239 239  
240 240  == 4.6 Calibration Values (K-value Adjustment) ==
241 241  
242 -In the event of deviations in the sensor values, calibration can be carried out by modifying the sensor's internal parameter: K-value parameter(**Default: 03E8(H)=1000(D)**).
242 +In the event of deviations in the sensor values, calibration can be carried out by modifying the sensor's internal parameter: K-value parameter(**Default: 03E8(h)=1000(D)**).
243 243  
244 -(% style="color:#037691" %)**Calibration Principle**
244 +For example
245 245  
246 -The K-value scales the sensor's output. If multiple sensors measure the same parameter but show different values, adjust the K-value of the secondary sensor(s) to match the primary sensor's reading.
246 +if the value of sensor 1 is 1500 and the value of sensor 2 is 2000, it is necessary to keep the values around 1500if the value of sensor 1 is 1500 and the value of sensor 2 is 2000, it is necessary to keep the values around 1500.
247 247  
248 -(% style="color:#037691" %)**Step-by-Step Example:**
248 +Modify the K value of sensor 2:
249 249  
250 -**Scenario:**
250 +* K value calculation as follows:
251 +* 1500/2000=0.75,  0.75*1000=750
252 +* Convert 750 to hexadecimal to 02EE.
253 +* Send command to sensor 2
254 +* Send:0B 06 00 53 02 EE F8 5D  Send successfully and return this command
251 251  
252 -* Sensor 1 (reference) reads 1500.
253 -* Sensor 2 reads 2000, but should match Sensor 1's value (~~1500).
254 254  
255 -**Steps:**
256 256  
257 -**~1. Calculate the K-value ratio**:
258 258  
259 -Ratio=Desired Value/Current Value=1500/2000=0.75
260 -
261 -Multiply by the default K-value (1000) to get the new K-value: 0.75×1000=750(Decimal)
262 -
263 -**2. Convert the (% style="color:#037691" %)new K-value(%%) to hexadecimal**:
264 -
265 -* 750(D) = 0x02EE(H)
266 -
267 -**3. Send the calibration command to Sensor 2**:
268 -
269 -* Command Format: **0B 06 00 53 [(% style="color:#037691" %)K-value(%%) (2 bytes)] [CRC]**
270 -
271 -Example for K-value 0x02EE: 0B 06 00 53 02 EE F8 5D
272 -
273 -Notes:
274 -Always use the sensor with the most trusted reading as the reference (do not modify its K-value).
275 -The default K-value 03E8 (1000) implies no scaling. Values below 03E8 reduce the output; values above increase it.
276 -CRC is auto-generated by the tool (no manual calculation needed).
277 -