<
From version < 104.1 >
edited by Bei Jinggeng
on 2024/03/30 17:44
To version < 107.1 >
edited by Bei Jinggeng
on 2024/08/02 17:02
>
Change comment: There is no comment for this version

Summary

Details

Page properties
Content
... ... @@ -1244,7 +1244,7 @@
1244 1244  In the case if user can see the console output but can't type input to the device. Please check if you already include the (% style="color:green" %)**ENTER**(%%) while sending out the command. Some serial tool doesn't send (% style="color:green" %)**ENTER**(%%) while press the send key, user need to add ENTER in their string.
1245 1245  
1246 1246  
1247 -== 5.Possible reasons why the device is unresponsive: ==
1247 +== 7.2 Possible reasons why the device is unresponsive: ==
1248 1248  
1249 1249  
1250 1250  ​1. Check whether the battery voltage is lower than 2.8V
... ... @@ -1256,9 +1256,46 @@
1256 1256  [[image:image-20240330174406-3.png]]
1257 1257  ​
1258 1258  
1259 -(% class="wikigeneratedid" %)
1260 -= =
1259 +== 7.3 The node cannot read the sensor data ==
1261 1261  
1261 +This may be caused by a software firmware(≤1.1.6 version) bug, which we fixed in the latest firmware (>1.0.0 version)
1262 +
1263 +The user can fix this problem via upgrade firmware.
1264 +
1265 +
1266 +By default, The latest firmware value of POWERIC is 1, while the 3322 version requires POWERIC to be set to 0 in order to function properly
1267 +
1268 +
1269 +* **//1. Check if the hardware version is 3322//**
1270 +
1271 +If the sensor hardware version is 3322 or earlier, the user can change the POWERIC value to 0 after a firmware upgrade using one of the following methods
1272 +
1273 +**a. Using AT command**
1274 +
1275 +(% class="box infomessage" %)
1276 +(((
1277 +AT+POWERIC=0.
1278 +)))
1279 +
1280 +
1281 +**b. Using Downlink**
1282 +
1283 +(% class="box infomessage" %)
1284 +(((
1285 +FF 00(AT+POWERIC=0).
1286 +)))
1287 +
1288 +
1289 +[[image:http://wiki.dragino.com/xwiki/bin/download/Main/User%20Manual%20for%20LoRaWAN%20End%20Nodes/LDDS75%20-%20LoRaWAN%20Distance%20Detection%20Sensor%20User%20Manual/WebHome/image-20240531090837-1.png?rev=1.1||alt="image-20240531090837-1.png"]]
1290 +
1291 +
1292 +Please check your hardware production date
1293 +
1294 +The first two digits are the week of the year, and the last two digits are the year.
1295 +
1296 +The number 3322 is the first batch we changed the power IC.
1297 +
1298 +
1262 1262  = 8. Order Info =
1263 1263  
1264 1264  
Copyright ©2010-2024 Dragino Technology Co., LTD. All rights reserved
Dragino Wiki v2.0