Changes for page Water Quality Sensors

Last modified by Karry Zhuang on 2025/07/18 16:37

From version 56.1
edited by Karry Zhuang
on 2024/12/24 09:55
Change comment: There is no comment for this version
To version 60.8
edited by Karry Zhuang
on 2025/07/15 15:39
Change comment: There is no comment for this version

Summary

Details

Page properties
Content
... ... @@ -21,9 +21,14 @@
21 21  ** **ECK0.1**: 0.2 ~~ 200.0 μS/cm
22 22  ** **ECK1.0** : 0 ~~ 2,000 μS/cm  Resolution: 1 μS/cm
23 23  ** **ECK10.0** : 10 ~~ 20,000 μS/cm  Resolution: 10 μS/cm
24 +** **ECK200.0** : 1 ~~ 200,000 μS/cm  Resolution: 1 μS/cm
24 24  
25 25  * **EC Accuracy**: ±1% FS
26 26  * **Temperature Accuracy: **±0.5 °C
28 +* **Working environment:**
29 +** Ambient Temperature: 0–60°C
30 +** Relative Humidity: <85% RH(Specifically refers to the cable male and female)
31 +** ECK200.0 Continuous monitoring of cross-section water quality, aquaculture, sewage treatment, environmental protection, pharmaceuticals, food, tap water, seawater and other high conductivity environments
27 27  * **IP Rated**: IP68
28 28  
29 29  * **Max Pressure**: 0.6MPa
... ... @@ -42,10 +42,12 @@
42 42  
43 43  == 1.4 Mechinical Drawing ==
44 44  
50 + ECK1 and ECK10  ECK200
45 45  
46 -[[image:image-20240714174241-2.png]]
47 47  
53 +[[image:image-20240714174241-2.png]] [[image:1752564223905-283.png||height="399" width="160"]]
48 48  
55 +
49 49  == 1.5 Installation ==
50 50  
51 51  
... ... @@ -104,6 +104,7 @@
104 104  
105 105  === 1.7.1 Query address ===
106 106  
114 +ECK1 and ECK10.0
107 107  
108 108  **send:**
109 109  
... ... @@ -120,6 +120,10 @@
120 120  |=(% style="width: 100px;background-color:#4F81BD;color:white" %)New address|=(% style="width: 110px;background-color:#4F81BD;color:white" %)Function code|=(% style="width: 106px; background-color: rgb(79, 129, 189); color: white;" %)Data length|=(% style="width: 93px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 low|=(% style="width: 104px; background-color: rgb(79, 129, 189); color: white;" %)CRC16 high
121 121  |(% style="width:99px" %)0X01|(% style="width:112px" %)0X03|(% style="width:106px" %)0X00|(% style="width:93px" %)0X20|(% style="width:104px" %)0XF0
122 122  
131 +
132 +
133 +ECK200.0
134 +
123 123  === 1.7.2 Change address ===
124 124  
125 125  
... ... @@ -264,7 +264,9 @@
264 264  
265 265  * **Temperature measurement error**: ±0.5°C
266 266  
267 -* **Temperature Measure Range**: -20 ~~ 60 °C
279 +* **Working environment:**
280 +** Ambient Temperature: 0–60°C
281 +** Relative Humidity: <85% RH(Specifically refers to the cable male and female)
268 268  
269 269  * **Temperature Accuracy: **±0.5 °C
270 270  
... ... @@ -471,7 +471,9 @@
471 471  
472 472  * **Stability**: ≤2mv/24 hours
473 473  
474 -* **Equipment working conditions**: Ambient temperature: 0-60°C Relative humidity: <85%RH
488 +* **Working environment:**
489 +** Ambient Temperature: 0–60°C
490 +** Relative Humidity: <85% RH(Specifically refers to the cable male and female)
475 475  
476 476  * **IP Rated**: IP68
477 477  
... ... @@ -642,11 +642,11 @@
642 642  == 4.1 Specification ==
643 643  
644 644  
645 -* **Measuring range**: 0-20mg/L, 0-50°C
661 +* **Measuring range**: 0-20mg/L, 050
646 646  
647 -* **Accuracy**: 3%, ±0.5°C
663 +* **Accuracy**: 3%, ±0.5
648 648  
649 -* **Resolution**: 0.01 mg/L, 0.01°C
665 +* **Resolution**: 0.01 mg/L, 0.01
650 650  
651 651  * **Maximum operating pressure**: 6 bar
652 652  
... ... @@ -654,7 +654,9 @@
654 654  
655 655  * **Power supply voltage**: 5-24V DC
656 656  
657 -* **Working environment**: temperature 0-60°C; humidity <95%RH
673 +* **Working environment:**
674 +** Ambient Temperature: 0–60°C
675 +** Relative Humidity: <85% RH(Specifically refers to the cable male and female)
658 658  
659 659  * **Power consumption**: ≤0.5W
660 660  
... ... @@ -781,7 +781,9 @@
781 781  
782 782  * **Power supply voltage**: 5~~24V DC (when output signal is RS485), 12~~24V DC (when output signal is 4~~20mA)
783 783  
784 -* **Working environment**: temperature 060°C; humidity ≤ 95%RH
802 +* **Working environment:**
803 +** Ambient Temperature: 0–60°C
804 +** Relative Humidity: <85% RH(Specifically refers to the cable male and female)
785 785  
786 786  * **Power consumption**: ≤ 0.5W
787 787  
1752564223905-283.png
Author
... ... @@ -1,0 +1,1 @@
1 +XWiki.karry
Size
... ... @@ -1,0 +1,1 @@
1 +144.4 KB
Content