240 -== 4.6 Calibration Values (K-value Adjustment) ==
241 -
242 -In the event of deviations in the sensor values, calibration can be carried out by modifying the sensor's internal parameter: K-value parameter(**Default: 03E8(H)=1000(D)**).
243 -
244 -(% style="color:#037691" %)**Calibration Principle**
245 -
246 -The K-value scales the sensor's output. If multiple sensors measure the same parameter but show different values, adjust the K-value of the secondary sensor(s) to match the primary sensor's reading.
247 -
248 -(% style="color:#037691" %)**Step-by-Step Example:**
249 -
250 -**Scenario:**
251 -
252 -* Sensor 1 (reference) reads 1500.
253 -* Sensor 2 reads 2000, but should match Sensor 1's value (~~1500).
254 -
255 -**Steps:**
256 -
257 -if the value of sensor 1 is 1500 and the value of sensor 2 is 2000, it is necessary to keep the values around 1500if the value of sensor 1 is 1500 and the value of sensor 2 is 2000, it is necessary to keep the values around 1500.
258 -
259 -Modify the K value of sensor 2:
260 -
261 -* K value calculation as follows:
262 -* 1500/2000=0.75,  0.75*1000=750
263 -* Convert 750 to hexadecimal to 02EE.
264 -* Send command to sensor 2
265 -* Send:0B 06 00 53 02 EE F8 5D  Send successfully and return this command
266 -
267 -
268 -
269 -