... ... @@ -28,7 +28,6 @@
28 28  * **Working environment:**
29 29  ** Ambient Temperature: 0–60°C
30 30  ** Relative Humidity: <85% RH(Specifically refers to the cable male and female)
31 -** ECK200.0 Continuous monitoring of cross-section water quality, aquaculture, sewage treatment, environmental protection, pharmaceuticals, food, tap water, seawater and other high conductivity environments
32 32  * **IP Rated**: IP68
33 33  
34 34  * **Max Pressure**: 0.6MPa
... ... @@ -47,12 +47,10 @@
47 47  
48 48  == 1.4 Mechinical Drawing ==
49 49  
50 - ECK1 and ECK10  ECK200
51 51  
50 +[[image:image-20240714174241-2.png]]
52 52  
53 -[[image:image-20240714174241-2.png]] [[image:1752564223905-283.png||height="399" width="160"]]
54 54  
55 -
56 56  == 1.5 Installation ==
57 57  
58 58  
... ... @@ -127,8 +127,6 @@
127 127  |=(% style="width: 100px;background-color:#4F81BD;color:white" %)New address|=(% style="width: 110px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 106px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 93px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 104px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
128 128  |(% style="width:99px" %)0X01|(% style="width:112px" %)0X03|(% style="width:106px" %)0X00|(% style="width:93px" %)0X20|(% style="width:104px" %)0XF0
129 129  
130 -
131 -
132 132  === 1.7.2 Change address ===
133 133  
134 134  
... ... @@ -207,11 +207,8 @@
207 207  **For example**, the returned data is 12 03 04 (% style="color:red" %)**02 AE**(%%) 01 64 B8 D0. 02 AE is converted to decimal 686,  K=1, EC: 686uS/cm,temperature: 35.6℃ Convert the returned data to decimal and divide by 10.
208 208  
209 209  
210 -ECK200
211 -
212 212  === 1.7.5 Calibration Method ===
213 213  
214 -ECK1 and ECK10.0
215 215  
216 216  This device uses one-point calibration, and you need to prepare a known E standard solution. When mileage K=1, 1~~2000 uses 1413μS/cm standard solution, and when mileage K=10, 10~~20000 uses 12.88mS/cm standard solution.
217 217  
... ... @@ -255,11 +255,6 @@
255 255  |=(% style="width: 42px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 68px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 68px;background-color:#4F81BD;color:white" %)CRC16 high
256 256  |(% style="width:99px" %)0X11|(% style="width:112px" %)0X06|(% style="width:135px" %)0X00|(% style="width:126px" %)0X26|(% style="width:85px" %)0X00|(% style="width:1px" %)0X02|(% style="width:1px" %)0XEB|(% style="width:1px" %)0X50
257 257  
258 -
259 -
260 -ECK200.0
261 -
262 -
263 263  = 2. DR-PH01 Water PH Sensor =
264 264  