -
281 281  = 5. WSS-25 ~-~- Temperature, Humidity, Illuminance, Pressure =