Hide last authors
Edwin Chen 1.1 1 **Table of Contents:**
2
3 {{toc/}}
4
5
Xiaoling 45.2 6
7
Edwin Chen 7.1 8 = 1. DR-ECK Water EC Probe =
9
10 == 1.1 Specification: ==
11
Xiaoling 45.2 12
Edwin Chen 7.1 13 * **Power Input**: DC7~~30
Xiaoling 45.63 14
Edwin Chen 7.1 15 * **Power Consumption** : < 0.5W
Xiaoling 45.63 16
Edwin Chen 7.1 17 * **Interface**: RS485. 9600 Baud Rate
Xiaoling 45.63 18
Edwin Chen 7.1 19 * **EC Range & Resolution:**
20 ** **ECK0.01** : 0.02 ~~ 20 μS/cm
21 ** **ECK0.1**: 0.2 ~~ 200.0 μS/cm
22 ** **ECK1.0** : 2 ~~ 2,000 μS/cm  Resolution: 1 μS/cm
23 ** **ECK10.0** : 20 ~~ 20,000 μS/cm  Resolution: 10 μS/cm
Xiaoling 45.63 24
Edwin Chen 7.1 25 * **EC Accuracy**: ±1% FS
Xiaoling 45.63 26
Edwin Chen 7.1 27 * **Temperature Measure Range**: -20 ~~ 60 °C
Xiaoling 45.63 28
Edwin Chen 7.1 29 * **Temperature Accuracy: **±0.5 °C
Xiaoling 45.63 30
Edwin Chen 7.1 31 * **IP Rated**: IP68
Xiaoling 45.63 32
Edwin Chen 7.1 33 * **Max Pressure**: 0.6MPa
34
Xiaoling 45.63 35
Edwin Chen 7.1 36 == 1.2 Application for Different Range ==
37
Xiaoling 45.2 38
Edwin Chen 7.1 39 [[image:image-20240714173018-1.png]]
40
41
42 == 1.3 Wiring ==
43
Xiaoling 45.2 44
Karry Zhuang 45.1 45 [[image:image-20240720172533-1.png||height="347" width="569"]]
Edwin Chen 7.1 46
Karry Zhuang 45.1 47
Edwin Chen 7.1 48 == 1.4 Mechinical Drawing ==
49
Xiaoling 45.2 50
Edwin Chen 7.1 51 [[image:image-20240714174241-2.png]]
52
53
54 == 1.5 Installation ==
55
56
Xiaoling 45.2 57 **Electrode installation form:**
Karry Zhuang 15.2 58
Xiaoling 45.2 59 A: Side wall installation
Karry Zhuang 15.2 60
Xiaoling 45.2 61 B: Top flange installation
Karry Zhuang 15.2 62
Xiaoling 45.2 63 C: Pipeline bend installation
Karry Zhuang 15.2 64
Xiaoling 45.2 65 D: Pipeline bend installation
Karry Zhuang 15.2 66
Xiaoling 45.2 67 E: Flow-through installation
Karry Zhuang 15.2 68
Xiaoling 45.2 69 F: Submerged installation
Karry Zhuang 15.2 70
Karry Zhuang 23.1 71 [[image:image-20240718190121-1.png||height="350" width="520"]]
Karry Zhuang 15.2 72
Karry Zhuang 18.1 73 **Several common installation methods of electrodes**
Karry Zhuang 15.2 74
Karry Zhuang 18.1 75 When installing the sensor on site, you should strictly follow the correct installation method shown in the following picture. Incorrect installation method will cause data deviation.
Karry Zhuang 15.2 76
Karry Zhuang 18.1 77 A. Several common incorrect installation methods
Karry Zhuang 15.2 78
Karry Zhuang 23.1 79 [[image:image-20240718190204-2.png||height="262" width="487"]]
Karry Zhuang 15.2 80
Xiaoling 45.2 81 **Error cause:** The electrode joint is too long, the extension part is too short, the sensor is easy to form a dead cavity, resulting in measurement error.
Karry Zhuang 15.2 82
Karry Zhuang 23.1 83 [[image:image-20240718190221-3.png||height="292" width="500"]]
Karry Zhuang 18.1 84
Xiaoling 45.2 85 **Error cause: **Measurement error or instability may occur due to water flow not being able to fill the pipe or air accumulation at high altitudes.
Karry Zhuang 18.1 86
87 B. Correct installation method
88
Karry Zhuang 23.1 89 [[image:image-20240718190249-4.png||height="287" width="515"]]
Karry Zhuang 18.1 90
91
Karry Zhuang 38.1 92 == 1.6 Maintenance ==
Edwin Chen 7.1 93
94
Karry Zhuang 26.1 95 * The equipment itself generally does not require daily maintenance. When an obvious fault occurs, please do not open it and repair it yourself, and contact us as soon as possible.
Xiaoling 45.66 96
Karry Zhuang 26.1 97 * If the electrode is not used for a long time, it can generally be stored in a dry place, but it must be placed (stored) in distilled water for several hours before use to activate the electrode. Electrodes that are frequently used can be placed (stored) in distilled water.
Xiaoling 45.66 98
Karry Zhuang 26.1 99 * Cleaning of conductivity electrodes: Organic stains on the electrode can be cleaned with warm water containing detergent, or with alcohol. Calcium and magnesium precipitates are best cleaned with 10% citric acid. The electrode plate or pole can only be cleaned by chemical methods or by shaking in water. Wiping the electrode plate will damage the coating (platinum black) on the electrode surface.
Xiaoling 45.66 100
Karry Zhuang 26.1 101 * The equipment should be calibrated before each use. It is recommended to calibrate it every 3 months for long-term use. The calibration frequency should be adjusted appropriately according to different application conditions (degree of dirt in the application, deposition of chemical substances, etc.).
Karry Zhuang 15.2 102
Edwin Chen 8.1 103 == 1.7 RS485 Commands ==
104
Karry Zhuang 15.2 105
106 RS485 signal (K1 default address 0x12; K10 default address 0x11):
107 Standard Modbus-RTU protocol, baud rate: 9600; check bit: none; data bit: 8; stop bit: 1
108
109
Karry Zhuang 16.1 110 === 1.7.1 Query address ===
Edwin Chen 8.1 111
Karry Zhuang 11.1 112
Xiaoling 45.48 113 **send:**
Xiaoling 45.2 114
115 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Xiaoling 45.40 116 |=(% style="width: 74.75px;background-color:#4F81BD;color:white" %)Original address|=(% style="width: 64.75px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)Quantity low|=(% style="width: 59.75px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 59.75px;background-color:#4F81BD;color:white" %)CRC16 high
Xiaoling 45.32 117 |(% style="width:99px" %)0XFE |(% style="width:72px" %)0X03|(% style="width:50px" %)0X00|(% style="width:42px" %)0X50|(% style="width:42px" %)0X00|(% style="width:42px" %)0X00|(% style="width:56px" %)0X51|(% style="width:56px" %)0XD4
Karry Zhuang 16.1 118
119 If you forget the original address of the sensor, you can use the broadcast address 0XFE instead. When using 0XFE, the host can only connect to one slave, which can be used as a method of address query.
120
121
Xiaoling 45.48 122 **response:**
Karry Zhuang 16.1 123
Xiaoling 45.14 124 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:512px" %)
Xiaoling 45.10 125 |=(% style="width: 100px;background-color:#4F81BD;color:white" %)New address|=(% style="width: 110px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 106px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 93px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 104px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
Karry Zhuang 27.1 126 |(% style="width:99px" %)0X01|(% style="width:112px" %)0X03|(% style="width:106px" %)0X00|(% style="width:93px" %)0X20|(% style="width:104px" %)0XF0
Karry Zhuang 16.1 127
128 === 1.7.2 Change address ===
129
Xiaoling 45.2 130
Karry Zhuang 16.1 131 For example: Change the address of the sensor with address 1 to 2, master → slave
132
Xiaoling 45.20 133 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Xiaoling 45.39 134 |=(% style="width: 74.75px; background-color: rgb(79, 129, 189); color: white;" %)Original address|=(% style="width: 64.75px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64.75px; background-color: rgb(79, 129, 189); color: white;" %)Address high|=(% style="width: 64.75px; background-color: rgb(79, 129, 189); color: white;" %)Address low|=(% style="width: 64.75px; background-color: rgb(79, 129, 189); color: white;" %)Quantity high|=(% style="width: 64.75px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 59.75px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 59.75px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
Xiaoling 45.32 135 |(% style="width:67px" %)0X01|(% style="width:76px" %)0X06|(% style="width:60px" %)0X00|(% style="width:50px" %)0X50|(% style="width:50px" %)0X00|(% style="width:50px" %)0X02|(% style="width:57px" %)0X08|(% style="width:56px" %)0X1A
Karry Zhuang 16.1 136
137 If the sensor receives correctly, the data is returned along the original path.
138
Xiaoling 45.32 139 (% style="color:red" %)**Note: If you forget the original address of the sensor, you can use the broadcast address 0XFE instead. When using 0XFE, the host can only connect to one slave, and the return address is still the original address, which can be used as a method of address query.**
Karry Zhuang 16.1 140
Xiaoling 45.32 141
Karry Zhuang 16.1 142 === 1.7.3 Modify intercept ===
143
144
Xiaoling 45.48 145 **send:**
Karry Zhuang 16.1 146
Xiaoling 45.36 147 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:512px" %)
Xiaoling 45.34 148 |=(% style="width: 64px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 64px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 64px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 64px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 64px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 64px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 64px;background-color:#4F81BD;color:white" %)CRC16 high
149 |(% style="width:64px" %)0X01|(% style="width:112px" %)0X06|(% style="width:135px" %)0X00|(% style="width:126px" %)0X23|(% style="width:85px" %)0X00|(% style="width:1px" %)0X01|(% style="width:1px" %)0XF8|(% style="width:1px" %)(((
Karry Zhuang 27.1 150 0X07
Karry Zhuang 16.1 151 )))
152
153 Change the intercept of the sensor with address 1 to 10 (default 0), which is 0X000A in the command.
154
Xiaoling 45.48 155 **response:**
Karry Zhuang 16.1 156
Xiaoling 45.36 157 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:512px" %)
Xiaoling 45.35 158 |=(% style="width: 64px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 64px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 64px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 64px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 64px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 64px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 64px;background-color:#4F81BD;color:white" %)CRC16 high
Karry Zhuang 16.1 159 |(% style="width:99px" %)0X01|(% style="width:112px" %)0X06|(% style="width:135px" %)(((
160 0X02
Karry Zhuang 27.1 161 )))|(% style="width:126px" %)0X00|(% style="width:85px" %)0X00|(% style="width:1px" %)0X0A|(% style="width:1px" %)0X38|(% style="width:1px" %)(((
162 0X8F
Karry Zhuang 16.1 163 )))
164
165 === 1.7.4 Query data ===
166
Karry Zhuang 37.1 167
168 Query the data (EC,temperature) of the sensor (address 11), host → slave
169
Xiaoling 45.20 170 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Xiaoling 45.43 171 |=(% style="width: 42px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address low|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
Karry Zhuang 37.1 172 |(% style="width:99px" %)0X11|(% style="width:72px" %)0X03|(% style="width:64px" %)0X00|(% style="width:68px" %)0X00|(% style="width:70px" %)0X00|(% style="width:72px" %)0X02|(% style="width:56px" %)0XC6|(% style="width:56px" %)0X9B
173
174 If the sensor receives correctly, the following data will be returned, slave → host
175
Xiaoling 45.20 176 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Xiaoling 45.42 177 |=(% style="width: 40px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 59.75px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 59.75px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 59.75px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data high|=(% style="width: 59.75px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data low|=(% style="width: 59.75px; background-color: rgb(79, 129, 189); color: white;" %)Register 1 Data high|=(% style="width: 59.75px; background-color: rgb(79, 129, 189); color: white;" %)Register 1 Data low|=(% style="width: 59.75px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 59.75px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
Karry Zhuang 37.1 178 |(% style="width:99px" %)0X11|(% style="width:72px" %)0X03|(% style="width:68px" %)0X04|(% style="width:70px" %)0X02|(% style="width:72px" %)0XAE|(% style="width:56px" %)0X01|(% style="width:56px" %)0X64|(% style="width:56px" %)0X8B|(% style="width:56px" %)0XD0
179
Karry Zhuang 16.3 180 The address of the EC K10 sensor is 11
Karry Zhuang 16.1 181
Karry Zhuang 10.1 182 The query data command is 11 03 00 00 00 02 C6 9B
183
Xiaoling 45.48 184 **For example**, the returned data is 11 03 04 (% style="color:red" %)**02 AE**(%%) 01 64 8B D0. 02 AE is converted to decimal 686,  K=10, EC: 6860uS/cm,temperature: 35.6℃ Convert the returned data to decimal and divide by 10.
Karry Zhuang 10.1 185
186
Karry Zhuang 37.1 187 Query the data (EC,temperature) of the sensor (address 11), host → slave
188
Xiaoling 45.20 189 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Xiaoling 45.44 190 |=(% style="width: 42px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address low|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
Karry Zhuang 37.1 191 |(% style="width:99px" %)0X12|(% style="width:72px" %)0X03|(% style="width:64px" %)0X00|(% style="width:68px" %)0X00|(% style="width:70px" %)0X00|(% style="width:72px" %)0X02|(% style="width:56px" %)0XC6|(% style="width:56px" %)0XA8
192
193 If the sensor receives correctly, the following data will be returned, slave → host
194
Xiaoling 45.20 195 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Xiaoling 45.44 196 |=(% style="width: 40px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 59.75px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 59.75px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 59.75px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data high|=(% style="width: 59.75px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data low|=(% style="width: 59.75px; background-color: rgb(79, 129, 189); color: white;" %)Register 1 Data high|=(% style="width: 59.75px; background-color: rgb(79, 129, 189); color: white;" %)Register 1 Data low|=(% style="width: 59.75px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 59.75px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
Karry Zhuang 37.1 197 |(% style="width:99px" %)0X12|(% style="width:72px" %)0X03|(% style="width:68px" %)0X04|(% style="width:70px" %)0X02|(% style="width:72px" %)0XAE|(% style="width:56px" %)0X01|(% style="width:56px" %)0X64|(% style="width:56px" %)0XB8|(% style="width:56px" %)0XD0
198
Karry Zhuang 10.1 199 The address of the EC K1 sensor is 12
200
201 The query data command is 12 03 00 00 00 02 C6 A8
202
Xiaoling 45.48 203 **For example**, the returned data is 12 03 04 (% style="color:red" %)**02 AE**(%%) 01 64 B8 D0. 02 AE is converted to decimal 686,  K=1, EC: 686uS/cm,temperature: 35.6℃ Convert the returned data to decimal and divide by 10.
Karry Zhuang 10.1 204
Karry Zhuang 11.1 205
Karry Zhuang 16.2 206 === 1.7.5 Calibration Method ===
Karry Zhuang 12.1 207
208
Karry Zhuang 15.1 209 This device uses one-point calibration, and you need to prepare a known E standard solution. When mileage K=1, 1~~2000 uses 1413μS/cm standard solution, and when mileage K=10, 10~~20000 uses 12.88mS/cm standard solution.
Karry Zhuang 12.1 210
Xiaoling 45.63 211 (% style="color:blue" %)**The calibration steps are as follows:**
212
Karry Zhuang 15.1 213 (1) Place the electrode in distilled water and clean it. When mileage 1~~2000 uses 1413μS/cm standard solution, enter the following calibration command after the data is stable.
214
Xiaoling 45.20 215 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Xiaoling 45.47 216 |=(% style="width: 42px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 53px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 53px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Data|=(% style="width: 53px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 53px;background-color:#4F81BD;color:white" %)CRC16 high
Karry Zhuang 14.1 217 |(% style="width:99px" %)0X12|(% style="width:112px" %)0X10|(% style="width:135px" %)0X00|(% style="width:126px" %)0X26|(% style="width:85px" %)0X00|(% style="width:1px" %)0X02|(% style="width:1px" %)0X04|(% style="width:1px" %)(((
218 0X00
219 0X00
220 0X37
221 0X32
222 )))|(% style="width:1px" %)0XBD|(% style="width:1px" %)0XFC
223
Karry Zhuang 15.1 224 1413*10 gives 0X00003732
Karry Zhuang 14.1 225
Xiaoling 45.48 226 **response:**
Karry Zhuang 15.1 227
Xiaoling 45.20 228 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Xiaoling 45.45 229 |=(% style="width: 42px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 68px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 68px;background-color:#4F81BD;color:white" %)CRC16 high
Karry Zhuang 14.1 230 |(% style="width:99px" %)0X12|(% style="width:112px" %)0X10|(% style="width:135px" %)0X00|(% style="width:126px" %)0X26|(% style="width:85px" %)0X00|(% style="width:1px" %)0X02|(% style="width:1px" %)0XA2|(% style="width:1px" %)0XA0
231
232 (2) Place the electrode in distilled water to clean it. Use 12.88mS/cm standard solution for the range of 10~~20000. After the data is stable, enter the following calibration command
233
Xiaoling 45.20 234 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Xiaoling 45.47 235 |=(% style="width: 42px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 53px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 53px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Data|=(% style="width: 53px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 53px;background-color:#4F81BD;color:white" %)CRC16 high
Karry Zhuang 14.1 236 |(% style="width:99px" %)0X11|(% style="width:112px" %)0X10|(% style="width:135px" %)0X00|(% style="width:126px" %)0X26|(% style="width:85px" %)0X00|(% style="width:1px" %)0X02|(% style="width:1px" %)0X04|(% style="width:1px" %)(((
237 0X00
238 0X01
239 0XF7
240 0X20
241 )))|(% style="width:1px" %)0X33|(% style="width:1px" %)0X75
242
Karry Zhuang 15.1 243 12880*10 gives 0X01F720
244
Xiaoling 45.48 245 **response:**
Karry Zhuang 14.1 246
Xiaoling 45.20 247 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Xiaoling 45.45 248 |=(% style="width: 42px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 68px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 68px;background-color:#4F81BD;color:white" %)CRC16 high
Karry Zhuang 14.1 249 |(% style="width:99px" %)0X11|(% style="width:112px" %)0X06|(% style="width:135px" %)0X00|(% style="width:126px" %)0X26|(% style="width:85px" %)0X00|(% style="width:1px" %)0X02|(% style="width:1px" %)0XEB|(% style="width:1px" %)0X50
250
Edwin Chen 8.1 251 = 2. DR-PH01 Water PH Sensor =
252
Karry Zhuang 28.2 253 == 2.1 Specification ==
Edwin Chen 9.1 254
Xiaoling 45.20 255
Karry Zhuang 26.1 256 * **Power Input**: DC7~~30
Xiaoling 45.62 257
Karry Zhuang 26.1 258 * **Power Consumption** : < 0.5W
Xiaoling 45.62 259
Karry Zhuang 26.1 260 * **Interface**: RS485. 9600 Baud Rate
Xiaoling 45.62 261
Karry Zhuang 26.1 262 * **pH measurement range**: 0~~14.00pH; resolution: 0.01pH
Xiaoling 45.62 263
264 * **pH measurement error**: ±0.15pH
265
266 * **Repeatability error**: ±0.02pH
267
268 * **Temperature measurement range**:0~~60°C; resolution: 0.1°C (set temperature for manual temperature compensation, default 25°C)
269
270 * **Temperature measurement error**: ±0.5°C
271
Karry Zhuang 26.1 272 * **Temperature Measure Range**: -20 ~~ 60 °C
Xiaoling 45.62 273
Karry Zhuang 26.1 274 * **Temperature Accuracy: **±0.5 °C
Xiaoling 45.62 275
Karry Zhuang 26.1 276 * **IP Rated**: IP68
Xiaoling 45.62 277
Karry Zhuang 26.1 278 * **Max Pressure**: 0.6MPa
279
280 == 2.2 Wiring ==
281
Xiaoling 45.49 282
Karry Zhuang 45.1 283 [[image:image-20240720172548-2.png||height="348" width="571"]]
Karry Zhuang 26.1 284
Karry Zhuang 45.1 285
Xiaoling 45.62 286 == 2.3 Mechinical Drawing ==
Karry Zhuang 26.1 287
Xiaoling 45.49 288
Karry Zhuang 26.1 289 [[image:image-20240714174241-2.png]]
290
291
292 == 2.4 Installation Notice ==
293
Xiaoling 45.49 294
Karry Zhuang 26.1 295 Do not power on while connect the cables. Double check the wiring before power on.
296
297 Installation Photo as reference:
298
Xiaoling 45.50 299 (% style="color:blue" %)**Submerged installation:**
Karry Zhuang 26.1 300
301 The lead wire of the equipment passes through the waterproof pipe, and the 3/4 thread on the top of the equipment is connected to the 3/4 thread of the waterproof pipe with raw tape. Ensure that the top of the equipment and the equipment wire are not flooded.
302
303 [[image:image-20240718191348-6.png]]
304
Xiaoling 45.50 305 (% style="color:blue" %)**Pipeline installation:**
Karry Zhuang 26.1 306
307 Connect the equipment to the pipeline through the 3/4 thread.
308
309 [[image:image-20240718191336-5.png||height="239" width="326"]]
310
Xiaoling 45.50 311 (% style="color:blue" %)**Sampling:**
Karry Zhuang 26.1 312
313 Take representative water samples according to sampling requirements. If it is inconvenient to take samples, you can also put the electrode into the solution to be tested and read the output data. After a period of time, take out the electrode and clean it.
314
Xiaoling 45.50 315 (% style="color:blue" %)**Measure the pH of the water sample:**
Karry Zhuang 26.1 316
317 First rinse the electrode with distilled water, then rinse it with the water sample, then immerse the electrode in the sample, carefully shake the test cup or stir it to accelerate the electrode balance, let it stand, and record the pH value when the reading is stable.
318
319
Karry Zhuang 39.1 320 == 2.5 Maintenance ==
Karry Zhuang 26.1 321
322
323 * The equipment itself generally does not require daily maintenance. When an obvious fault occurs, please do not open it and repair it yourself. Contact us as soon as possible!
Xiaoling 45.62 324
Karry Zhuang 26.1 325 * There is an appropriate amount of soaking solution in the protective bottle at the front end of the electrode. The electrode head is soaked in it to keep the glass bulb and the liquid junction activated. When measuring, loosen the bottle cap, pull out the electrode, and rinse it with pure water before use.
Xiaoling 45.62 326
Karry Zhuang 26.1 327 * Preparation of electrode soaking solution: Take a packet of PH4.00 buffer, dissolve it in 250 ml of pure water, and soak it in 3M potassium chloride solution. The preparation is as follows: Take 25 grams of analytical pure potassium chloride and dissolve it in 100 ml of pure water.
Xiaoling 45.62 328
Karry Zhuang 26.1 329 * The glass bulb at the front end of the electrode cannot come into contact with hard objects. Any damage and scratches will make the electrode ineffective.
Xiaoling 45.62 330
Karry Zhuang 26.1 331 * Before measurement, the bubbles in the electrode glass bulb should be shaken off, otherwise it will affect the measurement. When measuring, the electrode should be stirred in the measured solution and then placed still to accelerate the response.
Xiaoling 45.62 332
Karry Zhuang 26.1 333 * The electrode should be cleaned with deionized water before and after measurement to ensure accuracy.
Xiaoling 45.62 334
Karry Zhuang 26.1 335 * After long-term use, the pH electrode will become passivated, which is characterized by a decrease in sensitivity gradient, slow response, and inaccurate readings. At this time, the bulb at the bottom of the electrode can be soaked in 0.1M dilute hydrochloric acid for 24 hours (0.1M dilute hydrochloric acid preparation: 9 ml of hydrochloric acid is diluted to 1000 ml with distilled water), and then soaked in 3.3M potassium chloride solution for 24 hours. If the pH electrode is seriously passivated and soaking in 0.1M hydrochloric acid has no effect, the pH electrode bulb can be soaked in 4% HF (hydrofluoric acid) for 3-5 seconds, washed with pure water, and then soaked in 3.3M potassium chloride solution for 24 hours to restore its performance.
Xiaoling 45.62 336
Karry Zhuang 26.1 337 * Glass bulb contamination or liquid junction blockage can also cause electrode passivation. At this time, it should be cleaned with an appropriate solution according to the nature of the contaminant.
338
Xiaoling 45.62 339 * The equipment should be calibrated before each use. For long-term use, it is recommended to calibrate once every 3 months. The calibration frequency should be adjusted appropriately according to different application conditions (degree of dirt in the application, deposition of chemical substances, etc.). After aging, the electrodes should be replaced in time.
340
341
Karry Zhuang 26.1 342 == 2.6 RS485 Commands ==
343
Xiaoling 45.51 344
Karry Zhuang 27.1 345 RS485 signaldefault address 0x10
346 Standard Modbus-RTU protocol, baud rate: 9600; check bit: none; data bit: 8; stop bit: 1
Karry Zhuang 26.1 347
Xiaoling 45.51 348
Karry Zhuang 33.2 349 === 2.6.1 Query address ===
Karry Zhuang 27.1 350
Xiaoling 45.51 351
Xiaoling 45.52 352 **send:**
Karry Zhuang 27.1 353
Xiaoling 45.52 354 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Xiaoling 45.53 355 |=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)Original address|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 64.75px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)CRC16 high
Karry Zhuang 27.1 356 |(% style="width:99px" %)0XFE |(% style="width:112px" %)0X03|(% style="width:135px" %)0X00|(% style="width:126px" %)0X50|(% style="width:85px" %)0X00|(% style="width:1px" %)0X00|(% style="width:1px" %)0X51|(% style="width:1px" %)0XD4
357
Xiaoling 45.52 358 **response:**
Karry Zhuang 27.1 359
Xiaoling 45.52 360 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Xiaoling 45.57 361 |=(% style="width: 103.6px;background-color:#4F81BD;color:white" %)New address|=(% style="width: 103.6px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 103.6px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 103.6px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 103.6px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
Karry Zhuang 27.1 362 |(% style="width:99px" %)0X01|(% style="width:112px" %)0X03|(% style="width:106px" %)0X00|(% style="width:93px" %)0X20|(% style="width:104px" %)0XF0
363
364 === 2.6.2 Change address ===
365
Xiaoling 45.52 366
Karry Zhuang 27.1 367 For example: Change the address of the sensor with address 1 to 2, master → slave
368
Xiaoling 45.52 369 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Xiaoling 45.57 370 |=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)Original address|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 64.75px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 64.75px;background-color:#4F81BD;color:white" %)CRC16 high
Karry Zhuang 27.1 371 |(% style="width:99px" %)0X01|(% style="width:112px" %)0X06|(% style="width:135px" %)0X00|(% style="width:126px" %)0X50|(% style="width:85px" %)0X00|(% style="width:1px" %)0X02|(% style="width:1px" %)0X08|(% style="width:1px" %)0X1A
372
373 If the sensor receives correctly, the data is returned along the original path.
374
Xiaoling 45.52 375 (% style="color:red" %)**Note: If you forget the original address of the sensor, you can use the broadcast address 0XFE instead. When using 0XFE, the host can only connect to one slave, and the return address is still the original address, which can be used as a method of address query.**
Karry Zhuang 27.1 376
Xiaoling 45.52 377
Karry Zhuang 27.1 378 === 2.6.3 Modify intercept ===
379
380
Xiaoling 45.52 381 **send:**
Karry Zhuang 27.1 382
Xiaoling 45.52 383 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Xiaoling 45.56 384 |=(% style="width: 44.75px; background-color: rgb(79, 129, 189); color: white;" %)Address|=(% style="width: 64.75px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 69.75px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 69.75px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address  low|=(% style="width: 69.75px; background-color: rgb(79, 129, 189); color: white;" %)Register Length high|=(% style="width: 69.75px; background-color: rgb(79, 129, 189); color: white;" %)Register Length low|=(% style="width: 64.75px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 64.75px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
Karry Zhuang 34.3 385 |(% style="width:71px" %)0X10|(% style="width:74px" %)0X06|(% style="width:67px" %)0X00|(% style="width:68px" %)0X10|(% style="width:69px" %)0X00|(% style="width:66px" %)0X64|(% style="width:57px" %)0X8A|(% style="width:57px" %)(((
Karry Zhuang 27.1 386 0XA5
387 )))
388
Karry Zhuang 34.4 389 Change the intercept of the sensor at address 10 to 1 (default is 0). You need to pass the intercept 1*100 =100 into the command 0x006.
Karry Zhuang 27.1 390
Xiaoling 45.52 391 **response:**
Karry Zhuang 27.1 392
Xiaoling 45.52 393 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Xiaoling 45.55 394 |=(% style="width: 42px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 68px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 68px;background-color:#4F81BD;color:white" %)CRC16 high
Karry Zhuang 27.1 395 |(% style="width:99px" %)0X10|(% style="width:112px" %)0X06|(% style="width:135px" %)(((
396 0X00
397 )))|(% style="width:126px" %)0X10|(% style="width:85px" %)0X00|(% style="width:1px" %)0X64|(% style="width:1px" %)0X8A|(% style="width:1px" %)(((
398 0XA5
399 )))
400
401 === 2.6.4 Query data ===
402
403
Karry Zhuang 34.3 404 Query the data (PH) of the sensor (address 10), host → slave
Edwin Chen 9.1 405
Xiaoling 45.52 406 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Xiaoling 45.59 407 |=(% style="width: 42px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address low|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
Karry Zhuang 34.3 408 |(% style="width:99px" %)0X10|(% style="width:72px" %)0X03|(% style="width:64px" %)0X00|(% style="width:68px" %)0X00|(% style="width:70px" %)0X00|(% style="width:72px" %)0X01|(% style="width:56px" %)0X87|(% style="width:56px" %)0X4B
409
410 If the sensor receives correctly, the following data will be returned, slave → host
411
Xiaoling 45.60 412 (% border="1" cellspacing="3" style="background-color:#f2f2f2;width:518px" %)
Xiaoling 45.61 413 |=(% style="width: 44px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 79px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 79px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 79px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data high|=(% style="width: 79px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data low|=(% style="width: 79px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 79px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
Karry Zhuang 35.1 414 |(% style="width:99px" %)0X10|(% style="width:72px" %)0X03|(% style="width:68px" %)0X02|(% style="width:70px" %)0X02|(% style="width:72px" %)0XAE|(% style="width:56px" %)0XC4|(% style="width:56px" %)0X9B
Karry Zhuang 34.3 415
Karry Zhuang 11.1 416 The query data command is 10 03 00 00 00 01 87 4B. After the query, 7 bytes will be returned.
Karry Zhuang 10.1 417
Karry Zhuang 11.1 418 For example, the returned data is 10 03 02 (% style="color:red" %)**02 AE**(%%) C4 9B.
419
420 02 AE is the pH value, which is converted into decimal to get 686, and then two decimal places are added to get the actual value. 02 AE means the current pH value is 6.86.
421
422
Karry Zhuang 27.1 423 === 2.6.5 Calibration Method ===
424
425
426 This device uses three-point calibration, and three known pH standard solutions need to be prepared.
Xiaoling 45.62 427
428 (% style="color:blue" %)**The calibration steps are as follows:**
429
Karry Zhuang 27.1 430 (1) Place the electrode in distilled water to clean it, and then place it in 9.18 standard buffer solution. After the data stabilizes, enter the following calibration command, and the 9.18 calibration is completed.
431
Xiaoling 45.52 432 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Karry Zhuang 27.1 433 |=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 66px; background-color: rgb(79, 129, 189); color: white;" %)Address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Address low|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Quantity high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
434 |(% style="width:64px" %)0X10|(% style="width:72px" %)0X06|(% style="width:66px" %)(((
435 0X00
436 )))|(% style="width:68px" %)0X20|(% style="width:72px" %)0XFF|(% style="width:70px" %)0XFF|(% style="width:55px" %)0X8A|(% style="width:55px" %)(((
437 0XF1
438 )))
439
440 (2) Wash the electrode in distilled water and place it in 6.86 standard buffer. After the data stabilizes, enter the following calibration command. The 6.86 calibration is completed.
441
Xiaoling 45.52 442 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Karry Zhuang 27.1 443 |=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 66px; background-color: rgb(79, 129, 189); color: white;" %)Address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Address low|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Quantity high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
444 |(% style="width:64px" %)0X10|(% style="width:72px" %)0X06|(% style="width:66px" %)(((
445 0X00
446 )))|(% style="width:68px" %)0X21|(% style="width:72px" %)0XFF|(% style="width:70px" %)0XFF|(% style="width:55px" %)0XDB|(% style="width:55px" %)(((
447 0X31
448 )))
449
450 (3) Wash the electrode in distilled water and place it in 4.01 standard buffer. After the data stabilizes, enter the following calibration command, and the 4.00 calibration is completed.
451
Xiaoling 45.52 452 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Karry Zhuang 27.1 453 |=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 66px; background-color: rgb(79, 129, 189); color: white;" %)Address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Address low|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Quantity high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
454 |(% style="width:64px" %)0X10|(% style="width:72px" %)0X06|(% style="width:66px" %)(((
455 0X00
456 )))|(% style="width:68px" %)0X22|(% style="width:72px" %)0XFF|(% style="width:70px" %)0XFF|(% style="width:55px" %)0X2B|(% style="width:55px" %)(((
457 0X31
458 )))
459
460 After the above three steps are completed, the calibration is successful. The advantage of three-point calibration compared to two-point calibration is that the electrode is calibrated separately in the acid and alkali parts, thereby achieving accurate calibration of the full range and making the measurement data more accurate.
461
462
Edwin Chen 8.1 463 = 3. DR-ORP1 Water ORP Sensor =
464
Karry Zhuang 28.2 465 == 3.1 Specification ==
Karry Zhuang 27.2 466
467 * **Power Input**: DC7~~30
Karry Zhuang 32.1 468 * **Measuring range**:** **-1999~~1999mV
469 **Resolution**: 1mV
Karry Zhuang 27.2 470 * **Interface**: RS485. 9600 Baud Rate
471 * **Measurement error**: ±3mV
472 * **Stability**: ≤2mv/24 hours
473 * **Equipment working conditions**: Ambient temperature: 0-60℃ Relative humidity: <85%RH
474 * **IP Rated**: IP68
475 * **Max Pressure**: 0.6MPa
476
477 == 3.2 Wiring ==
478
Karry Zhuang 45.1 479 [[image:image-20240720172620-3.png||height="378" width="620"]]
Karry Zhuang 27.2 480
Karry Zhuang 45.1 481
Karry Zhuang 27.2 482 == 3.3 Mechinical Drawing ==
483
484 [[image:image-20240714174241-2.png]]
485
486 == 3.4 Installation Notice ==
487
488 Do not power on while connect the cables. Double check the wiring before power on.
489
490 Installation Photo as reference:
491
492 **~ Submerged installation:**
493
494 The lead wire of the equipment passes through the waterproof pipe, and the 3/4 thread on the top of the equipment is connected to the 3/4 thread of the waterproof pipe with raw tape. Ensure that the top of the equipment and the equipment wire are not flooded.
495
496 [[image:image-20240718191348-6.png]]
497
498 **~ Pipeline installation:**
499
500 Connect the equipment to the pipeline through the 3/4 thread.
501
502 [[image:image-20240718191336-5.png||height="239" width="326"]]
503
504
Karry Zhuang 39.1 505 == 3.5 Maintenance ==
Edwin Chen 8.1 506
Edwin Chen 9.1 507
Karry Zhuang 29.1 508 (1) The equipment itself generally does not require daily maintenance. When an obvious fault occurs, please do not open it and repair it yourself, and contact us as soon as possible.
Karry Zhuang 32.3 509
Karry Zhuang 29.1 510 (2) In general, ORP electrodes do not need to be calibrated and can be used directly. When there is doubt about the quality and test results of the ORP electrode, the electrode potential can be checked with an ORP standard solution to determine whether the ORP electrode meets the measurement requirements, and the electrode can be recalibrated or replaced with a new ORP electrode. The frequency of calibration or inspection of the measuring electrode depends on different application conditions (the degree of dirt in the application, the deposition of chemical substances, etc.).
Karry Zhuang 32.3 511
Karry Zhuang 29.1 512 (3) There is an appropriate soaking solution in the protective bottle at the front end of the electrode, and the electrode head is soaked in it to ensure the activation of the platinum sheet and the liquid junction. When measuring, loosen the bottle cap, pull out the electrode, and rinse it with pure water before use.
Karry Zhuang 32.3 513
Karry Zhuang 29.1 514 (4) Preparation of electrode soaking solution: Take 25 grams of analytical pure potassium chloride and dissolve it in 100 ml of pure water to prepare a 3.3M potassium chloride solution.
Karry Zhuang 32.3 515
Karry Zhuang 29.1 516 (5) Before measuring, the bubbles in the electrode glass bulb should be shaken off, otherwise it will affect the measurement. When measuring, the electrode should be stirred in the measured solution and then placed still to accelerate the response.
Karry Zhuang 32.3 517
Karry Zhuang 29.1 518 (6) The electrode should be cleaned with deionized water before and after the measurement to ensure the measurement accuracy.
Karry Zhuang 32.3 519
Karry Zhuang 29.1 520 (7) After long-term use, the ORP electrode will be passivated, which is manifested as a decrease in sensitivity gradient, slow response, and inaccurate readings. At this time, the platinum sheet at the bottom of the electrode can be soaked in 0.1M dilute hydrochloric acid for 24 hours (0.1M dilute hydrochloric acid preparation: 9 ml of hydrochloric acid is diluted to 1000 ml with distilled water), and then soaked in 3.3M potassium chloride solution for 24 hours to restore its performance.
Karry Zhuang 32.3 521
Karry Zhuang 29.1 522 (8) Electrode contamination or liquid junction blockage can also cause electrode passivation. At this time, it should be cleaned with an appropriate solution according to the nature of the contaminant. If the platinum of the electrode is severely contaminated and an oxide film is formed, toothpaste can be applied to the platinum surface and then gently scrubbed to restore the platinum's luster.
Karry Zhuang 32.3 523
Karry Zhuang 29.1 524 (9) The equipment should be calibrated before each use. It is recommended to calibrate once every 3 months for long-term use. The calibration frequency should be adjusted appropriately according to different application conditions (degree of dirt in the application, deposition of chemical substances, etc.). After aging, the electrodes should be replaced in time.
525
526 == 3.6 RS485 Commands ==
527
528
529 RS485 signaldefault address 0x13
530 Standard Modbus-RTU protocol, baud rate: 9600; check bit: none; data bit: 8; stop bit: 1
531
Karry Zhuang 33.2 532 === 3.6.1 Query address ===
Karry Zhuang 29.1 533
534 send
535
536 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:676.25px" %)
537 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Original address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 1px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 high
538 |(% style="width:99px" %)0XFE |(% style="width:112px" %)0X03|(% style="width:135px" %)0X00|(% style="width:126px" %)0X50|(% style="width:85px" %)0X00|(% style="width:1px" %)0X00|(% style="width:1px" %)0X51|(% style="width:1px" %)0XD4
539
540 response
541
542 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:561.333px" %)
543 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)New address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 106px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 93px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 104px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
544 |(% style="width:99px" %)0X01|(% style="width:112px" %)0X03|(% style="width:106px" %)0X00|(% style="width:93px" %)0X20|(% style="width:104px" %)0XF0
545
546 === 3.6.2 Change address ===
547
548 For example: Change the address of the sensor with address 1 to 2, master → slave
549
550 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:676.25px" %)
551 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Original address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 1px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 high
552 |(% style="width:99px" %)0X01|(% style="width:112px" %)0X06|(% style="width:135px" %)0X00|(% style="width:126px" %)0X50|(% style="width:85px" %)0X00|(% style="width:1px" %)0X02|(% style="width:1px" %)0X08|(% style="width:1px" %)0X1A
553
554 If the sensor receives correctly, the data is returned along the original path.
555 Note: If you forget the original address of the sensor, you can use the broadcast address 0XFE instead. When using 0XFE, the host can only connect to one slave, and the return address is still the original address, which can be used as a method of address query.
556
557
558 === 3.6.3 Modify intercept ===
559
560 send
561
562 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:676.25px" %)
Karry Zhuang 35.1 563 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 67px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address  low|=(% style="width: 69px; background-color: rgb(79, 129, 189); color: white;" %)Register Length high|=(% style="width: 66px; background-color: rgb(79, 129, 189); color: white;" %)Register Length low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 high
Karry Zhuang 29.1 564 |(% style="width:99px" %)0X13|(% style="width:112px" %)0X06|(% style="width:135px" %)0X00|(% style="width:126px" %)0X10|(% style="width:85px" %)0X00|(% style="width:1px" %)0X64|(% style="width:1px" %)0X8A|(% style="width:1px" %)(((
565 0X96
566 )))
567
568 Change the intercept of the sensor with address 1 to 10 (default 0), which is 0X000A in the command.
569
570 response
571
572 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:676.25px" %)
573 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 1px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 high
574 |(% style="width:99px" %)0X13|(% style="width:112px" %)0X06|(% style="width:135px" %)(((
575 0X00
576 )))|(% style="width:126px" %)0X10|(% style="width:85px" %)0X00|(% style="width:1px" %)0X64|(% style="width:1px" %)0X8A|(% style="width:1px" %)(((
577 0X96
578 )))
579
580 === 3.6.4 Query data ===
581
Edwin Chen 9.1 582
Karry Zhuang 37.1 583 Query the data (ORP) of the sensor (address 13), host → slave
Karry Zhuang 35.1 584
585 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
586 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
587 |(% style="width:99px" %)0X13|(% style="width:72px" %)0X03|(% style="width:64px" %)0X00|(% style="width:68px" %)0X00|(% style="width:70px" %)0X00|(% style="width:72px" %)0X01|(% style="width:56px" %)0X87|(% style="width:56px" %)0X78
588
589 If the sensor receives correctly, the following data will be returned, slave → host
590
591 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
592 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
593 |(% style="width:99px" %)0X13|(% style="width:72px" %)0X03|(% style="width:68px" %)0X02|(% style="width:70px" %)0X02|(% style="width:72px" %)0XAE|(% style="width:56px" %)0X80|(% style="width:56px" %)0X9B
594
Karry Zhuang 11.1 595 The query data command is 13 03 00 00 00 01 87 78
Karry Zhuang 10.1 596
Karry Zhuang 11.1 597 For example, the returned data is 13 03 02 (% style="color:red" %)**02 AE**(%%) 80 9B.
Karry Zhuang 10.1 598
Karry Zhuang 11.1 599 02 AE is the ORP value, converted to decimal, the actual value is 686, 02 AE means the current ORP value is 686mV
600
601
Karry Zhuang 29.1 602 === 3.6.5 Calibration Method ===
603
604 This device uses two-point calibration, and two known ORP standard solutions need to be prepared. The calibration steps are as follows:
605 (1) Place the electrode in distilled water to clean it, and then place it in 86mV standard buffer solution. After the data stabilizes,
606 enter the following calibration command, and the 86mV point calibration is completed;
607
608 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:575.333px" %)
609 |=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 66px; background-color: rgb(79, 129, 189); color: white;" %)Address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Address low|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Quantity high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
610 |(% style="width:64px" %)0X13|(% style="width:72px" %)0X06|(% style="width:66px" %)(((
611 0X00
612 )))|(% style="width:68px" %)0X24|(% style="width:72px" %)0XFF|(% style="width:70px" %)0XFF|(% style="width:55px" %)0XCB|(% style="width:55px" %)(((
613 0X03
614 )))
615
616 Wash the electrode in distilled water and place it in 256mV standard buffer. After the data is stable, enter the following calibration command to complete the 256mV point calibration.
617
618 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:575.333px" %)
619 |=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 66px; background-color: rgb(79, 129, 189); color: white;" %)Address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Address low|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Quantity high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
620 |(% style="width:64px" %)0X13|(% style="width:72px" %)0X06|(% style="width:66px" %)(((
621 0X00
622 )))|(% style="width:68px" %)0X25|(% style="width:72px" %)0XFF|(% style="width:70px" %)0XFF|(% style="width:55px" %)0X9A|(% style="width:55px" %)(((
623 0XC3
624 )))
625
Edwin Chen 8.1 626 = 4. DR-DO1 Dissolved Oxygen Sensor =
627
628
Karry Zhuang 11.1 629
Karry Zhuang 32.1 630 == 4.1 Specification ==
631
Karry Zhuang 33.2 632
633 * **Measuring range**: 0-20mg/L, 0-50℃
634 * **Accuracy**: 3%, ±0.5℃
635 * **Resolution**: 0.01 mg/L, 0.01℃
636 * **Maximum operating pressure**: 6 bar
637 * **Output signal**: A: 4-20mA (current loop)B: RS485 (standard Modbus-RTU protocol, device default address: 01)
638 * **Power supply voltage**: 5-24V DC
639 * **Working environment**: temperature 0-60℃; humidity <95%RH
Karry Zhuang 32.1 640 * **Power consumption**: ≤0.5W
641
Karry Zhuang 33.2 642 == 4.2 wiring ==
Karry Zhuang 32.1 643
Karry Zhuang 45.1 644 [[image:image-20240720172632-4.png||height="390" width="640"]]
Karry Zhuang 33.2 645
646
647 == (% id="cke_bm_224234S" style="display:none" %) (%%)4.3 Impedance requirements for current signals ==
648
Karry Zhuang 32.1 649 [[image:image-20240718195414-8.png||height="100" width="575"]]
650
651
652 == 4.4 Mechinical Drawing ==
653
654
Karry Zhuang 33.2 655 [[image:image-20240719155308-1.png||height="226" width="527"]]
Karry Zhuang 32.1 656
Karry Zhuang 33.2 657
Karry Zhuang 39.1 658 == 4.5 Instructions for use and maintenance ==
Karry Zhuang 32.1 659
660 * It can be directly put into water without adding a protective tube, ensuring the long-term stability, reliability and accuracy of the sensor.
661 * If the water conditions are complex and you want accurate data, you need to wipe the sensor probe frequently.
662
663 == 4.6 RS485 Commands ==
664
Karry Zhuang 34.1 665 RS485 signaldefault address 0x14
666 Standard Modbus-RTU protocol, baud rate: 9600; check bit: none; data bit: 8; stop bit: 1
667
Karry Zhuang 32.3 668 === 4.6.1 Query address ===
Karry Zhuang 32.1 669
Karry Zhuang 32.3 670 send
671
672 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
Karry Zhuang 34.1 673 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Original address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
674 |(% style="width:99px" %)0XFF|(% style="width:72px" %)0X03|(% style="width:64px" %)0X00|(% style="width:68px" %)0X0A|(% style="width:70px" %)0X00|(% style="width:72px" %)0X02|(% style="width:56px" %)0XF1|(% style="width:56px" %)0XD7
Karry Zhuang 32.3 675
Karry Zhuang 34.1 676 If you forget the original address of the sensor, you can use the broadcast address 0XFF instead. When using 0XFE, the host can only connect to one slave, which can be used as a method of address query.
Karry Zhuang 32.3 677
678
679 response
680
Karry Zhuang 34.1 681 Register 0 data high and register 0 data low indicate the actual address of the sensor: 1
682 Register 1 data high and register 1 data low indicate the sensor version
Karry Zhuang 32.3 683
Karry Zhuang 34.1 684 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
685 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register 1 Data high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Register 1 Data low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
686 |(% style="width:99px" %)0XFF|(% style="width:72px" %)0X03|(% style="width:64px" %)0X04|(% style="width:68px" %)0X00|(% style="width:70px" %)0X01|(% style="width:72px" %)0X00|(% style="width:56px" %)0X00|(% style="width:56px" %)0XB4|(% style="width:56px" %)0X3C
687
Karry Zhuang 33.2 688 === 4.6.2 Change address ===
Karry Zhuang 32.3 689
Karry Zhuang 34.1 690 For example: Change the address of the sensor with address 1 to 2(address range: 1-119), master → slave
Karry Zhuang 33.2 691
692 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:907.333px" %)
693 |=(% style="width: 67px; background-color: rgb(79, 129, 189); color: white;" %)Original address|=(% style="width: 71px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 65px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 65px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 53px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 53px; background-color: rgb(79, 129, 189); color: white;" %)Start address high|=(% style="width: 53px; background-color: rgb(79, 129, 189); color: white;" %)Start address low|=(% style="width: 53px; background-color: rgb(79, 129, 189); color: white;" %)Sensor version|=(% style="width: 53px; background-color: rgb(79, 129, 189); color: white;" %)Sensor version|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low
694 |(% style="width:67px" %)0X01|(% style="width:71px" %)0X10|(% style="width:65px" %)0X00|(% style="width:65px" %)0X0A|(% style="width:70px" %)0X00|(% style="width:72px" %)0X02|(% style="width:53px" %)0X04|(% style="width:53px" %)0X00|(% style="width:72px" %)0X02|(% style="width:53px" %)0X00|(% style="width:53px" %)0X00|(% style="width:56px" %)0XD2|(% style="width:53px" %)0X10
695
Karry Zhuang 34.1 696 response
Karry Zhuang 32.3 697
Karry Zhuang 34.1 698 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
699 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
700 |(% style="width:99px" %)0X01|(% style="width:72px" %)0X10|(% style="width:64px" %)0X00|(% style="width:68px" %)0X0A|(% style="width:70px" %)0X00|(% style="width:72px" %)0X02|(% style="width:56px" %)0X61|(% style="width:56px" %)0XCA
Karry Zhuang 32.3 701
Karry Zhuang 34.1 702 === 4.6.3 Query data ===
Edwin Chen 9.1 703
704
Karry Zhuang 34.2 705 Query the data (dissolved oxygen) of the sensor (address 14), host → slave
Karry Zhuang 34.1 706
707 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
708 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
709 |(% style="width:99px" %)0X14|(% style="width:72px" %)0X03|(% style="width:64px" %)0X00|(% style="width:68px" %)0X14|(% style="width:70px" %)0X00|(% style="width:72px" %)0X01|(% style="width:56px" %)0XC6|(% style="width:56px" %)0XCB
710
Karry Zhuang 34.2 711 If the sensor receives correctly, the following data will be returned, slave → host
Karry Zhuang 34.1 712
713 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
714 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
Karry Zhuang 34.2 715 |(% style="width:99px" %)0X14|(% style="width:72px" %)0X03|(% style="width:68px" %)0X02|(% style="width:70px" %)0X03|(% style="width:72px" %)0X78|(% style="width:56px" %)0XB5|(% style="width:56px" %)0X55
Karry Zhuang 34.1 716
Karry Zhuang 11.1 717 After the query, 7 bytes will be returned. For example, the returned data is 14 03 02 (% style="color:red" %)**03 78**(%%) B5 55. 03 78 is the value of dissolved oxygen.
Karry Zhuang 10.1 718
719 Converted to decimal, it is 888. Add two decimal places to get the actual value. 03 78 means the current dissolved oxygen is 8.88mg/L
720
Karry Zhuang 11.1 721
Karry Zhuang 34.2 722 Query the data (temperature) of the sensor (address 14), host → slave
723
724 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
725 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
726 |(% style="width:99px" %)0X14|(% style="width:72px" %)0X03|(% style="width:64px" %)0X00|(% style="width:68px" %)0X11|(% style="width:70px" %)0X00|(% style="width:72px" %)0X01|(% style="width:56px" %)0XD6|(% style="width:56px" %)0XCA
727
728 If the sensor receives correctly, the following data will be returned, slave → host
729
730 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
731 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
732 |(% style="width:99px" %)0X14|(% style="width:72px" %)0X03|(% style="width:68px" %)0X02|(% style="width:70px" %)0X09|(% style="width:72px" %)0XA4|(% style="width:56px" %)0XB2|(% style="width:56px" %)0X6C
733
734 After the query, 7 bytes will be returned. For example, the returned data is 14 03 02 (% style="color:red" %)**09 A4**(%%) B2 6C. 03 78 is the value of dissolved oxygen temperature.
735
736 Converted to decimal, it is 2468. Add two decimal places to get the actual value. 09 A4 means the current dissolved oxygen temperature is 24.68℃
737
738
Edwin Chen 8.1 739 = 5. DR-TS1 Water Turbidity Sensor =
740
Edwin Chen 9.1 741
Karry Zhuang 10.1 742
Karry Zhuang 32.3 743 == (% id="cke_bm_81470S" style="display:none" %) (%%)5.1 Specification ==
744
745 * **Measuring range**: 0.1~1000.0NTU
746 * **Accuracy**: ±5%
747 * **Resolution**: 0.1NTU
748 * **Stability**: ≤3mV/24 hours
749 * **Output signal**: A: 4~20 mA (current loop)B: RS485 (standard Modbus-RTU protocol, device default address: 01)
750 * **Power supply voltage**: 5~24V DC (when output signal is RS485)12~24V DC (when output signal is 4~20mA)
751 * **Working environment**: temperature 0~60℃; humidity ≤95%RH
752 * **Power consumption**: ≤0.5W
753
754 == 5.2 wiring ==
755
Karry Zhuang 45.1 756 [[image:image-20240720172640-5.png||height="387" width="635"]]
Karry Zhuang 32.3 757
Karry Zhuang 45.1 758
Karry Zhuang 32.3 759 == 5.3 Impedance requirements for current signals ==
760
761 [[image:image-20240718195414-8.png||height="100" width="575"]]
762
763
764 == 5.4 Mechinical Drawing ==
765
766 [[image:image-20240718195058-7.png||height="305" width="593"]]
767
768
Karry Zhuang 39.1 769 == 5.5 Instructions for use and maintenance ==
Karry Zhuang 32.3 770
771 * It can be directly put into water without adding a protective tube, ensuring the long-term stability, reliability and accuracy of the sensor.
772 * If the water conditions are complex and you want accurate data, you need to wipe the sensor probe frequently.
773
774 == 5.6 RS485 Commands ==
775
776
Karry Zhuang 36.1 777 RS485 signaldefault address 0x15
778 Standard Modbus-RTU protocol, baud rate: 9600; check bit: none; data bit: 8; stop bit: 1
779
780 === 5.6.1 Query address ===
781
Karry Zhuang 32.3 782 send
783
784 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
785 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Original address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Quantity high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
786 |(% style="width:99px" %)0XFE |(% style="width:72px" %)0X03|(% style="width:64px" %)0X00|(% style="width:68px" %)0X50|(% style="width:70px" %)0X00|(% style="width:72px" %)0X00|(% style="width:56px" %)0X51|(% style="width:56px" %)0XD4
787
788 If you forget the original address of the sensor, you can use the broadcast address 0XFE instead. When using 0XFE, the host can only connect to one slave, which can be used as a method of address query.
789
790
791 response
792
793 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:561.333px" %)
794 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)New address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 106px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 93px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 104px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
795 |(% style="width:99px" %)0X01|(% style="width:112px" %)0X03|(% style="width:106px" %)0X00|(% style="width:93px" %)0X20|(% style="width:104px" %)0XF0
796
Karry Zhuang 36.1 797 === 5.6.2 Change address ===
Karry Zhuang 32.3 798
Karry Zhuang 36.1 799 For example: Change the address of the sensor with address 1 to 2, master → slave
Karry Zhuang 10.1 800
Karry Zhuang 36.1 801 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:676.25px" %)
802 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Original address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 1px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 high
803 |(% style="width:99px" %)0X01|(% style="width:112px" %)0X06|(% style="width:135px" %)0X00|(% style="width:126px" %)0X50|(% style="width:85px" %)0X00|(% style="width:1px" %)0X02|(% style="width:1px" %)0X08|(% style="width:1px" %)0X1A
804
805 If the sensor receives correctly, the data is returned along the original path.
806 Note: If you forget the original address of the sensor, you can use the broadcast address 0XFE instead. When using 0XFE, the host can only connect to one slave, and the return address is still the original address, which can be used as a method of address query.
807
808 === 5.6.3 Query data ===
809
810
811 Query the data (turbidity) of the sensor (address 15), host → slave
812
813 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
814 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
815 |(% style="width:99px" %)0X15|(% style="width:72px" %)0X03|(% style="width:64px" %)0X00|(% style="width:68px" %)0X00|(% style="width:70px" %)0X00|(% style="width:72px" %)0X01|(% style="width:56px" %)0X87|(% style="width:56px" %)0X1E
816
817 If the sensor receives correctly, the following data will be returned, slave → host
818
819 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
820 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
821 |(% style="width:99px" %)0X15|(% style="width:72px" %)0X03|(% style="width:68px" %)0X02|(% style="width:70px" %)0X02|(% style="width:72px" %)0X9A|(% style="width:56px" %)0X09|(% style="width:56px" %)0X4C
822
Karry Zhuang 11.1 823 The query data command is 15 03 00 00 00 01 87 1E
Karry Zhuang 10.1 824
Karry Zhuang 11.1 825 For example, the returned data is 15 03 02 (% style="color:red" %)**02 9A**(%%) 09 4C
826
827 02 9A is the turbidity value, converted to decimal, it is 666, and then divided by 10, the actual value is 66.6, 02 9A means the current turbidity value is 66.6 NTU
Copyright ©2010-2024 Dragino Technology Co., LTD. All rights reserved