... ... @@ -58,6 +58,7 @@
58 58  * IP66 Waterproof Enclosure
59 59  * 4000mAh or 8500mAh Battery for long term use
60 60  
61 +
61 61  == 1.3 Specification ==
62 62  
63 63  Measure Volume: Base on the centra pin of the probe, a cylinder with 7cm diameter and 10cm height.
... ... @@ -89,7 +89,7 @@
89 89  )))
90 90  
91 91  (((
92 -In case you can’t set the OTAA keys in the LoRaWAN OTAA server, and you have to use the keys from the server, you can [[use AT Commands >>||anchor="H3.UsingtheATCommands"]].
93 +In case you can’t set the OTAA keys in the LoRaWAN OTAA server, and you have to use the keys from the server, you can [[use AT Commands >>||anchor="H3.200BUsingtheATCommands"]].
93 93  )))
94 94  
95 95  
... ... @@ -630,6 +630,8 @@
630 630  * Solid ON for 5 seconds once device successful Join the network.
631 631  * Blink once when device transmit a packet.
632 632  
634 +
635 +
633 633  == 2.9 Installation in Soil ==
634 634  
635 635  **Measurement the soil surface**
... ... @@ -947,7 +947,9 @@
947 947  
948 948  (% style="color:#4f81bd" %)**Cause for this issue:**
949 949  
953 +(((
950 950  The fuse on LSE01 is not large enough, some of the soil probe require large current up to 5v 800mA, in a short pulse. When this happen, it cause the device reboot so user see rejoin.
955 +)))
951 951  
952 952  
953 953  (% style="color:#4f81bd" %)**Solution: **
... ... @@ -954,7 +954,7 @@
954 954  
955 955  All new shipped LSE01 after 2020-May-30 will have this to fix. For the customer who see this issue, please bypass the fuse as below:
956 956  
957 -[[image:1654500929571-736.png]]
962 +[[image:1654500929571-736.png||height="458" width="832"]]
958 958  
959 959  
960 960  = 6. ​Order Info =