Hide last authors
Edwin Chen 1.1 1 **Table of Contents:**
2
3 {{toc/}}
4
5
Xiaoling 45.2 6
7
Edwin Chen 7.1 8 = 1. DR-ECK Water EC Probe =
9
10 == 1.1 Specification: ==
11
Xiaoling 45.2 12
Edwin Chen 7.1 13 * **Power Input**: DC7~~30
14 * **Power Consumption** : < 0.5W
15 * **Interface**: RS485. 9600 Baud Rate
16 * **EC Range & Resolution:**
17 ** **ECK0.01** : 0.02 ~~ 20 μS/cm
18 ** **ECK0.1**: 0.2 ~~ 200.0 μS/cm
19 ** **ECK1.0** : 2 ~~ 2,000 μS/cm  Resolution: 1 μS/cm
20 ** **ECK10.0** : 20 ~~ 20,000 μS/cm  Resolution: 10 μS/cm
21 * **EC Accuracy**: ±1% FS
22 * **Temperature Measure Range**: -20 ~~ 60 °C
23 * **Temperature Accuracy: **±0.5 °C
24 * **IP Rated**: IP68
25 * **Max Pressure**: 0.6MPa
26
Xiaoling 45.2 27
Edwin Chen 7.1 28 == 1.2 Application for Different Range ==
29
Xiaoling 45.2 30
Edwin Chen 7.1 31 [[image:image-20240714173018-1.png]]
32
33
34 == 1.3 Wiring ==
35
Xiaoling 45.2 36
Karry Zhuang 45.1 37 [[image:image-20240720172533-1.png||height="347" width="569"]]
Edwin Chen 7.1 38
Karry Zhuang 45.1 39
Edwin Chen 7.1 40 == 1.4 Mechinical Drawing ==
41
Xiaoling 45.2 42
Edwin Chen 7.1 43 [[image:image-20240714174241-2.png]]
44
45
46 == 1.5 Installation ==
47
48
Xiaoling 45.2 49 **Electrode installation form:**
Karry Zhuang 15.2 50
Xiaoling 45.2 51 A: Side wall installation
Karry Zhuang 15.2 52
Xiaoling 45.2 53 B: Top flange installation
Karry Zhuang 15.2 54
Xiaoling 45.2 55 C: Pipeline bend installation
Karry Zhuang 15.2 56
Xiaoling 45.2 57 D: Pipeline bend installation
Karry Zhuang 15.2 58
Xiaoling 45.2 59 E: Flow-through installation
Karry Zhuang 15.2 60
Xiaoling 45.2 61 F: Submerged installation
Karry Zhuang 15.2 62
Karry Zhuang 23.1 63 [[image:image-20240718190121-1.png||height="350" width="520"]]
Karry Zhuang 15.2 64
Karry Zhuang 18.1 65 **Several common installation methods of electrodes**
Karry Zhuang 15.2 66
Karry Zhuang 18.1 67 When installing the sensor on site, you should strictly follow the correct installation method shown in the following picture. Incorrect installation method will cause data deviation.
Karry Zhuang 15.2 68
Karry Zhuang 18.1 69 A. Several common incorrect installation methods
Karry Zhuang 15.2 70
Karry Zhuang 23.1 71 [[image:image-20240718190204-2.png||height="262" width="487"]]
Karry Zhuang 15.2 72
Xiaoling 45.2 73 **Error cause:** The electrode joint is too long, the extension part is too short, the sensor is easy to form a dead cavity, resulting in measurement error.
Karry Zhuang 15.2 74
Karry Zhuang 23.1 75 [[image:image-20240718190221-3.png||height="292" width="500"]]
Karry Zhuang 18.1 76
Xiaoling 45.2 77 **Error cause: **Measurement error or instability may occur due to water flow not being able to fill the pipe or air accumulation at high altitudes.
Karry Zhuang 18.1 78
79 B. Correct installation method
80
Karry Zhuang 23.1 81 [[image:image-20240718190249-4.png||height="287" width="515"]]
Karry Zhuang 18.1 82
83
Karry Zhuang 38.1 84 == 1.6 Maintenance ==
Edwin Chen 7.1 85
86
Karry Zhuang 26.1 87 * The equipment itself generally does not require daily maintenance. When an obvious fault occurs, please do not open it and repair it yourself, and contact us as soon as possible.
88 * If the electrode is not used for a long time, it can generally be stored in a dry place, but it must be placed (stored) in distilled water for several hours before use to activate the electrode. Electrodes that are frequently used can be placed (stored) in distilled water.
89 * Cleaning of conductivity electrodes: Organic stains on the electrode can be cleaned with warm water containing detergent, or with alcohol. Calcium and magnesium precipitates are best cleaned with 10% citric acid. The electrode plate or pole can only be cleaned by chemical methods or by shaking in water. Wiping the electrode plate will damage the coating (platinum black) on the electrode surface.
90 * The equipment should be calibrated before each use. It is recommended to calibrate it every 3 months for long-term use. The calibration frequency should be adjusted appropriately according to different application conditions (degree of dirt in the application, deposition of chemical substances, etc.).
Karry Zhuang 15.2 91
Xiaoling 45.2 92
Edwin Chen 8.1 93 == 1.7 RS485 Commands ==
94
Karry Zhuang 15.2 95
96 RS485 signal (K1 default address 0x12; K10 default address 0x11):
97 Standard Modbus-RTU protocol, baud rate: 9600; check bit: none; data bit: 8; stop bit: 1
98
99
Karry Zhuang 16.1 100 === 1.7.1 Query address ===
Edwin Chen 8.1 101
Karry Zhuang 11.1 102
Xiaoling 45.2 103 **send**
104
105 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Xiaoling 45.12 106 |=(% style="width: 74px;background-color:#4F81BD;color:white" %)Original address|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 60px; background-color: rgb(79, 129, 189); color: white;" %)Address high|=(% style="width: 66px; background-color: rgb(79, 129, 189); color: white;" %)Address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Quantity high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
Xiaoling 45.29 107 |(% style="width:99px" %)0XFE |(% style="width:72px" %)0X03|(% style="width:64px" %)0X00|(% style="width:50px" %)0X50|(% style="width:50px" %)0X00|(% style="width:50px" %)0X00|(% style="width:56px" %)0X51|(% style="width:56px" %)0XD4
Karry Zhuang 16.1 108
109 If you forget the original address of the sensor, you can use the broadcast address 0XFE instead. When using 0XFE, the host can only connect to one slave, which can be used as a method of address query.
110
111
Xiaoling 45.2 112 **response**
Karry Zhuang 16.1 113
Xiaoling 45.14 114 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:512px" %)
Xiaoling 45.10 115 |=(% style="width: 100px;background-color:#4F81BD;color:white" %)New address|=(% style="width: 110px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 106px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 93px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 104px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
Karry Zhuang 27.1 116 |(% style="width:99px" %)0X01|(% style="width:112px" %)0X03|(% style="width:106px" %)0X00|(% style="width:93px" %)0X20|(% style="width:104px" %)0XF0
Karry Zhuang 16.1 117
Xiaoling 45.2 118
Karry Zhuang 16.1 119 === 1.7.2 Change address ===
120
Xiaoling 45.2 121
Karry Zhuang 16.1 122 For example: Change the address of the sensor with address 1 to 2, master → slave
123
Xiaoling 45.20 124 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Xiaoling 45.29 125 |=(% style="width: 75px; background-color: rgb(79, 129, 189); color: white;" %)Original address|=(% style="width: 65px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 65px; background-color: rgb(79, 129, 189); color: white;" %)Address high|=(% style="width: 66px; background-color: rgb(79, 129, 189); color: white;" %)Address low|=(% style="width: 71px; background-color: rgb(79, 129, 189); color: white;" %)Quantity high|=(% style="width: 71px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 54px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
126 |(% style="width:67px" %)0X01|(% style="width:76px" %)0X06|(% style="width:62px" %)0X00|(% style="width:54px" %)0X50|(% style="width:64px" %)0X00|(% style="width:64px" %)0X02|(% style="width:57px" %)0X08|(% style="width:56px" %)0X1A
Karry Zhuang 16.1 127
128 If the sensor receives correctly, the data is returned along the original path.
129 Note: If you forget the original address of the sensor, you can use the broadcast address 0XFE instead. When using 0XFE, the host can only connect to one slave, and the return address is still the original address, which can be used as a method of address query.
130
131
132 === 1.7.3 Modify intercept ===
133
134
135 send
136
Xiaoling 45.20 137 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Karry Zhuang 16.1 138 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 1px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 high
Karry Zhuang 27.1 139 |(% style="width:99px" %)0X01|(% style="width:112px" %)0X06|(% style="width:135px" %)0X00|(% style="width:126px" %)0X23|(% style="width:85px" %)0X00|(% style="width:1px" %)0X01|(% style="width:1px" %)0XF8|(% style="width:1px" %)(((
140 0X07
Karry Zhuang 16.1 141 )))
142
143 Change the intercept of the sensor with address 1 to 10 (default 0), which is 0X000A in the command.
144
145 response
146
Xiaoling 45.20 147 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Karry Zhuang 16.1 148 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 1px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 high
149 |(% style="width:99px" %)0X01|(% style="width:112px" %)0X06|(% style="width:135px" %)(((
150 0X02
Karry Zhuang 27.1 151 )))|(% style="width:126px" %)0X00|(% style="width:85px" %)0X00|(% style="width:1px" %)0X0A|(% style="width:1px" %)0X38|(% style="width:1px" %)(((
152 0X8F
Karry Zhuang 16.1 153 )))
154
155 === 1.7.4 Query data ===
156
Karry Zhuang 37.1 157
158 Query the data (EC,temperature) of the sensor (address 11), host → slave
159
Xiaoling 45.20 160 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Karry Zhuang 37.1 161 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
162 |(% style="width:99px" %)0X11|(% style="width:72px" %)0X03|(% style="width:64px" %)0X00|(% style="width:68px" %)0X00|(% style="width:70px" %)0X00|(% style="width:72px" %)0X02|(% style="width:56px" %)0XC6|(% style="width:56px" %)0X9B
163
164 If the sensor receives correctly, the following data will be returned, slave → host
165
Xiaoling 45.20 166 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Karry Zhuang 37.1 167 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data low|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register 1 Data high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register 1 Data low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
168 |(% style="width:99px" %)0X11|(% style="width:72px" %)0X03|(% style="width:68px" %)0X04|(% style="width:70px" %)0X02|(% style="width:72px" %)0XAE|(% style="width:56px" %)0X01|(% style="width:56px" %)0X64|(% style="width:56px" %)0X8B|(% style="width:56px" %)0XD0
169
Karry Zhuang 16.3 170 The address of the EC K10 sensor is 11
Karry Zhuang 16.1 171
Karry Zhuang 10.1 172 The query data command is 11 03 00 00 00 02 C6 9B
173
Karry Zhuang 37.1 174 For example, the returned data is 11 03 04 (% style="color:red" %)**02 AE**(%%) 01 64 8B D0. 02 AE is converted to decimal 686,  K=10, EC: 6860uS/cm,temperature: 35.6℃ Convert the returned data to decimal and divide by 10.
Karry Zhuang 10.1 175
176
Karry Zhuang 37.1 177 Query the data (EC,temperature) of the sensor (address 11), host → slave
178
Xiaoling 45.20 179 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Karry Zhuang 37.1 180 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
181 |(% style="width:99px" %)0X12|(% style="width:72px" %)0X03|(% style="width:64px" %)0X00|(% style="width:68px" %)0X00|(% style="width:70px" %)0X00|(% style="width:72px" %)0X02|(% style="width:56px" %)0XC6|(% style="width:56px" %)0XA8
182
183 If the sensor receives correctly, the following data will be returned, slave → host
184
Xiaoling 45.20 185 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Karry Zhuang 37.1 186 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data low|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register 1 Data high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register 1 Data low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
187 |(% style="width:99px" %)0X12|(% style="width:72px" %)0X03|(% style="width:68px" %)0X04|(% style="width:70px" %)0X02|(% style="width:72px" %)0XAE|(% style="width:56px" %)0X01|(% style="width:56px" %)0X64|(% style="width:56px" %)0XB8|(% style="width:56px" %)0XD0
188
Karry Zhuang 10.1 189 The address of the EC K1 sensor is 12
190
191 The query data command is 12 03 00 00 00 02 C6 A8
192
Karry Zhuang 37.1 193 For example, the returned data is 12 03 04 (% style="color:red" %)**02 AE**(%%) 01 64 B8 D0. 02 AE is converted to decimal 686,  K=1, EC: 686uS/cm,temperature: 35.6℃ Convert the returned data to decimal and divide by 10.
Karry Zhuang 10.1 194
Karry Zhuang 11.1 195
Karry Zhuang 16.2 196 === 1.7.5 Calibration Method ===
Karry Zhuang 12.1 197
198
Karry Zhuang 15.1 199 This device uses one-point calibration, and you need to prepare a known E standard solution. When mileage K=1, 1~~2000 uses 1413μS/cm standard solution, and when mileage K=10, 10~~20000 uses 12.88mS/cm standard solution.
Karry Zhuang 12.1 200
Karry Zhuang 15.1 201 The calibration steps are as follows:
202 (1) Place the electrode in distilled water and clean it. When mileage 1~~2000 uses 1413μS/cm standard solution, enter the following calibration command after the data is stable.
203
Xiaoling 45.20 204 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Karry Zhuang 14.1 205 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 1px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 139.083px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Data|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 high
206 |(% style="width:99px" %)0X12|(% style="width:112px" %)0X10|(% style="width:135px" %)0X00|(% style="width:126px" %)0X26|(% style="width:85px" %)0X00|(% style="width:1px" %)0X02|(% style="width:1px" %)0X04|(% style="width:1px" %)(((
207 0X00
Karry Zhuang 12.1 208
Karry Zhuang 14.1 209 0X00
210
211 0X37
212
213 0X32
214 )))|(% style="width:1px" %)0XBD|(% style="width:1px" %)0XFC
215
Karry Zhuang 15.1 216 1413*10 gives 0X00003732
Karry Zhuang 14.1 217
Karry Zhuang 16.1 218 response
Karry Zhuang 15.1 219
Xiaoling 45.20 220 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Karry Zhuang 14.1 221 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 1px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 high
222 |(% style="width:99px" %)0X12|(% style="width:112px" %)0X10|(% style="width:135px" %)0X00|(% style="width:126px" %)0X26|(% style="width:85px" %)0X00|(% style="width:1px" %)0X02|(% style="width:1px" %)0XA2|(% style="width:1px" %)0XA0
223
224 (2) Place the electrode in distilled water to clean it. Use 12.88mS/cm standard solution for the range of 10~~20000. After the data is stable, enter the following calibration command
225
Xiaoling 45.20 226 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Karry Zhuang 13.1 227 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 1px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 139.083px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Data|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 high
Karry Zhuang 14.1 228 |(% style="width:99px" %)0X11|(% style="width:112px" %)0X10|(% style="width:135px" %)0X00|(% style="width:126px" %)0X26|(% style="width:85px" %)0X00|(% style="width:1px" %)0X02|(% style="width:1px" %)0X04|(% style="width:1px" %)(((
229 0X00
Karry Zhuang 12.1 230
Karry Zhuang 14.1 231 0X01
232
233 0XF7
234
235 0X20
236 )))|(% style="width:1px" %)0X33|(% style="width:1px" %)0X75
237
Karry Zhuang 15.1 238 12880*10 gives 0X01F720
239
Karry Zhuang 16.1 240 response
Karry Zhuang 14.1 241
Xiaoling 45.20 242 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
Karry Zhuang 14.1 243 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 1px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 high
244 |(% style="width:99px" %)0X11|(% style="width:112px" %)0X06|(% style="width:135px" %)0X00|(% style="width:126px" %)0X26|(% style="width:85px" %)0X00|(% style="width:1px" %)0X02|(% style="width:1px" %)0XEB|(% style="width:1px" %)0X50
245
Xiaoling 45.20 246
247
Edwin Chen 8.1 248 = 2. DR-PH01 Water PH Sensor =
249
Karry Zhuang 28.2 250 == 2.1 Specification ==
Edwin Chen 9.1 251
Xiaoling 45.20 252
Karry Zhuang 26.1 253 * **Power Input**: DC7~~30
254 * **Power Consumption** : < 0.5W
255 * **Interface**: RS485. 9600 Baud Rate
256 * **pH measurement range**: 0~~14.00pH; resolution: 0.01pH
257 * **pH measurement error**:±0.15pH
258 * **Repeatability error**:±0.02pH
259 * **Temperature measurement range**:0~~60℃; resolution: 0.1℃ (set temperature for manual temperature compensation, default 25℃)
260 * **Temperature measurement error**: ±0.5℃
261 * **Temperature Measure Range**: -20 ~~ 60 °C
262 * **Temperature Accuracy: **±0.5 °C
263 * **IP Rated**: IP68
264 * **Max Pressure**: 0.6MPa
265
266 == 2.2 Wiring ==
267
Karry Zhuang 45.1 268 [[image:image-20240720172548-2.png||height="348" width="571"]]
Karry Zhuang 26.1 269
Karry Zhuang 45.1 270
Karry Zhuang 26.1 271 == (% style="color:inherit; font-family:inherit" %)2.3 (% style="color:inherit; font-family:inherit; font-size:26px" %)Mechinical Drawing(%%) ==
272
273 [[image:image-20240714174241-2.png]]
274
275
276 == 2.4 Installation Notice ==
277
278 Do not power on while connect the cables. Double check the wiring before power on.
279
280 Installation Photo as reference:
281
282 **~ Submerged installation:**
283
284 The lead wire of the equipment passes through the waterproof pipe, and the 3/4 thread on the top of the equipment is connected to the 3/4 thread of the waterproof pipe with raw tape. Ensure that the top of the equipment and the equipment wire are not flooded.
285
286 [[image:image-20240718191348-6.png]]
287
288 **~ Pipeline installation:**
289
290 Connect the equipment to the pipeline through the 3/4 thread.
291
292 [[image:image-20240718191336-5.png||height="239" width="326"]]
293
294 **Sampling:**
295
296 Take representative water samples according to sampling requirements. If it is inconvenient to take samples, you can also put the electrode into the solution to be tested and read the output data. After a period of time, take out the electrode and clean it.
297
298 **Measure the pH of the water sample:**
299
300 First rinse the electrode with distilled water, then rinse it with the water sample, then immerse the electrode in the sample, carefully shake the test cup or stir it to accelerate the electrode balance, let it stand, and record the pH value when the reading is stable.
301
302
Karry Zhuang 39.1 303 == 2.5 Maintenance ==
Karry Zhuang 26.1 304
305
306 * The equipment itself generally does not require daily maintenance. When an obvious fault occurs, please do not open it and repair it yourself. Contact us as soon as possible!
307 * There is an appropriate amount of soaking solution in the protective bottle at the front end of the electrode. The electrode head is soaked in it to keep the glass bulb and the liquid junction activated. When measuring, loosen the bottle cap, pull out the electrode, and rinse it with pure water before use.
308 * Preparation of electrode soaking solution: Take a packet of PH4.00 buffer, dissolve it in 250 ml of pure water, and soak it in 3M potassium chloride solution. The preparation is as follows: Take 25 grams of analytical pure potassium chloride and dissolve it in 100 ml of pure water.
309 * The glass bulb at the front end of the electrode cannot come into contact with hard objects. Any damage and scratches will make the electrode ineffective.
310 * Before measurement, the bubbles in the electrode glass bulb should be shaken off, otherwise it will affect the measurement. When measuring, the electrode should be stirred in the measured solution and then placed still to accelerate the response.
311 * The electrode should be cleaned with deionized water before and after measurement to ensure accuracy.
312 * After long-term use, the pH electrode will become passivated, which is characterized by a decrease in sensitivity gradient, slow response, and inaccurate readings. At this time, the bulb at the bottom of the electrode can be soaked in 0.1M dilute hydrochloric acid for 24 hours (0.1M dilute hydrochloric acid preparation: 9 ml of hydrochloric acid is diluted to 1000 ml with distilled water), and then soaked in 3.3M potassium chloride solution for 24 hours. If the pH electrode is seriously passivated and soaking in 0.1M hydrochloric acid has no effect, the pH electrode bulb can be soaked in 4% HF (hydrofluoric acid) for 3-5 seconds, washed with pure water, and then soaked in 3.3M potassium chloride solution for 24 hours to restore its performance.
313 * Glass bulb contamination or liquid junction blockage can also cause electrode passivation. At this time, it should be cleaned with an appropriate solution according to the nature of the contaminant.
314 * (((
315 The equipment should be calibrated before each use. For long-term use, it is recommended to calibrate once every 3 months. The calibration frequency should be adjusted appropriately according to different application conditions (degree of dirt in the application, deposition of chemical substances, etc.). After aging, the electrodes should be replaced in time.
316 )))
317
318 == 2.6 RS485 Commands ==
319
Karry Zhuang 27.1 320 RS485 signaldefault address 0x10
321 Standard Modbus-RTU protocol, baud rate: 9600; check bit: none; data bit: 8; stop bit: 1
Karry Zhuang 26.1 322
Karry Zhuang 33.2 323 === 2.6.1 Query address ===
Karry Zhuang 27.1 324
325 send
326
327 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:676.25px" %)
328 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Original address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 1px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 high
329 |(% style="width:99px" %)0XFE |(% style="width:112px" %)0X03|(% style="width:135px" %)0X00|(% style="width:126px" %)0X50|(% style="width:85px" %)0X00|(% style="width:1px" %)0X00|(% style="width:1px" %)0X51|(% style="width:1px" %)0XD4
330
331 response
332
333 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:561.333px" %)
334 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)New address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 106px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 93px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 104px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
335 |(% style="width:99px" %)0X01|(% style="width:112px" %)0X03|(% style="width:106px" %)0X00|(% style="width:93px" %)0X20|(% style="width:104px" %)0XF0
336
337 === 2.6.2 Change address ===
338
339 For example: Change the address of the sensor with address 1 to 2, master → slave
340
341 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:676.25px" %)
342 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Original address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 1px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 high
343 |(% style="width:99px" %)0X01|(% style="width:112px" %)0X06|(% style="width:135px" %)0X00|(% style="width:126px" %)0X50|(% style="width:85px" %)0X00|(% style="width:1px" %)0X02|(% style="width:1px" %)0X08|(% style="width:1px" %)0X1A
344
345 If the sensor receives correctly, the data is returned along the original path.
346 Note: If you forget the original address of the sensor, you can use the broadcast address 0XFE instead. When using 0XFE, the host can only connect to one slave, and the return address is still the original address, which can be used as a method of address query.
347
348
349 === 2.6.3 Modify intercept ===
350
351
352 send
353
Karry Zhuang 34.3 354 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:570.333px" %)
355 |=(% style="width: 71px; background-color: rgb(79, 129, 189); color: white;" %)Address|=(% style="width: 74px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 67px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address  low|=(% style="width: 69px; background-color: rgb(79, 129, 189); color: white;" %)Register Length high|=(% style="width: 66px; background-color: rgb(79, 129, 189); color: white;" %)Register Length low|=(% style="width: 57px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 57px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
356 |(% style="width:71px" %)0X10|(% style="width:74px" %)0X06|(% style="width:67px" %)0X00|(% style="width:68px" %)0X10|(% style="width:69px" %)0X00|(% style="width:66px" %)0X64|(% style="width:57px" %)0X8A|(% style="width:57px" %)(((
Karry Zhuang 27.1 357 0XA5
358 )))
359
Karry Zhuang 34.4 360 Change the intercept of the sensor at address 10 to 1 (default is 0). You need to pass the intercept 1*100 =100 into the command 0x006.
Karry Zhuang 27.1 361
362 response
363
364 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:676.25px" %)
365 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 1px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 high
366 |(% style="width:99px" %)0X10|(% style="width:112px" %)0X06|(% style="width:135px" %)(((
367 0X00
368 )))|(% style="width:126px" %)0X10|(% style="width:85px" %)0X00|(% style="width:1px" %)0X64|(% style="width:1px" %)0X8A|(% style="width:1px" %)(((
369 0XA5
370 )))
371
372 === 2.6.4 Query data ===
373
374
Karry Zhuang 34.3 375 Query the data (PH) of the sensor (address 10), host → slave
Edwin Chen 9.1 376
Karry Zhuang 34.3 377 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
378 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
379 |(% style="width:99px" %)0X10|(% style="width:72px" %)0X03|(% style="width:64px" %)0X00|(% style="width:68px" %)0X00|(% style="width:70px" %)0X00|(% style="width:72px" %)0X01|(% style="width:56px" %)0X87|(% style="width:56px" %)0X4B
380
381 If the sensor receives correctly, the following data will be returned, slave → host
382
383 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
384 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
Karry Zhuang 35.1 385 |(% style="width:99px" %)0X10|(% style="width:72px" %)0X03|(% style="width:68px" %)0X02|(% style="width:70px" %)0X02|(% style="width:72px" %)0XAE|(% style="width:56px" %)0XC4|(% style="width:56px" %)0X9B
Karry Zhuang 34.3 386
Karry Zhuang 11.1 387 The query data command is 10 03 00 00 00 01 87 4B. After the query, 7 bytes will be returned.
Karry Zhuang 10.1 388
Karry Zhuang 11.1 389 For example, the returned data is 10 03 02 (% style="color:red" %)**02 AE**(%%) C4 9B.
390
391 02 AE is the pH value, which is converted into decimal to get 686, and then two decimal places are added to get the actual value. 02 AE means the current pH value is 6.86.
392
393
Karry Zhuang 27.1 394 === 2.6.5 Calibration Method ===
395
396
397 This device uses three-point calibration, and three known pH standard solutions need to be prepared.
398 The calibration steps are as follows:
399 (1) Place the electrode in distilled water to clean it, and then place it in 9.18 standard buffer solution. After the data stabilizes, enter the following calibration command, and the 9.18 calibration is completed.
400
401 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:575.333px" %)
402 |=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 66px; background-color: rgb(79, 129, 189); color: white;" %)Address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Address low|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Quantity high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
403 |(% style="width:64px" %)0X10|(% style="width:72px" %)0X06|(% style="width:66px" %)(((
404 0X00
405 )))|(% style="width:68px" %)0X20|(% style="width:72px" %)0XFF|(% style="width:70px" %)0XFF|(% style="width:55px" %)0X8A|(% style="width:55px" %)(((
406 0XF1
407 )))
408
409 (2) Wash the electrode in distilled water and place it in 6.86 standard buffer. After the data stabilizes, enter the following calibration command. The 6.86 calibration is completed.
410
411 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:575.333px" %)
412 |=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 66px; background-color: rgb(79, 129, 189); color: white;" %)Address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Address low|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Quantity high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
413 |(% style="width:64px" %)0X10|(% style="width:72px" %)0X06|(% style="width:66px" %)(((
414 0X00
415 )))|(% style="width:68px" %)0X21|(% style="width:72px" %)0XFF|(% style="width:70px" %)0XFF|(% style="width:55px" %)0XDB|(% style="width:55px" %)(((
416 0X31
417 )))
418
419 (3) Wash the electrode in distilled water and place it in 4.01 standard buffer. After the data stabilizes, enter the following calibration command, and the 4.00 calibration is completed.
420
421 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:575.333px" %)
422 |=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 66px; background-color: rgb(79, 129, 189); color: white;" %)Address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Address low|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Quantity high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
423 |(% style="width:64px" %)0X10|(% style="width:72px" %)0X06|(% style="width:66px" %)(((
424 0X00
425 )))|(% style="width:68px" %)0X22|(% style="width:72px" %)0XFF|(% style="width:70px" %)0XFF|(% style="width:55px" %)0X2B|(% style="width:55px" %)(((
426 0X31
427 )))
428
429 After the above three steps are completed, the calibration is successful. The advantage of three-point calibration compared to two-point calibration is that the electrode is calibrated separately in the acid and alkali parts, thereby achieving accurate calibration of the full range and making the measurement data more accurate.
430
431
Edwin Chen 8.1 432 = 3. DR-ORP1 Water ORP Sensor =
433
Karry Zhuang 27.2 434
Karry Zhuang 28.2 435 == 3.1 Specification ==
Karry Zhuang 27.2 436
437 * **Power Input**: DC7~~30
Karry Zhuang 32.1 438 * **Measuring range**:** **-1999~~1999mV
439 **Resolution**: 1mV
Karry Zhuang 27.2 440 * **Interface**: RS485. 9600 Baud Rate
441 * **Measurement error**: ±3mV
442 * **Stability**: ≤2mv/24 hours
443 * **Equipment working conditions**: Ambient temperature: 0-60℃ Relative humidity: <85%RH
444 * **IP Rated**: IP68
445 * **Max Pressure**: 0.6MPa
446
447 == 3.2 Wiring ==
448
Karry Zhuang 45.1 449 [[image:image-20240720172620-3.png||height="378" width="620"]]
Karry Zhuang 27.2 450
Karry Zhuang 45.1 451
Karry Zhuang 27.2 452 == 3.3 Mechinical Drawing ==
453
454 [[image:image-20240714174241-2.png]]
455
456 == 3.4 Installation Notice ==
457
458 Do not power on while connect the cables. Double check the wiring before power on.
459
460 Installation Photo as reference:
461
462 **~ Submerged installation:**
463
464 The lead wire of the equipment passes through the waterproof pipe, and the 3/4 thread on the top of the equipment is connected to the 3/4 thread of the waterproof pipe with raw tape. Ensure that the top of the equipment and the equipment wire are not flooded.
465
466 [[image:image-20240718191348-6.png]]
467
468 **~ Pipeline installation:**
469
470 Connect the equipment to the pipeline through the 3/4 thread.
471
472 [[image:image-20240718191336-5.png||height="239" width="326"]]
473
474
Karry Zhuang 39.1 475 == 3.5 Maintenance ==
Edwin Chen 8.1 476
Edwin Chen 9.1 477
Karry Zhuang 29.1 478 (1) The equipment itself generally does not require daily maintenance. When an obvious fault occurs, please do not open it and repair it yourself, and contact us as soon as possible.
Karry Zhuang 32.3 479
Karry Zhuang 29.1 480 (2) In general, ORP electrodes do not need to be calibrated and can be used directly. When there is doubt about the quality and test results of the ORP electrode, the electrode potential can be checked with an ORP standard solution to determine whether the ORP electrode meets the measurement requirements, and the electrode can be recalibrated or replaced with a new ORP electrode. The frequency of calibration or inspection of the measuring electrode depends on different application conditions (the degree of dirt in the application, the deposition of chemical substances, etc.).
Karry Zhuang 32.3 481
Karry Zhuang 29.1 482 (3) There is an appropriate soaking solution in the protective bottle at the front end of the electrode, and the electrode head is soaked in it to ensure the activation of the platinum sheet and the liquid junction. When measuring, loosen the bottle cap, pull out the electrode, and rinse it with pure water before use.
Karry Zhuang 32.3 483
Karry Zhuang 29.1 484 (4) Preparation of electrode soaking solution: Take 25 grams of analytical pure potassium chloride and dissolve it in 100 ml of pure water to prepare a 3.3M potassium chloride solution.
Karry Zhuang 32.3 485
Karry Zhuang 29.1 486 (5) Before measuring, the bubbles in the electrode glass bulb should be shaken off, otherwise it will affect the measurement. When measuring, the electrode should be stirred in the measured solution and then placed still to accelerate the response.
Karry Zhuang 32.3 487
Karry Zhuang 29.1 488 (6) The electrode should be cleaned with deionized water before and after the measurement to ensure the measurement accuracy.
Karry Zhuang 32.3 489
Karry Zhuang 29.1 490 (7) After long-term use, the ORP electrode will be passivated, which is manifested as a decrease in sensitivity gradient, slow response, and inaccurate readings. At this time, the platinum sheet at the bottom of the electrode can be soaked in 0.1M dilute hydrochloric acid for 24 hours (0.1M dilute hydrochloric acid preparation: 9 ml of hydrochloric acid is diluted to 1000 ml with distilled water), and then soaked in 3.3M potassium chloride solution for 24 hours to restore its performance.
Karry Zhuang 32.3 491
Karry Zhuang 29.1 492 (8) Electrode contamination or liquid junction blockage can also cause electrode passivation. At this time, it should be cleaned with an appropriate solution according to the nature of the contaminant. If the platinum of the electrode is severely contaminated and an oxide film is formed, toothpaste can be applied to the platinum surface and then gently scrubbed to restore the platinum's luster.
Karry Zhuang 32.3 493
Karry Zhuang 29.1 494 (9) The equipment should be calibrated before each use. It is recommended to calibrate once every 3 months for long-term use. The calibration frequency should be adjusted appropriately according to different application conditions (degree of dirt in the application, deposition of chemical substances, etc.). After aging, the electrodes should be replaced in time.
495
496 == 3.6 RS485 Commands ==
497
498
499 RS485 signaldefault address 0x13
500 Standard Modbus-RTU protocol, baud rate: 9600; check bit: none; data bit: 8; stop bit: 1
501
Karry Zhuang 33.2 502 === 3.6.1 Query address ===
Karry Zhuang 29.1 503
504 send
505
506 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:676.25px" %)
507 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Original address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 1px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 high
508 |(% style="width:99px" %)0XFE |(% style="width:112px" %)0X03|(% style="width:135px" %)0X00|(% style="width:126px" %)0X50|(% style="width:85px" %)0X00|(% style="width:1px" %)0X00|(% style="width:1px" %)0X51|(% style="width:1px" %)0XD4
509
510 response
511
512 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:561.333px" %)
513 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)New address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 106px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 93px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 104px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
514 |(% style="width:99px" %)0X01|(% style="width:112px" %)0X03|(% style="width:106px" %)0X00|(% style="width:93px" %)0X20|(% style="width:104px" %)0XF0
515
516 === 3.6.2 Change address ===
517
518 For example: Change the address of the sensor with address 1 to 2, master → slave
519
520 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:676.25px" %)
521 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Original address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 1px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 high
522 |(% style="width:99px" %)0X01|(% style="width:112px" %)0X06|(% style="width:135px" %)0X00|(% style="width:126px" %)0X50|(% style="width:85px" %)0X00|(% style="width:1px" %)0X02|(% style="width:1px" %)0X08|(% style="width:1px" %)0X1A
523
524 If the sensor receives correctly, the data is returned along the original path.
525 Note: If you forget the original address of the sensor, you can use the broadcast address 0XFE instead. When using 0XFE, the host can only connect to one slave, and the return address is still the original address, which can be used as a method of address query.
526
527
528 === 3.6.3 Modify intercept ===
529
530 send
531
532 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:676.25px" %)
Karry Zhuang 35.1 533 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 67px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address  low|=(% style="width: 69px; background-color: rgb(79, 129, 189); color: white;" %)Register Length high|=(% style="width: 66px; background-color: rgb(79, 129, 189); color: white;" %)Register Length low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 high
Karry Zhuang 29.1 534 |(% style="width:99px" %)0X13|(% style="width:112px" %)0X06|(% style="width:135px" %)0X00|(% style="width:126px" %)0X10|(% style="width:85px" %)0X00|(% style="width:1px" %)0X64|(% style="width:1px" %)0X8A|(% style="width:1px" %)(((
535 0X96
536 )))
537
538 Change the intercept of the sensor with address 1 to 10 (default 0), which is 0X000A in the command.
539
540 response
541
542 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:676.25px" %)
543 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 1px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 high
544 |(% style="width:99px" %)0X13|(% style="width:112px" %)0X06|(% style="width:135px" %)(((
545 0X00
546 )))|(% style="width:126px" %)0X10|(% style="width:85px" %)0X00|(% style="width:1px" %)0X64|(% style="width:1px" %)0X8A|(% style="width:1px" %)(((
547 0X96
548 )))
549
550 === 3.6.4 Query data ===
551
Edwin Chen 9.1 552
Karry Zhuang 37.1 553 Query the data (ORP) of the sensor (address 13), host → slave
Karry Zhuang 35.1 554
555 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
556 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
557 |(% style="width:99px" %)0X13|(% style="width:72px" %)0X03|(% style="width:64px" %)0X00|(% style="width:68px" %)0X00|(% style="width:70px" %)0X00|(% style="width:72px" %)0X01|(% style="width:56px" %)0X87|(% style="width:56px" %)0X78
558
559 If the sensor receives correctly, the following data will be returned, slave → host
560
561 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
562 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
563 |(% style="width:99px" %)0X13|(% style="width:72px" %)0X03|(% style="width:68px" %)0X02|(% style="width:70px" %)0X02|(% style="width:72px" %)0XAE|(% style="width:56px" %)0X80|(% style="width:56px" %)0X9B
564
Karry Zhuang 11.1 565 The query data command is 13 03 00 00 00 01 87 78
Karry Zhuang 10.1 566
Karry Zhuang 11.1 567 For example, the returned data is 13 03 02 (% style="color:red" %)**02 AE**(%%) 80 9B.
Karry Zhuang 10.1 568
Karry Zhuang 11.1 569 02 AE is the ORP value, converted to decimal, the actual value is 686, 02 AE means the current ORP value is 686mV
570
571
Karry Zhuang 29.1 572 === 3.6.5 Calibration Method ===
573
574 This device uses two-point calibration, and two known ORP standard solutions need to be prepared. The calibration steps are as follows:
575 (1) Place the electrode in distilled water to clean it, and then place it in 86mV standard buffer solution. After the data stabilizes,
576 enter the following calibration command, and the 86mV point calibration is completed;
577
578 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:575.333px" %)
579 |=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 66px; background-color: rgb(79, 129, 189); color: white;" %)Address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Address low|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Quantity high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
580 |(% style="width:64px" %)0X13|(% style="width:72px" %)0X06|(% style="width:66px" %)(((
581 0X00
582 )))|(% style="width:68px" %)0X24|(% style="width:72px" %)0XFF|(% style="width:70px" %)0XFF|(% style="width:55px" %)0XCB|(% style="width:55px" %)(((
583 0X03
584 )))
585
586 Wash the electrode in distilled water and place it in 256mV standard buffer. After the data is stable, enter the following calibration command to complete the 256mV point calibration.
587
588 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:575.333px" %)
589 |=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 66px; background-color: rgb(79, 129, 189); color: white;" %)Address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Address low|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Quantity high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 55px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
590 |(% style="width:64px" %)0X13|(% style="width:72px" %)0X06|(% style="width:66px" %)(((
591 0X00
592 )))|(% style="width:68px" %)0X25|(% style="width:72px" %)0XFF|(% style="width:70px" %)0XFF|(% style="width:55px" %)0X9A|(% style="width:55px" %)(((
593 0XC3
594 )))
595
Edwin Chen 8.1 596 = 4. DR-DO1 Dissolved Oxygen Sensor =
597
598
Karry Zhuang 11.1 599
Karry Zhuang 32.1 600 == 4.1 Specification ==
601
Karry Zhuang 33.2 602
603 * **Measuring range**: 0-20mg/L, 0-50℃
604 * **Accuracy**: 3%, ±0.5℃
605 * **Resolution**: 0.01 mg/L, 0.01℃
606 * **Maximum operating pressure**: 6 bar
607 * **Output signal**: A: 4-20mA (current loop)B: RS485 (standard Modbus-RTU protocol, device default address: 01)
608 * **Power supply voltage**: 5-24V DC
609 * **Working environment**: temperature 0-60℃; humidity <95%RH
Karry Zhuang 32.1 610 * **Power consumption**: ≤0.5W
611
Karry Zhuang 33.2 612 == 4.2 wiring ==
Karry Zhuang 32.1 613
Karry Zhuang 45.1 614 [[image:image-20240720172632-4.png||height="390" width="640"]]
Karry Zhuang 33.2 615
616
617 == (% id="cke_bm_224234S" style="display:none" %) (%%)4.3 Impedance requirements for current signals ==
618
Karry Zhuang 32.1 619 [[image:image-20240718195414-8.png||height="100" width="575"]]
620
621
622 == 4.4 Mechinical Drawing ==
623
624
Karry Zhuang 33.2 625 [[image:image-20240719155308-1.png||height="226" width="527"]]
Karry Zhuang 32.1 626
Karry Zhuang 33.2 627
Karry Zhuang 39.1 628 == 4.5 Instructions for use and maintenance ==
Karry Zhuang 32.1 629
630 * It can be directly put into water without adding a protective tube, ensuring the long-term stability, reliability and accuracy of the sensor.
631 * If the water conditions are complex and you want accurate data, you need to wipe the sensor probe frequently.
632
633 == 4.6 RS485 Commands ==
634
Karry Zhuang 34.1 635 RS485 signaldefault address 0x14
636 Standard Modbus-RTU protocol, baud rate: 9600; check bit: none; data bit: 8; stop bit: 1
637
Karry Zhuang 32.3 638 === 4.6.1 Query address ===
Karry Zhuang 32.1 639
Karry Zhuang 32.3 640 send
641
642 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
Karry Zhuang 34.1 643 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Original address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
644 |(% style="width:99px" %)0XFF|(% style="width:72px" %)0X03|(% style="width:64px" %)0X00|(% style="width:68px" %)0X0A|(% style="width:70px" %)0X00|(% style="width:72px" %)0X02|(% style="width:56px" %)0XF1|(% style="width:56px" %)0XD7
Karry Zhuang 32.3 645
Karry Zhuang 34.1 646 If you forget the original address of the sensor, you can use the broadcast address 0XFF instead. When using 0XFE, the host can only connect to one slave, which can be used as a method of address query.
Karry Zhuang 32.3 647
648
649 response
650
Karry Zhuang 34.1 651 Register 0 data high and register 0 data low indicate the actual address of the sensor: 1
652 Register 1 data high and register 1 data low indicate the sensor version
Karry Zhuang 32.3 653
Karry Zhuang 34.1 654 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
655 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register 1 Data high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Register 1 Data low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
656 |(% style="width:99px" %)0XFF|(% style="width:72px" %)0X03|(% style="width:64px" %)0X04|(% style="width:68px" %)0X00|(% style="width:70px" %)0X01|(% style="width:72px" %)0X00|(% style="width:56px" %)0X00|(% style="width:56px" %)0XB4|(% style="width:56px" %)0X3C
657
Karry Zhuang 33.2 658 === 4.6.2 Change address ===
Karry Zhuang 32.3 659
Karry Zhuang 34.1 660 For example: Change the address of the sensor with address 1 to 2(address range: 1-119), master → slave
Karry Zhuang 33.2 661
662 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:907.333px" %)
663 |=(% style="width: 67px; background-color: rgb(79, 129, 189); color: white;" %)Original address|=(% style="width: 71px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 65px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 65px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 53px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 53px; background-color: rgb(79, 129, 189); color: white;" %)Start address high|=(% style="width: 53px; background-color: rgb(79, 129, 189); color: white;" %)Start address low|=(% style="width: 53px; background-color: rgb(79, 129, 189); color: white;" %)Sensor version|=(% style="width: 53px; background-color: rgb(79, 129, 189); color: white;" %)Sensor version|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low
664 |(% style="width:67px" %)0X01|(% style="width:71px" %)0X10|(% style="width:65px" %)0X00|(% style="width:65px" %)0X0A|(% style="width:70px" %)0X00|(% style="width:72px" %)0X02|(% style="width:53px" %)0X04|(% style="width:53px" %)0X00|(% style="width:72px" %)0X02|(% style="width:53px" %)0X00|(% style="width:53px" %)0X00|(% style="width:56px" %)0XD2|(% style="width:53px" %)0X10
665
Karry Zhuang 34.1 666 response
Karry Zhuang 32.3 667
Karry Zhuang 34.1 668 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
669 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
670 |(% style="width:99px" %)0X01|(% style="width:72px" %)0X10|(% style="width:64px" %)0X00|(% style="width:68px" %)0X0A|(% style="width:70px" %)0X00|(% style="width:72px" %)0X02|(% style="width:56px" %)0X61|(% style="width:56px" %)0XCA
Karry Zhuang 32.3 671
Karry Zhuang 34.1 672 === 4.6.3 Query data ===
Edwin Chen 9.1 673
674
Karry Zhuang 34.2 675 Query the data (dissolved oxygen) of the sensor (address 14), host → slave
Karry Zhuang 34.1 676
677 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
678 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
679 |(% style="width:99px" %)0X14|(% style="width:72px" %)0X03|(% style="width:64px" %)0X00|(% style="width:68px" %)0X14|(% style="width:70px" %)0X00|(% style="width:72px" %)0X01|(% style="width:56px" %)0XC6|(% style="width:56px" %)0XCB
680
Karry Zhuang 34.2 681 If the sensor receives correctly, the following data will be returned, slave → host
Karry Zhuang 34.1 682
683 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
684 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
Karry Zhuang 34.2 685 |(% style="width:99px" %)0X14|(% style="width:72px" %)0X03|(% style="width:68px" %)0X02|(% style="width:70px" %)0X03|(% style="width:72px" %)0X78|(% style="width:56px" %)0XB5|(% style="width:56px" %)0X55
Karry Zhuang 34.1 686
Karry Zhuang 11.1 687 After the query, 7 bytes will be returned. For example, the returned data is 14 03 02 (% style="color:red" %)**03 78**(%%) B5 55. 03 78 is the value of dissolved oxygen.
Karry Zhuang 10.1 688
689 Converted to decimal, it is 888. Add two decimal places to get the actual value. 03 78 means the current dissolved oxygen is 8.88mg/L
690
Karry Zhuang 11.1 691
Karry Zhuang 34.2 692 Query the data (temperature) of the sensor (address 14), host → slave
693
694 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
695 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
696 |(% style="width:99px" %)0X14|(% style="width:72px" %)0X03|(% style="width:64px" %)0X00|(% style="width:68px" %)0X11|(% style="width:70px" %)0X00|(% style="width:72px" %)0X01|(% style="width:56px" %)0XD6|(% style="width:56px" %)0XCA
697
698 If the sensor receives correctly, the following data will be returned, slave → host
699
700 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
701 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
702 |(% style="width:99px" %)0X14|(% style="width:72px" %)0X03|(% style="width:68px" %)0X02|(% style="width:70px" %)0X09|(% style="width:72px" %)0XA4|(% style="width:56px" %)0XB2|(% style="width:56px" %)0X6C
703
704 After the query, 7 bytes will be returned. For example, the returned data is 14 03 02 (% style="color:red" %)**09 A4**(%%) B2 6C. 03 78 is the value of dissolved oxygen temperature.
705
706 Converted to decimal, it is 2468. Add two decimal places to get the actual value. 09 A4 means the current dissolved oxygen temperature is 24.68℃
707
708
Edwin Chen 8.1 709 = 5. DR-TS1 Water Turbidity Sensor =
710
Edwin Chen 9.1 711
Karry Zhuang 10.1 712
Karry Zhuang 32.3 713 == (% id="cke_bm_81470S" style="display:none" %) (%%)5.1 Specification ==
714
715 * **Measuring range**: 0.1~1000.0NTU
716 * **Accuracy**: ±5%
717 * **Resolution**: 0.1NTU
718 * **Stability**: ≤3mV/24 hours
719 * **Output signal**: A: 4~20 mA (current loop)B: RS485 (standard Modbus-RTU protocol, device default address: 01)
720 * **Power supply voltage**: 5~24V DC (when output signal is RS485)12~24V DC (when output signal is 4~20mA)
721 * **Working environment**: temperature 0~60℃; humidity ≤95%RH
722 * **Power consumption**: ≤0.5W
723
724 == 5.2 wiring ==
725
Karry Zhuang 45.1 726 [[image:image-20240720172640-5.png||height="387" width="635"]]
Karry Zhuang 32.3 727
Karry Zhuang 45.1 728
Karry Zhuang 32.3 729 == 5.3 Impedance requirements for current signals ==
730
731 [[image:image-20240718195414-8.png||height="100" width="575"]]
732
733
734 == 5.4 Mechinical Drawing ==
735
736 [[image:image-20240718195058-7.png||height="305" width="593"]]
737
738
Karry Zhuang 39.1 739 == 5.5 Instructions for use and maintenance ==
Karry Zhuang 32.3 740
741 * It can be directly put into water without adding a protective tube, ensuring the long-term stability, reliability and accuracy of the sensor.
742 * If the water conditions are complex and you want accurate data, you need to wipe the sensor probe frequently.
743
744 == 5.6 RS485 Commands ==
745
746
Karry Zhuang 36.1 747 RS485 signaldefault address 0x15
748 Standard Modbus-RTU protocol, baud rate: 9600; check bit: none; data bit: 8; stop bit: 1
749
750 === 5.6.1 Query address ===
751
Karry Zhuang 32.3 752 send
753
754 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
755 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Original address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Quantity high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
756 |(% style="width:99px" %)0XFE |(% style="width:72px" %)0X03|(% style="width:64px" %)0X00|(% style="width:68px" %)0X50|(% style="width:70px" %)0X00|(% style="width:72px" %)0X00|(% style="width:56px" %)0X51|(% style="width:56px" %)0XD4
757
758 If you forget the original address of the sensor, you can use the broadcast address 0XFE instead. When using 0XFE, the host can only connect to one slave, which can be used as a method of address query.
759
760
761 response
762
763 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:561.333px" %)
764 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)New address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 106px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 93px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 104px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
765 |(% style="width:99px" %)0X01|(% style="width:112px" %)0X03|(% style="width:106px" %)0X00|(% style="width:93px" %)0X20|(% style="width:104px" %)0XF0
766
Karry Zhuang 36.1 767 === 5.6.2 Change address ===
Karry Zhuang 32.3 768
Karry Zhuang 36.1 769 For example: Change the address of the sensor with address 1 to 2, master → slave
Karry Zhuang 10.1 770
Karry Zhuang 36.1 771 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:676.25px" %)
772 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Original address|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 1px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 50px;background-color:#4F81BD;color:white" %)CRC16 high
773 |(% style="width:99px" %)0X01|(% style="width:112px" %)0X06|(% style="width:135px" %)0X00|(% style="width:126px" %)0X50|(% style="width:85px" %)0X00|(% style="width:1px" %)0X02|(% style="width:1px" %)0X08|(% style="width:1px" %)0X1A
774
775 If the sensor receives correctly, the data is returned along the original path.
776 Note: If you forget the original address of the sensor, you can use the broadcast address 0XFE instead. When using 0XFE, the host can only connect to one slave, and the return address is still the original address, which can be used as a method of address query.
777
778 === 5.6.3 Query data ===
779
780
781 Query the data (turbidity) of the sensor (address 15), host → slave
782
783 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
784 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 64px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Starting register address low|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register length high|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Register length low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
785 |(% style="width:99px" %)0X15|(% style="width:72px" %)0X03|(% style="width:64px" %)0X00|(% style="width:68px" %)0X00|(% style="width:70px" %)0X00|(% style="width:72px" %)0X01|(% style="width:56px" %)0X87|(% style="width:56px" %)0X1E
786
787 If the sensor receives correctly, the following data will be returned, slave → host
788
789 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:599.333px" %)
790 |=(% style="width: 50px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 72px; background-color: rgb(79, 129, 189); color: white;" %)Function code|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data high|=(% style="width: 70px; background-color: rgb(79, 129, 189); color: white;" %)Register 0 Data low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 56px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
791 |(% style="width:99px" %)0X15|(% style="width:72px" %)0X03|(% style="width:68px" %)0X02|(% style="width:70px" %)0X02|(% style="width:72px" %)0X9A|(% style="width:56px" %)0X09|(% style="width:56px" %)0X4C
792
Karry Zhuang 11.1 793 The query data command is 15 03 00 00 00 01 87 1E
Karry Zhuang 10.1 794
Karry Zhuang 11.1 795 For example, the returned data is 15 03 02 (% style="color:red" %)**02 9A**(%%) 09 4C
796
797 02 9A is the turbidity value, converted to decimal, it is 666, and then divided by 10, the actual value is 66.6, 02 9A means the current turbidity value is 66.6 NTU
Copyright ©2010-2024 Dragino Technology Co., LTD. All rights reserved