... ... @@ -1244,6 +1244,58 @@
1244 1244  In the case if user can see the console output but can't type input to the device. Please check if you already include the (% style="color:green" %)**ENTER**(%%) while sending out the command. Some serial tool doesn't send (% style="color:green" %)**ENTER**(%%) while press the send key, user need to add ENTER in their string.
1245 1245  
1246 1246  
1247 +== 7.2 Possible reasons why the device is unresponsive: ==
1248 +
1249 +
1250 +​1. Check whether the battery voltage is lower than 2.8V
1251 +2. Check whether the jumper of the device is correctly connected
1252 +
1253 +[[image:image-20240330174400-2.png]]
1254 +3. Check whether the switch here of the device is at the ISP(The switch can operate normally only when it is in RUN)
1255 +
1256 +[[image:image-20240330174406-3.png]]
1257 +​
1258 +
1259 +== 7.3 The node cannot read the sensor data ==
1260 +
1261 +This may be caused by a software firmware(≤1.1.6 version) bug, which we fixed in the latest firmware (>1.0.0 version)
1262 +
1263 +The user can fix this problem via upgrade firmware.
1264 +
1265 +
1266 +By default, The latest firmware value of POWERIC is 1, while the 3322 version requires POWERIC to be set to 0 in order to function properly
1267 +
1268 +
1269 +* **//1. Check if the hardware version is 3322//**
1270 +
1271 +If the sensor hardware version is 3322 or earlier, the user can change the POWERIC value to 0 after a firmware upgrade using one of the following methods
1272 +
1273 +**a. Using AT command**
1274 +
1275 +(% class="box infomessage" %)
1276 +(((
1277 +AT+POWERIC=0.
1278 +)))
1279 +
1280 +
1281 +**b. Using Downlink**
1282 +
1283 +(% class="box infomessage" %)
1284 +(((
1285 +FF 00(AT+POWERIC=0).
1286 +)))
1287 +
1288 +
1289 +[[image:http://wiki.dragino.com/xwiki/bin/download/Main/User%20Manual%20for%20LoRaWAN%20End%20Nodes/LDDS75%20-%20LoRaWAN%20Distance%20Detection%20Sensor%20User%20Manual/WebHome/image-20240531090837-1.png?rev=1.1||alt="image-20240531090837-1.png"]]
1290 +
1291 +
1292 +Please check your hardware production date
1293 +
1294 +The first two digits are the week of the year, and the last two digits are the year.
1295 +
1296 +The number 3322 is the first batch we changed the power IC.
1297 +
1298 +