... ... @@ -239,10 +239,19 @@
239 239  
240 240  == 4.6 Calibration Values (K-value Adjustment) ==
241 241  
242 -In the event of deviations in the sensor values, calibration can be carried out by modifying the sensor's internal parameter: K-value parameter(**Default: 03E8(h)=1000(D)**).
242 +In the event of deviations in the sensor values, calibration can be carried out by modifying the sensor's internal parameter: K-value parameter(**Default: 03E8(H)=1000(D)**).
243 243  
244 -For example
244 +**Calibration Principle**
245 245  
246 +The K-value scales the sensor's output. If multiple sensors measure the same parameter but show different values, adjust the K-value of the secondary sensor(s) to match the primary sensor's reading.
247 +
248 +**Step-by-Step Example:**
249 +
250 +Scenario:
251 +
252 +* Sensor 1 (reference) reads 1500.
253 +* Sensor 2 reads 2000, but should match Sensor 1's value (~~1500).
254 +
246 246  if the value of sensor 1 is 1500 and the value of sensor 2 is 2000, it is necessary to keep the values around 1500if the value of sensor 1 is 1500 and the value of sensor 2 is 2000, it is necessary to keep the values around 1500.
247 247  
248 248  Modify the K value of sensor 2: