567 567  User can do calibration for the probe. It is limited to use below pH buffer solution to calibrate: 4.00, 6.86, 9.18. When calibration, user need to clean the electrode and put the probe in the pH buffer solution to wait the value stable ( a new clean electrode might need max 24 hours to be stable).
568 568  
569 -**Note:Only one ph buffer with one of the ph values needs to be calibrated during calibration. Wait for the sensor measurement value to be stable, it does not need to be the same as the ph marked by the ph buffer**
569 +(% style="color:red" %)**Note:Only one ph buffer with one of the ph values needs to be calibrated during calibration. Wait for the sensor measurement value to be stable, it does not need to be the same as the ph marked by the ph buffer**
570 570  
571 571  
572 572  (% style="color:blue" %)**Step3 :(%%) The numerically stable sensors were calibrated using commands**