Changes for page Water Quality Sensors

Last modified by Karry Zhuang on 2025/07/18 16:37

From version 60.10
edited by Karry Zhuang
on 2025/07/15 15:42
Change comment: There is no comment for this version
To version 61.3
edited by Karry Zhuang
on 2025/07/15 15:54
Change comment: There is no comment for this version

Summary

Details

Page properties
Content
... ... @@ -211,6 +211,7 @@
211 211  
212 212  === 1.7.5 Calibration Method ===
213 213  
214 +ECK1 and ECK10.0
214 214  
215 215  This device uses one-point calibration, and you need to prepare a known E standard solution. When mileage K=1, 1~~2000 uses 1413μS/cm standard solution, and when mileage K=10, 10~~20000 uses 12.88mS/cm standard solution.
216 216  
... ... @@ -254,6 +254,23 @@
254 254  |=(% style="width: 42px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 68px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 68px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 68px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 68px;background-color:#4F81BD;color:white" %)CRC16 high
255 255  |(% style="width:99px" %)0X11|(% style="width:112px" %)0X06|(% style="width:135px" %)0X00|(% style="width:126px" %)0X26|(% style="width:85px" %)0X00|(% style="width:1px" %)0X02|(% style="width:1px" %)0XEB|(% style="width:1px" %)0X50
256 256  
258 +
259 +
260 +ECK200.0
261 +
262 +For the device with address 01, use 1413uS/cm standard solution to calibrate the first point. Send frame: 1413. Convert hexadecimal to 585. Write 0001, 00 00, 0585 to 0x0120, 0x0121, 0x0122 respectively.
263 +
264 +(% border="1" cellspacing="3" style="background-color:#f2f2f2; width:518px" %)
265 +|=(% style="width: 42px;background-color:#4F81BD;color:white" %)Address|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Address high|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Address low|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Quantity high|=(% style="width: 53px; background-color: rgb(79, 129, 189); color: white;" %)Quantity low|=(% style="width: 53px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 53px;background-color:#4F81BD;color:white" %)Data|=(% style="width: 53px;background-color:#4F81BD;color:white" %)CRC16 low|=(% style="width: 53px;background-color:#4F81BD;color:white" %)CRC16 high
266 +|(% style="width:99px" %)0X01|(% style="width:112px" %)0X10|(% style="width:135px" %)0X01|(% style="width:126px" %)0X02|(% style="width:85px" %)0X00|(% style="width:1px" %)0X03|(% style="width:1px" %)0X04|(% style="width:1px" %)(((
267 +0X00
268 +0X00
269 +0X37
270 +0X32
271 +)))|(% style="width:1px" %)0XBD|(% style="width:1px" %)0XFC
272 +
273 +
274 +
257 257  = 2. DR-PH01 Water PH Sensor =
258 258  