... ... @@ -556,20 +556,19 @@
556 556  
557 557  == 2.7 Calibration ==
558 558  
559 +Step1 : Clean probe
559 559  
560 -(% style="color:blue" %)**Step1 :(%%) Clean probe**
561 -
562 562  user can use high grade sandpaper to polish it or put in 5% hydrochloric acid for several minutes. After the metal probe looks new, users can clean it to remove hydrochloric acid or other liquids from the surface.
563 563  
564 -(% style="color:blue" %)**Step2 :(%%) Immerse the sensor in ph buffer solution to make the sensor values stable**
563 +Step2 : Immerse the sensor in ph buffer solution to make the sensor values stable
565 565  
566 566  (((
567 567  User can do calibration for the probe. It is limited to use below pH buffer solution to calibrate: 4.00, 6.86, 9.18. When calibration, user need to clean the electrode and put the probe in the pH buffer solution to wait the value stable ( a new clean electrode might need max 24 hours to be stable).
568 568  
569 -**Note:Only one ph buffer with one of the ph values needs to be calibrated during calibration. Wait for the sensor measurement value to be stable, it does not need to be the same as the ph marked by the ph buffer**
568 +Note:Only one ph buffer with one of the ph values needs to be calibrated during calibration. Wait for the sensor measurement value to be stable, it does not need to be the same as the ph marked by the ph buffer
570 570  
571 571  
572 -(% style="color:blue" %)**Step3 :(%%) The numerically stable sensors were calibrated using commands**
571 +Step3 : The numerically stable sensors were calibrated using commands
573 573  )))
574 574  
575 575  (((
... ... @@ -611,6 +611,7 @@
611 611  
612 612  * LoRaWAN Downlink.  Instruction for different platforms: See [[IoT LoRaWAN Server>>url:http://wiki.dragino.com/xwiki/bin/view/Main/]] section.
613 613  
613 +
614 614  When the command is successfully used, the node sends a uplink.
615 615  This is the payload for uplink
616 616  
... ... @@ -639,8 +639,10 @@
639 639  
640 640  * Reply to non-confirmed packet: 14 00
641 641  
642 -(% style="color:blue" %)**Step4:(%%)After the calibration is completed, its measurement results in this ph buffer solution should be consistent with the numerical value of the command used in the calibration.**
642 +Step4:
643 643  
644 +After the calibration is completed, its measurement results in this ph buffer solution should be consistent with the numerical value of the command used in the calibration
645 +
644 644  Examples:
645 645  After using the command AT+PH=9 in 9.18ph buffer
646 646  The readout returned by the sensor in 9.18ph buffer should be pH=9