Changes for page

Last modified by Kilight Cao on 2025/07/22 16:13

From version 81.18
edited by Mengting Qiu
on 2025/07/01 14:12
Change comment: There is no comment for this version
To version 81.5
edited by Mengting Qiu
on 2025/07/01 13:55
Change comment: There is no comment for this version

Summary

Details

Page properties
Content
... ... @@ -241,42 +241,30 @@
241 241  
242 242  In the event of deviations in the sensor values, calibration can be carried out by modifying the sensor's internal parameter: K-value parameter(**Default: 03E8(H)=1000(D)**).
243 243  
244 -(% style="color:#037691" %)**Calibration Principle**
244 +**Calibration Principle**
245 245  
246 246  The K-value scales the sensor's output. If multiple sensors measure the same parameter but show different values, adjust the K-value of the secondary sensor(s) to match the primary sensor's reading.
247 247  
248 -(% style="color:#037691" %)**Step-by-Step Example:**
248 +**Step-by-Step Example:**
249 249  
250 -**Scenario:**
250 +Scenario:
251 251  
252 252  * Sensor 1 (reference) reads 1500.
253 253  * Sensor 2 reads 2000, but should match Sensor 1's value (~~1500).
254 254  
255 -**Steps:**
255 +if the value of sensor 1 is 1500 and the value of sensor 2 is 2000, it is necessary to keep the values around 1500if the value of sensor 1 is 1500 and the value of sensor 2 is 2000, it is necessary to keep the values around 1500.
256 256  
257 -**~1. Calculate the K-value ratio**:
257 +Modify the K value of sensor 2:
258 258  
259 -Ratio=Desired Value/Current Value=1500/2000=0.75
259 +* K value calculation as follows:
260 +* 1500/2000=0.75,  0.75*1000=750
261 +* Convert 750 to hexadecimal to 02EE.
262 +* Send command to sensor 2
263 +* Send:0B 06 00 53 02 EE F8 5D  Send successfully and return this command
260 260  
261 -Multiply by the default K-value (1000) to get the new K-value: 0.75×1000=750(Decimal)
262 262  
263 -**2. Convert the (% style="color:#037691" %)new K-value(%%) to hexadecimal**:
264 264  
265 -* 750(D) = 0x02EE(H)
266 266  
267 -**3. Send the calibration command to Sensor 2**:
268 -
269 -* Command Format: **0B 06 00 53 [(% style="color:#037691" %)K-value(%%) (2 bytes)] [CRC]**
270 -
271 -Example for K-value 0x02EE: 0B 06 00 53 02 EE F8 5D
272 -
273 -Notes:
274 -Always use the sensor with the most trusted reading as the reference (do not modify its K-value).
275 -The default K-value 03E8 (1000) implies no scaling. Values below 03E8 reduce the output; values above increase it.
276 -CRC is auto-generated by the tool (no manual calculation needed).
277 -
278 -
279 -
280 280  = 5. WSS-25 ~-~- Temperature, Humidity, Illuminance, Pressure =
281 281  
282 282