401 401  User can do calibration for the probe. It is limited to use below pH buffer solution to calibrate: 4.00, 6.86, 9.18. When calibration, user need to clean the electrode and put the probe in the pH buffer solution to wait the value stable ( a new clean electrode might need max 24 hours to be stable).
402 402  
403 -After stable, user can use below command to calibrate.
404 404  
404 +(% style="color:#037691" %)**After stable, user can use below command to calibrate.**
405 +
405 405  [[image:image-20220607144936-3.png]]
406 406  
407 407  
409 +
408 408  (% style="color:#037691" %)**Calibration Payload**
409 409  
410 410  (% border="1" cellspacing="10" style="background-color:#ffffcc; width:510px" %)
... ... @@ -429,24 +429,24 @@
429 429  
430 430  User can also send 0x14 downlink command to poll the current calibration payload.
431 431  
432 -[[image:image-20220607145603-4.png]]
433 433  
435 +|**Downlink Control Type**|**FPort**|**Type Code**|**Downlink payload size(bytes)**
436 +|Get Calibration Version Info|Any|14|2
437 +
434 434  * Reply to the confirmation package: 14 01
435 435  * Reply to non-confirmed packet: 14 00
436 436  
441 +1.
442 +11. Frequency Plans
437 437  
438 -
439 -== 2.8 Frequency Plans ==
440 -
441 -(((
442 442  The LSPH01 uses OTAA mode and below frequency plans by default. If user want to use it with different frequency plan, please refer the AT command sets.
443 -)))
444 444  
446 +1.
447 +11.
448 +111. EU863-870 (EU868)
445 445  
446 -=== 2.8.1 EU863-870 (EU868) ===
450 +Uplink:
447 447  
448 -(% style="color:blue" %)**Uplink:**
449 -
450 450  868.1 - SF7BW125 to SF12BW125
451 451  
452 452  868.3 - SF7BW125 to SF12BW125 and SF7BW250
... ... @@ -466,7 +466,7 @@
466 466  868.8 - FSK
467 467  
468 468  
469 -(% style="color:blue" %)**Downlink:**
471 +Downlink:
470 470  
471 471  Uplink channels 1-9 (RX1)
472 472  
... ... @@ -473,31 +473,28 @@
473 473  869.525 - SF9BW125 (RX2 downlink only)
474 474  
475 475  
478 +1.
479 +11.
480 +111. US902-928(US915)
476 476  
477 -=== 2.8.2 US902-928(US915) ===
478 -
479 -(((
480 480  Used in USA, Canada and South America. Frequency band as per definition in LoRaWAN 1.0.3 Regional document.
481 -)))
482 482  
483 -(((
484 +
484 484  To make sure the end node supports all sub band by default. In the OTAA Join process, the end node will use frequency 1 from sub-band1, then frequency 1 from sub-band2, then frequency 1 from sub-band3, etc to process the OTAA join.
485 -)))
486 486  
487 -(((
487 +
488 488  After Join success, the end node will switch to the correct sub band by:
489 -)))
490 490  
491 491  * Check what sub-band the LoRaWAN server ask from the OTAA Join Accept message and switch to that sub-band
492 492  * Use the Join successful sub-band if the server doesn’t include sub-band info in the OTAA Join Accept message ( TTN v2 doesn't include)
493 493  
493 +1.
494 +11.
495 +111. CN470-510 (CN470)
494 494  
495 -
496 -=== 2.8.3 CN470-510 (CN470) ===
497 -
498 498  Used in China, Default use CHE=1
499 499  
500 -(% style="color:blue" %)**Uplink:**
499 +Uplink:
501 501  
502 502  486.3 - SF7BW125 to SF12BW125
503 503  
... ... @@ -516,7 +516,7 @@
516 516  487.7 - SF7BW125 to SF12BW125
517 517  
518 518  
519 -(% style="color:blue" %)**Downlink:**
518 +Downlink:
520 520  
521 521  506.7 - SF7BW125 to SF12BW125
522 522  
... ... @@ -537,31 +537,26 @@
537 537  505.3 - SF12BW125 (RX2 downlink only)
538 538  
539 539  
539 +1.
540 +11.
541 +111. AU915-928(AU915)
540 540  
541 -=== 2.8.4 AU915-928(AU915) ===
542 -
543 -(((
544 544  Frequency band as per definition in LoRaWAN 1.0.3 Regional document.
545 -)))
546 546  
547 -(((
545 +
548 548  To make sure the end node supports all sub band by default. In the OTAA Join process, the end node will use frequency 1 from sub-band1, then frequency 1 from sub-band2, then frequency 1 from sub-band3, etc to process the OTAA join.
549 -)))
550 550  
551 -(((
552 -
553 -)))
554 554  
555 -(((
556 556  After Join success, the end node will switch to the correct sub band by:
557 -)))
558 558  
559 559  * Check what sub-band the LoRaWAN server ask from the OTAA Join Accept message and switch to that sub-band
560 560  * Use the Join successful sub-band if the server doesn’t include sub-band info in the OTAA Join Accept message ( TTN v2 doesn't include)
561 561  
562 -=== 2.8.5 AS920-923 & AS923-925 (AS923) ===
554 +1.
555 +11.
556 +111. AS920-923 & AS923-925 (AS923)
563 563  
564 -(% style="color:blue" %)**Default Uplink channel:**
558 +**Default Uplink channel:**
565 565  
566 566  923.2 - SF7BW125 to SF10BW125
567 567  
... ... @@ -568,13 +568,12 @@
568 568  923.4 - SF7BW125 to SF10BW125
569 569  
570 570  
571 -(% style="color:blue" %)**Additional Uplink Channel**:
565 +**Additional Uplink Channel**:
572 572  
573 573  (OTAA mode, channel added by JoinAccept message)
574 574  
569 +**AS920~~AS923 for Japan, Malaysia, Singapore**:
575 575  
576 -(% style="color:blue" %)**AS920~~AS923 for Japan, Malaysia, Singapore**:
577 -
578 578  922.2 - SF7BW125 to SF10BW125
579 579  
580 580  922.4 - SF7BW125 to SF10BW125
... ... @@ -588,7 +588,7 @@
588 588  922.0 - SF7BW125 to SF10BW125
589 589  
590 590  
591 -(% style="color:blue" %)**AS923 ~~ AS925 for Brunei, Cambodia, Hong Kong, Indonesia, Laos, Taiwan, Thailand, Vietnam**:
584 +**AS923 ~~ AS925 for Brunei, Cambodia, Hong Kong, Indonesia, Laos, Taiwan, Thailand, Vietnam**:
592 592  
593 593  923.6 - SF7BW125 to SF10BW125
594 594  
... ... @@ -603,18 +603,20 @@
603 603  924.6 - SF7BW125 to SF10BW125
604 604  
605 605  
606 -(% style="color:blue" %)**Downlink:**
607 607  
600 +**Downlink:**
601 +
608 608  Uplink channels 1-8 (RX1)
609 609  
610 610  923.2 - SF10BW125 (RX2)
611 611  
612 612  
607 +1.
608 +11.
609 +111. KR920-923 (KR920)
613 613  
614 -=== 2.8.6 KR920-923 (KR920) ===
611 +Default channel:
615 615  
616 -(% style="color:blue" %)**Default channel:**
617 -
618 618  922.1 - SF7BW125 to SF12BW125
619 619  
620 620  922.3 - SF7BW125 to SF12BW125
... ... @@ -622,7 +622,7 @@
622 622  922.5 - SF7BW125 to SF12BW125
623 623  
624 624  
625 -(% style="color:blue" %)**Uplink: (OTAA mode, channel added by JoinAccept message)**
620 +Uplink: (OTAA mode, channel added by JoinAccept message)
626 626  
627 627  922.1 - SF7BW125 to SF12BW125
628 628  
... ... @@ -639,7 +639,7 @@
639 639  923.3 - SF7BW125 to SF12BW125
640 640  
641 641  
642 -(% style="color:blue" %)**Downlink:**
637 +Downlink:
643 643  
644 644  Uplink channels 1-7(RX1)
645 645  
... ... @@ -646,9 +646,11 @@
646 646  921.9 - SF12BW125 (RX2 downlink only; SF12BW125 might be changed to SF9BW125)
647 647  
648 648  
649 -=== 2.8.7 IN865-867 (IN865) ===
644 +1.
645 +11.
646 +111. IN865-867 (IN865)
650 650  
651 -(% style="color:blue" %)**Uplink:**
648 +Uplink:
652 652  
653 653  865.0625 - SF7BW125 to SF12BW125
654 654  
... ... @@ -657,7 +657,7 @@
657 657  865.9850 - SF7BW125 to SF12BW125
658 658  
659 659  
660 -(% style="color:blue" %)**Downlink:**
657 +Downlink:
661 661  
662 662  Uplink channels 1-3 (RX1)
663 663  
... ... @@ -664,36 +664,45 @@
664 664  866.550 - SF10BW125 (RX2)
665 665  
666 666  
664 +1.
665 +11. LED Indicator
667 667  
668 -== 2.9 LED Indicator ==
669 -
670 670  The LSPH01 has an internal LED which is to show the status of different state.
671 671  
669 +
672 672  * The sensor is detected when the device is turned on, and it will flash 4 times quickly when it is detected.
673 673  * Blink once when device transmit a packet.
674 674  
675 -== 2.10 ​Firmware Change Log ==
673 +1.
674 +11. ​Firmware Change Log
676 676  
677 -
678 678  **Firmware download link:**
679 679  
680 680  [[http:~~/~~/www.dragino.com/downloads/index.pHp?dir=LoRa_End_Node/LSPH01/Firmware/>>url:http://www.dragino.com/downloads/index.php?dir=LoRa_End_Node/LSE01/Firmware/]]
681 681  
682 682  
683 -**Firmware Upgrade Method: **[[Firmware Upgrade Instruction>>doc:Main.Firmware Upgrade Instruction for STM32 base products.WebHome]]
681 +**Firmware Upgrade Method:**
684 684  
683 +[[http:~~/~~/wiki.dragino.com/index.pHp?title=Firmware_Upgrade_Instruction_for_STM32_base_products#Introduction>>url:http://wiki.dragino.com/index.php?title=Firmware_Upgrade_Instruction_for_STM32_base_products#Introduction]]
685 685  
686 686  
687 -= 3. Configure LSPH01 via AT Command or LoRaWAN Downlink =
688 688  
687 +
688 +
689 +
690 +1. Configure LSPH01 via AT Command or LoRaWAN Downlink
691 +
689 689  Use can configure LSPH01 via AT Command or LoRaWAN Downlink.
690 690  
691 -* AT Command Connection: See [[FAQ>>||anchor="H6.FAQ"]].
692 -* LoRaWAN Downlink instruction for different platforms: [[IoT LoRaWAN Server>>doc:Main.WebHome]]
694 +* AT Command Connection: See [[FAQ>>path:#AT_COMMAND]].
695 +* LoRaWAN Downlink instruction for different platforms:
693 693  
697 +[[http:~~/~~/wiki.dragino.com/index.pHp?title=Main_Page#Use_Note_for_Server>>url:http://wiki.dragino.com/index.php?title=Main_Page#Use_Note_for_Server]]
698 +
699 +
694 694  There are two kinds of commands to configure LSPH01, they are:
695 695  
696 -* (% style="color:#4f81bd" %)** General Commands**.
702 +* **General Commands**.
697 697  
698 698  These commands are to configure:
699 699  
... ... @@ -700,25 +700,25 @@
700 700  * General system settings like: uplink interval.
701 701  * LoRaWAN protocol & radio related command.
702 702  
703 -They are same for all Dragino Device which support DLWS-005 LoRaWAN Stack. These commands can be found on the wiki:[[End Device AT Commands and Downlink Command>>doc:Main.End Device AT Commands and Downlink Command.WebHome]]
709 +They are same for all Dragino Device which support DLWS-005 LoRaWAN Stack. These commands can be found on the wiki:
704 704  
711 +[[http:~~/~~/wiki.dragino.com/index.pHp?title=End_Device_Downlink_Command>>url:http://wiki.dragino.com/index.php?title=End_Device_Downlink_Command]]
705 705  
706 -* (% style="color:#4f81bd" %)** Commands special design for LSPH01**
707 707  
714 +* **Commands special design for LSPH01**
715 +
708 708  These commands only valid for LSPH01, as below:
709 709  
710 710  
719 +1.
720 +11. Set Transmit Interval Time
711 711  
712 -== 3.1 Set Transmit Interval Time ==
713 -
714 714  Feature: Change LoRaWAN End Node Transmit Interval.
715 715  
716 -(% style="color:#037691" %)**AT Command: AT+TDC**
724 +**AT Command: AT+TDC**
717 717  
718 -
719 -(% border="1" cellspacing="10" style="background-color:#ffffcc; color:green; width:710px" %)
720 -|(% style="width:154px" %)**Command Example**|(% style="width:223px" %)**Function**|(% style="width:330px" %)**Response**
721 -|(% style="width:154px" %)AT+TDC=?|(% style="width:223px" %)Show current transmit Interval|(% style="width:330px" %)(((
726 +|**Command Example**|**Function**|**Response**
727 +|AT+TDC=?|Show current transmit Interval|(((
722 722  30000
723 723  
724 724  OK
... ... @@ -725,13 +725,13 @@
725 725  
726 726  the interval is 30000ms = 30s
727 727  )))
728 -|(% style="width:154px" %)AT+TDC=60000|(% style="width:223px" %)Set Transmit Interval|(% style="width:330px" %)(((
734 +|AT+TDC=60000|Set Transmit Interval|(((
729 729  OK
730 730  
731 731  Set transmit interval to 60000ms = 60 seconds
732 732  )))
733 733  
734 -(% style="color:#037691" %)**Downlink Command: 0x01**
740 +**Downlink Command: 0x01**
735 735  
736 736  Format: Command Code (0x01) followed by 3 bytes time value.
737 737  
... ... @@ -740,18 +740,32 @@
740 740  * Example 1: Downlink Payload: 0100001E ~/~/ Set Transmit Interval (TDC) = 30 seconds
741 741  * Example 2: Downlink Payload: 0100003C ~/~/ Set Transmit Interval (TDC) = 60 seconds
742 742  
749 +1.
750 +11. Set Interrupt Mode
743 743  
744 -== 3.2 Set Interrupt Mode ==
745 -
746 746  Feature, Set Interrupt mode for GPIO_EXIT.
747 747  
748 -(% style="color:#037691" %)**AT Command: AT+INTMOD**
754 +**AT Command: AT+INTMOD**
749 749  
750 -[[image:image-20220607153759-6.png]]
756 +|**Command Example**|**Function**|**Response**
757 +|AT+INTMOD=?|Show current interrupt mode|(((
758 +0
751 751  
760 +OK
752 752  
753 -(% style="color:#037691" %)**Downlink Command: 0x06**
762 +the mode is 0 = No interruption
763 +)))
764 +|AT+INTMOD=2|(((
765 +Set Transmit Interval
754 754  
767 +1. (Disable Interrupt),
768 +1. (Trigger by rising and falling edge),
769 +1. (Trigger by falling edge)
770 +1. (Trigger by rising edge)
771 +)))|OK
772 +
773 +**Downlink Command: 0x06**
774 +
755 755  Format: Command Code (0x06) followed by 3 bytes.
756 756  
757 757  This means that the interrupt mode of the end node is set to 0x000003=3 (rising edge trigger), and the type code is 06.
... ... @@ -759,21 +759,25 @@
759 759  * Example 1: Downlink Payload: 06000000 ~/~/ Turn off interrupt mode
760 760  * Example 2: Downlink Payload: 06000003 ~/~/ Set the interrupt mode to rising edge trigger
761 761  
782 +1.
783 +11. Calibrate Sensor
762 762  
763 -== 3.3 Calibrate Sensor ==
785 +Detail See [[Calibration Guide>>path:#Calibration]] for the user of 0x13 and 0x14 downlink commands
764 764  
765 -Detail See [[Calibration Guide>>||anchor="2.7Calibration"]] for the user of 0x13 and 0x14 downlink commands
766 766  
767 767  
789 +1.
790 +11. Get Firmware Version Info
768 768  
769 -== 3.4 Get Firmware Version Info ==
770 -
771 771  Feature: use downlink to get firmware version.
772 772  
773 -(% style="color:#037691" %)**Downlink Command: 0x26**
774 774  
775 -[[image:image-20220607154718-7.png]]
795 +**Downlink Command: 0x26**
776 776  
797 +
798 +|**Downlink Control Type**|**FPort**|**Type Code**|**Downlink payload size(bytes)**
799 +|Get Firmware Version Info|Any|26|2
800 +
777 777  * Reply to the confirmation package: 26 01
778 778  * Reply to non-confirmed packet: 26 00
779 779  
... ... @@ -781,10 +781,11 @@
781 781  
782 782  Configures info payload:
783 783  
784 -(% border="1" cellspacing="10" style="background-color:#ffffcc; color:green; width:510px" %)
785 -|=(((
786 -**Size(bytes)**
787 -)))|=**1**|=**1**|=**1**|=**1**|=**1**|=**5**|=**1**
808 +|(((
809 +**Size**
810 +
811 +**(bytes)**
812 +)))|**1**|**1**|**1**|**1**|**1**|**5**|**1**
788 788  |**Value**|Software Type|(((
789 789  Frequency
790 790  
... ... @@ -794,7 +794,10 @@
794 794  
795 795  Version
796 796  )))|Sensor Type|Reserve|(((
797 -[[Message Type>>||anchor="H2.3.6MessageType"]]
822 +[[Message>>path:#Message_Type]]
823 +
824 +[[Type>>path:#Message_Type]]
825 +
798 798  Always 0x02
799 799  )))
800 800  
image-20220607153237-5.png
Author
... ... @@ -1,1 +1,0 @@
1 -XWiki.Xiaoling
Size
... ... @@ -1,1 +1,0 @@
1 -12.9 KB
Content
image-20220607153759-6.png
Author
... ... @@ -1,1 +1,0 @@