... ... @@ -1222,7 +1222,7 @@
1222 1222  * **Power supply**: DC7~~30V
1223 1223  * **Power consumption**: ≤0.5W
1224 1224  * **Communication interface**: RS485; standard MODBUS-RTU protocol; communication baud rate: default 9600
1225 -* **Conductivity measurement range**: K=1: 0~~2000μs/cm; resolution: 1μs/cm K=10: 10~~20000μs/cm; resolution: 10μs/cm
1225 +* **Conductivity measurement range**: K=1: 0~~2000μs/cm; resolution: 1μs/cm K=10: 10~~20000μs/cm; resolution: 10μs/cm 200:0-200000 μS/cm; resolution: 1 μS/cm
1226 1226  * **Conductivity measurement error**: ±1%FS
1227 1227  * **Working environment:**
1228 1228  ** Ambient Temperature: 0–60°C
... ... @@ -1554,6 +1554,48 @@
1554 1554  * The sensor is a precision device. When using it, please do not disassemble it by yourself or contact the sensor surface with sharp objects or corrosive liquids to avoid damaging the product.
1555 1555  * Do not power on while connect the cables. Double check the wiring before power on.
1556 1556  
1557 +
1558 +
1559 += 4.6 Residual Chlorine Sensor =
1560 +
1561 +== 4.6.1 Specification: ==
1562 +
1563 +* **Power Input**: DC7~~30
1564 +
1565 +* **Power Consumption** : 0.19W
1566 +
1567 +* **Interface**: RS485. 9600 Baud Rate
1568 +
1569 +* **CL Range & Resolution:**
1570 +** **CL2ML:**0-2mg/L
1571 +** **CL10ML:**0-10mg/L
1572 +** **Resolution:**0.01mg/L
1573 +
1574 +* **CL Accuracy**: ±5% FS
1575 +* **Temperature Accuracy: **±0.5 °C
1576 +* **Working environment:**
1577 +** Ambient Temperature: 0–50°C
1578 +** pH:4-9
1579 +** Flow rate: 30L/h~~60L/h (flow tank installation)
1580 +* **IP Rated**: IP68
1581 +
1582 +* **Max Pressure**: 0.6MPa
1583 +
1584 +== 4.6.2 Mechinical Drawing ==
1585 +
1586 +[[image:1752573238705-910.png||height="467" width="187"]]
1587 +
1588 +== 4.6.3 Maintenance ==
1589 +
1590 +* The device itself generally does not require daily maintenance. When an obvious fault occurs, please do not open it and repair it yourself, and contact us as soon as possible!
1591 +* After using the electrode, please clean the electrode head with clean water and cover it with a protective cover.
1592 +* When measuring the device, the measured liquid should flow and the flow rate should be uniform, and there should be no bubbles attached to the measuring end of the device.
1593 +* If the electrode diaphragm is attached with dirt and mineral components, the sensitivity will be reduced, and it may not be possible to perform sufficient measurement. Please ensure that the platinum ring is clean.
1594 +* The platinum induction ring of a good residual chlorine electrode should always be kept clean and bright. If the platinum ring of the electrode becomes rough or covered with pollutants after measurement, please clean it according to the following method: (For reference) Inorganic pollution: immerse the electrode in 0.1mol/L dilute hydrochloric acid for 15 minutes, gently wipe the platinum ring of the residual chlorine electrode with a cotton swab, and then wash it with tap water.
1595 +* Organic or oil pollution: immerse the electrode in tap water with a small amount of detergent, such as dishwashing liquid, and thoroughly clean the sensing surface of the electrode sensor. Gently wipe the platinum ring of the electrode with a cotton swab, then rinse with tap water, and the cleaning is complete. If the platinum ring of the electrode has formed an oxide film, please use toothpaste or 1000-grit fine sandpaper to properly polish the sensing surface, and then clean it with tap water. The platinum ring is connected to the glass, so please handle it carefully when polishing.
1596 +The electrode has a service life of about one year, and a new electrode should be replaced in time after aging.
1597 +* Before the cable plug and the device plug are locked, do not put the plug part into water.
1598 +
1557 1557  = 5. Battery & Power Consumption =
1558 1558  
1559 1559  
... ... @@ -1562,9 +1562,14 @@
1562 1562  [[**Battery Info & Power Consumption Analyze**>>url:http://wiki.dragino.com/xwiki/bin/view/Main/How%20to%20calculate%20the%20battery%20life%20of%20Dragino%20sensors%3F/]] .
1563 1563  
1564 1564  
1565 -= 6. Firmware update =
1607 += 6. FAQ =
1566 1566  
1567 1567  
1610 +
1611 +
1612 += 7. Firmware update =
1613 +
1614 +
1568 1568  User can change device firmware to:
1569 1569  
1570 1570  * Update with new features.
... ... @@ -1579,7 +1579,7 @@
1579 1579  
1580 1580  * Update through UART TTL interface : **[[Instruction>>url:http://wiki.dragino.com/xwiki/bin/view/Main/UART_Access_for_NB_ST_BC660K-GL/#H4.2UpdateFirmware28Assumethedevicealreadyhaveabootloader29]]**.
1581 1581  
1582 -= 7. ​Packing Info =
1629 += 8. ​Packing Info =
1583 1583  
1584 1584  
1585 1585  (% style="color:#037691" %)**Package Includes**:
... ... @@ -1598,7 +1598,7 @@
1598 1598  
1599 1599  * Weight / pcs : 180g
1600 1600  
1601 -= 8. Sensors =
1648 += 9. Sensors =
1602 1602  
1603 1603  
1604 1604  (% border="1" cellspacing="3" style="background-color:#f2f2f2; width:500px" %)
... ... @@ -1610,7 +1610,7 @@
1610 1610  |(% style="width:462px" %)Dissolved Oxygen Sensor|(% style="width:120px" %)DR-DO1
1611 1611  |(% style="width:462px" %)Turbidity Sensor|(% style="width:120px" %)DR-TS1
1612 1612  
1613 -= 9. Support =
1660 += 10. Support =
1614 1614  
1615 1615  
1616 1616  * Support is provided Monday to Friday, from 09:00 to 18:00 GMT+8. Due to different timezones we cannot offer live support. However, your questions will be answered as soon as possible in the before-mentioned schedule.