2
3 {{toc/}}
4
5
Edwin Chen 7.1 6 = 1. DR-ECK Water EC Probe =
7
8 == 1.1 Specification: ==
9
10 * **Power Input**: DC7~~30
11 * **Power Consumption** : < 0.5W
12 * **Interface**: RS485. 9600 Baud Rate
13 * **EC Range & Resolution:**
14 ** **ECK0.01** : 0.02 ~~ 20 μS/cm
15 ** **ECK0.1**: 0.2 ~~ 200.0 μS/cm
16 ** **ECK1.0** : 2 ~~ 2,000 μS/cm  Resolution: 1 μS/cm
17 ** **ECK10.0** : 20 ~~ 20,000 μS/cm  Resolution: 10 μS/cm
18 * **EC Accuracy**: ±1% FS
19 * **Temperature Measure Range**: -20 ~~ 60 °C
20 * **Temperature Accuracy: **±0.5 °C
21 * **IP Rated**: IP68
22 * **Max Pressure**: 0.6MPa
23
24 == 1.2 Application for Different Range ==
25
26 [[image:image-20240714173018-1.png]]
27
28
29 == 1.3 Wiring ==
30
31
32 == 1.4 Mechinical Drawing ==
33
34 [[image:image-20240714174241-2.png]]
35
36
37 == 1.5 Installation ==
38
39
40 == 1.6 Maintain ==
41
42
Edwin Chen 8.1 43 == 1.7 RS485 Commands ==
44
Karry Zhuang 12.1 45 === 1.7.1 Query data ===
Edwin Chen 8.1 46
Karry Zhuang 11.1 47 The address of the EC K10 sensor is 11
48
Karry Zhuang 10.1 49 The query data command is 11 03 00 00 00 02 C6 9B
50
Karry Zhuang 11.1 51 For example, the returned data is 11 03 04 (% style="color:red" %)**02 AE**(%%) 01 64 8B D0. 02 AE is converted to decimal 686,  K=10, EC: 6860uS/cm
Karry Zhuang 10.1 52
53
54 The address of the EC K1 sensor is 12
55
56 The query data command is 12 03 00 00 00 02 C6 A8
57
Karry Zhuang 11.1 58 For example, the returned data is 12 03 04 (% style="color:red" %)**02 AE**(%%) 01 64 B8 D0. 02 AE is converted to decimal 686,  K=1, EC: 686uS/cm.
Karry Zhuang 10.1 59
Karry Zhuang 11.1 60
Karry Zhuang 12.1 61 === 1.7.2 Calibration Method ===
62
63
64 This device uses one-point calibration, and you need to prepare a known E standard solution. When mileage K=1, 1~~2000 uses 1413uS/cm standard solution, and when mileage K=10, 10~~20000 uses 12.88mS/cm standard solution. The calibration steps are as follows:
65 (1) Place the electrode in distilled water and clean it. When mileage 1~~2000 uses 1413HS/cm standard solution, enter the following calibration command after the data is stable.
66
67
68 (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:500px" %)
69 |=(% style="width: 70px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 60px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 80px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 80px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 60px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 50px;background-color:#4F81BD;color:white" %)Quantity low
70 |(% style="width:99px" %) |(% style="width:112px" %) |(% style="width:135px" %) |(% style="width:126px" %) |(% style="width:85px" %) |(% style="width:46px" %)
71
Edwin Chen 8.1 72 = 2. DR-PH01 Water PH Sensor =
73
Edwin Chen 9.1 74 == 2.7 RS485 Commands ==
Edwin Chen 8.1 75
Edwin Chen 9.1 76
Karry Zhuang 11.1 77 The address of the pH  sensor is 10
Edwin Chen 9.1 78
Karry Zhuang 11.1 79 The query data command is 10 03 00 00 00 01 87 4B. After the query, 7 bytes will be returned.
Karry Zhuang 10.1 80
Karry Zhuang 11.1 81 For example, the returned data is 10 03 02 (% style="color:red" %)**02 AE**(%%) C4 9B.
82
83 02 AE is the pH value, which is converted into decimal to get 686, and then two decimal places are added to get the actual value. 02 AE means the current pH value is 6.86.
84
85
Edwin Chen 8.1 86 = 3. DR-ORP1 Water ORP Sensor =
87
Edwin Chen 9.1 88 == 3.7 RS485 Commands ==
Edwin Chen 8.1 89
Edwin Chen 9.1 90
Karry Zhuang 11.1 91 The address of the ORP sensor is 13
Edwin Chen 9.1 92
Karry Zhuang 11.1 93 The query data command is 13 03 00 00 00 01 87 78
Karry Zhuang 10.1 94
Karry Zhuang 11.1 95 For example, the returned data is 13 03 02 (% style="color:red" %)**02 AE**(%%) 80 9B.
Karry Zhuang 10.1 96
Karry Zhuang 11.1 97 02 AE is the ORP value, converted to decimal, the actual value is 686, 02 AE means the current ORP value is 686mV
98
99
Edwin Chen 8.1 100 = 4. DR-DO1 Dissolved Oxygen Sensor =
101
Edwin Chen 9.1 102 == 4.7 RS485 Commands ==
Edwin Chen 8.1 103
Karry Zhuang 11.1 104
Karry Zhuang 10.1 105 The address of the dissolved oxygen sensor is 14
Edwin Chen 9.1 106
Karry Zhuang 10.1 107 The query data command is 14 03 00 14 00 01 C6 CB
Edwin Chen 9.1 108
Karry Zhuang 11.1 109 After the query, 7 bytes will be returned. For example, the returned data is 14 03 02 (% style="color:red" %)**03 78**(%%) B5 55. 03 78 is the value of dissolved oxygen.
Karry Zhuang 10.1 110
111 Converted to decimal, it is 888. Add two decimal places to get the actual value. 03 78 means the current dissolved oxygen is 8.88mg/L
112
Karry Zhuang 11.1 113
Edwin Chen 8.1 114 = 5. DR-TS1 Water Turbidity Sensor =
115
Edwin Chen 9.1 116 == 5.7 RS485 Commands ==
117
Karry Zhuang 10.1 118
119 The address of the dissolved oxygen sensor is 15
120
Karry Zhuang 11.1 121 The query data command is 15 03 00 00 00 01 87 1E
Karry Zhuang 10.1 122
Karry Zhuang 11.1 123 For example, the returned data is 15 03 02 (% style="color:red" %)**02 9A**(%%) 09 4C
124
125 02 9A is the turbidity value, converted to decimal, it is 666, and then divided by 10, the actual value is 66.6, 02 9A means the current turbidity value is 66.6 NTU
126
Edwin Chen 7.1 127