... ... @@ -259,7 +259,19 @@
259 259  
260 260  ECK200.0
261 261  
262 +For the device with address 01, use 1413uS/cm standard solution to calibrate the first point. Send frame: 1413. Convert hexadecimal to 585. Write 0001, 00 00, 0585 to 0x0120, 0x0121, 0x0122 respectively.
262 262  
264 +(% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
265 +|=(% style="width: 42px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 53px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 53px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Data|=(% style="width: 53px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 53px;background-color:#4F81BD;color:white" %)CRC16 high
266 +|(% style="width:99px" %)0X01|(% style="width:112px" %)0X10|(% style="width:135px" %)0X01|(% style="width:126px" %)0X02|(% style="width:85px" %)0X00|(% style="width:1px" %)0X03|(% style="width:1px" %)0X04|(% style="width:1px" %)(((
267 +0X00
268 +0X00
269 +0X37
270 +0X32
271 +)))|(% style="width:1px" %)0XBD|(% style="width:1px" %)0XFC
272 +
273 +
274 +
263 263  = 2. DR-PH01 Water PH Sensor =
264 264  